Page 1
Connection box 1-Phase A 1626 Connection box 3-Phase A 1627 Universal Connection box A 1688 Instruction manual Version 2.1.1, Code No. 20 753 084...
Page 2
® ® The trade names Metrel , Smartec , Eurotest , Auto Sequence are trademarks registered worldwide. No part of this publication may be reproduced or utilized in any form or by any means without permission in writing from METREL.
General description ....................... 4 Warnings and notes ..................... 4 1.1.1 General safety warnings ..................4 1.1.2 Safety warnings related to operation of A 1626, A 1627, and A 1688 ...... 4 1.1.3 Markings on the instrument ..................5 Standards applied ......................5 Connection box sets and accessories .................
Connection box are units intended for connection of various tested appliances to the A 1600 CE Switch 5kV in test setups with the MI 3394 XS. Each one of the A 1626, the A 1627, and A 1688 has unique code identifier. This unique code determines predefined characteristics for the applied connection box, like its configuration and some limiting values rated to the connection box;...
EU regulations. This equipment should be recycled as electronic waste. 1.2 Standards applied The A 1626, A 1627, and A 1688 Connection boxes are manufactured and tested according to the following regulations, listed below. Electromagnetic compatibility (EMC) void...
A 1626, A 1627, A 1688 Connection boxes Connection box sets and accessories 2 Connection box sets and accessories 2.1 Standard set for the A 1626 Connection box 1-Phase A 1626 Connection box 1-Phase Shorting plug (3 pieces) ...
A 1626, A 1627, A 1688 Connection boxes Connection boxes description 3 Connection boxes description Figure 3.1 Universal Connection box (A 1688)
Page 8
A 1626, A 1627, A 1688 Connection boxes Connection boxes description Figure 3.2 Connection box 3-Phase (A 1627)
Page 9
A 1626, A 1627, A 1688 Connection boxes Connection boxes description Figure 3.3 Connection box 1-Phase (A 1626)
Page 10
Note: Test sockets (2) and (3) are connected in parallel to wiring connections (9) in case the shorting jumpers (8) are inserted. (A 1626, A 1627) In case the shorting jumpers (8) are inserted, next connections are enabled: L connected to HV(~,+) / channel 1, N connected to HV(~,+) / channel 2, PE connected to HV(~,-) / channel 1.
Installing the Connection box 4 Installing the Connection box Connection box (A 1626, A 1627, or A 1688) can be no-fixed as desktop unit or can be fixed in proper place inside the test chamber. Consider prohibition zone and other requirements of EN 50191.
5 Operation of the Connection boxes 5.1 Wiring principle in the Connection boxes The A 1626, A 1627, and A 1688 Connection boxes are passive devices used as a wiring interface for tested equipment connection set-up. Figure 5.1 Wiring principle in A 1626, A 1627 Wiring principle is the same for both the A 1626 and the A1627 Connection boxes with omitted items in the A 1626.
5.1.1 Application examples for the Connection boxes The following figures show typical test setups for application of the A 1626, the A 1627, and the A 1688 Connection boxes. Optional tip commander A 1511 can only be used in setups, with light...
Page 14
Operation of the Connection boxes Figure 5.3 Application example for the A 1627 with the tip commander A 1511 in the test environment protected by the light barrier Figure 5.4 Application example for the A 1626 in the test chamber...
Page 15
A 1626, A 1627, A 1688 Connection boxes Operation of the Connection boxes Figure 5.5 Application example for the A 1688 with the tip commander A 1511 in the test environment protected by the light barrier Figure 5.6 Application example for the A 1688 in the test chamber Notes: ...
Operation of the Connection boxes 5.1.2 Tip commander interface The tip commander A 1511 is intended for optional application with the A 1626, the A 1627, and A 1688 Connection boxes. Connections are as follows. Figure 5.7: Commander - pin layout and test connections...
6.2 Cleaning Use a soft, slightly moistened cloth with soap water or alcohol to clean the surface of the A 1626, A 1627, or A 1688 Connection boxes. Leave the Connection box to dry totally before using it.
A 1626, A 1627, A 1688 Connection boxes Technical specifications 7 Technical specifications Parameters defined with built-in identification Parameter A 1626 A 1627 A 1688 Maximum a.c. 4 kV 4 kV 5.1 kV withstanding voltage Maximum d.c. 4 kV 4 kV...