  • Page 1 METREL test and measurement accessories Discharge Link Adapter A 1513 Instruction manual Version 1.1.1, Code no. 20 752 541...
  • Page 2 Mark on your equipment certifies that this equipment meets the requirements of the all applicable EU (European Union) regulations © 2016 METREL No part of this publication may be reproduced or utilized in any form or by any means without permission in writing from METREL.
  • Page 3: Table Of Contents

    SAFETY AND OPERATIONAL CONSIDERATIONS..........2 ..................2 ARNINGS AND NOTE A 1513 .................. 3 ARKINGS ON THE .................... 3 TANDARD APPLIED WORKING WITH A 1513 ..................4 ..................... 4 ESCRIPTION A 1513 .................... 4 OPERATION ................5 XAMPLE OF DISCHARGING TIME MAINTENANCE ......................
  • Page 4: Preface

    1 Preface Congratulations for purchasing and using METREL A 1513 Discharge Link accessory. The A 1513 Discharge link is optional accessory intended for manual discharging of capacitive objects. This unit contains a high ohm discharge resistor (10 kV) enclosed in double insulation plastic package.
  • Page 5: Safety And Operational Considerations

    2 Safety and operational considerations 2.1 Warnings and note In order to reach high level of operator’s safety during application of the A 1513 Discharge Link adapter as well as to keep the test accessory and equipment undamaged, it is necessary to consider the following general warnings: symbol on the A 1513 means »Read the Instruction manual with special...
  • Page 6: Markings On Thea 1513

    EMC, LVD, and ROHS regulations. This equipment should be recycled as electronic waste. 2.3 Standard applied The A 1513 Discharge Link Adapter is manufactured and tested according to the following standard, listed below. Safety (LVD) Safety requirements for electrical equipment for measurement, control, and laboratory use –...
  • Page 7: Working With A 1513

    Guarde A 1513 Figure 3.2: Manual discharge example Step 1 After the Insulation measurement, connect the measuring object to the A 1513 Discharge Link terminals. Use individual protective measures. Step 2 Wait until the visual Discharge indication stops shining. Step 3...
  • Page 8: Example Of Discharging Time

    3.3 Example of discharging time Device under test (DUT) was charged to U and has an internal capacity of C How fast will the A 1513 discharge the internal capacity of the DUT to the safe voltage value (U < 50 V)? Discharging resistance R of the A 1513 is 100 k.
  • Page 9: Maintenance

    A 1513 Discharge Link. 4.1 Cleaning No special maintenance is required for the housing. To clean the surface of the A 1513 Discharge Link adapter, use a soft cloth slightly moistened with soapy water or alcohol.
  • Page 10: Technical Specifications

    A 1513 Discharge Link Technical specifications 5 Technical specifications Discharge resistors: Discharging rate for capacitive load ....< 0.6 s / µF at 10 kV Discharging resistance ........100 k  15 % Max discharging capacitance ......10 μF at 10 kV Max discharging current .......