Electrode Offset Settings - Thermo Scientific ALPHA PH 550 Manual

Monitor ph / orp
Table of Contents

Advertisement

5.2

Electrode Offset Settings

NOTE:
'Electrode Offset Setting' is not available when the pH Monitor is
configured for ORP measurement mode. (Refer Section 5.5 for
switching measurement modes)
In applications where continuous pH measurement is required, it may not be
convenient to remove the electrode for calibration. In such cases, an on-line offset
adjustment is recommended. The pH Monitor allows you set an offset of up to ±
2.00pH to compensate for errors in the pH electrode.
The pH Monitor adds or subtracts the offset value from the measured pH value and
displays the corrected value. However, if you need to offset the value beyond the
average offset you would expect in your application type, consider a full calibration
or even electrode replacement.
To Determine the Offset Value:
1.
Take a sample from the liquid of pH measurement. Record the pH reading
of the pH Monitor at the time the sample was taken.
2.
Measure the pH value of your sample using another calibrated pH meter
having its own electrode (pH tester, hand-held meter or bench meter)
3.
The difference between these two readings is the offset
SETUP
From pH measurement mode press ENT key to enter setup mode as
1 1
described in section 5.1. The LCD shows the first screen of setup mode
(OFS). Press ENT key to access electrode offset setting (OFS).
The upper display shows the last configured offset value (if any), otherwise
2
zero. The lower display shows currently measured pH reading (including
last configured offset value). Allow the reading to stabilize. LCD shows
'READY' annunciator when the reading is stable.
Press ▲ or ▼ key to set the upper display to the newly calculated offset
value.
As the upper display value (offset) changes, the pH Monitor adjust the lower
3
reading automatically to suit the new offset value. Up to ± 2.00 pH offset is
1
SETUP
ENT
READY
2
pH
pH
- 24 -
3
SETUP
pH
READY
pH
ENT