Download Print this page

Advertisement

Quick Links

The PGAxxxEVM-034 supplies a platform to test the PGA900, PGA300, and PGA305 in the QFN
package. The EVM is shipped as either the PGA900EVM-034, the PGA300EVM-034, or the
PGA305EVM-034, with the name identifying the signal conditioner included in the EVM package. The
PGAxxxEVM-034 supplies an interchangeable platform all of these devices can use.
...................................................................................................................
1
Introduction
2
Default Configuration
3
Inputs and Output Configurations
3.1
Pressure Input
3.2
Temperature Input
3.3
Output
3.4
Voltage Mode
3.5
Current Mode
3.6
OWI and SPI/I
..........................................................................................................................
4
OWI
4.1
Activation Pulse
4.2
Data Sent Through OWI
5
XDS200 and USB2ANY Connectors
6
Power Supplies in the PGAxxxEVM-034
..................................................................................................................
7
Schematics
1
PGAxxxEVM-034
2
Pressure Stimulus in the PGAxxxEVM-034
3
PGA900 in Voltage Mode Configuration
4
PGAxxxEVM-034 in Current Mode
5
Additional Loop Current in 4-mA to 20-mA Mode When Using SPI/I2C/SWD as Digital Interface
6
Additional Loop Current in 4-mA to 20-mA Mode When Using OWI as Digital Interface
7
OWI Block Diagram
8
OWI Activation Pulse for the PGA900 Generated by the PGAxxxEVM-034 in Voltage Mode
9
OWI Data at 320 bps; Oscilloscope Probe is Connected at TP20 in the PGAxxxEVM-034
10
XDS200 Emulator Connection to the PGAxxxEVM-034
11
USB2ANY Connection to the PGAxxxEVM-034
...................................................................................................................
12
USB2ANY
13
USB2ANY Pinout
14
Power Distribution in PGAxxxEVM-034
15
PGAxxxEVM-034 Main Schematic
16
Input and Output Schematic
17
USB2ANY Schematic
18
OWI Activation Pulse and Data Schematic
19
OWI Power Amplifier Schematic
20
Power Supplies Schematic
SLDU011B – May 2015 – Revised August 2018
Submit Documentation Feedback
........................................................................................................
.........................................................................................
.......................................................................................................
..................................................................................................
................................................................................................................
........................................................................................................
........................................................................................................
2
C/SWD in 4-mA to 20-mA Current Loop
....................................................................................................
..........................................................................................
....................................................................................
................................................................................
............................................................................................................
..................................................................................
........................................................................................
........................................................................................................
...........................................................................................................
.................................................................................
.......................................................................................
..............................................................................................
......................................................................................................
.........................................................................................
................................................................................................
Copyright © 2015–2018, Texas Instruments Incorporated
SLDU011B – May 2015 – Revised August 2018
PGAxxxEVM-034 User Guide
Contents
......................................................
List of Figures
..............................................................................
..............................................................
.......................................................................
.............................................................................
User's Guide
..............
..........................
...................
.....................
PGAxxxEVM-034 User Guide
2
3
4
4
4
4
5
6
7
10
11
12
13
16
17
2
4
5
6
8
9
10
11
12
13
14
14
15
16
17
18
19
20
21
22
1

Advertisement

loading
Need help?

Need help?

Do you have a question about the PGA EVM-034 Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Texas Instruments PGA EVM-034 Series

  • Page 1 Input and Output Schematic ...................... USB2ANY Schematic ................OWI Activation Pulse and Data Schematic ..................OWI Power Amplifier Schematic ....................Power Supplies Schematic SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 2 The application components section only lists the required components for real-life applications using the PGA900/PGA300/PGA305 device. In this case, the application is configured for current mode. PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 3 Closed Connect ASIC_GND to IRETURN Pins 1-2 closed Connect VDD cap to ASIC_GND Pins 2-3 closed Connect USBGND to IRETURN SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 4 The PGAxxxEVM-034 can be configured for voltage or current mode (4-mA to 20-mA loop). Each mode requires different jumper settings as shown in the following sections. PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 5 NOTE: If the designer wants the VDD to be higher than 5 V in voltage mode, open J12 and close J13. VDD will be equal to the power applied to the PGAxxxEVM-034. SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 6 NOTE: When in current mode, the input voltage to the EVM (J20 and J21) should be at least 20 V. PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 7 Disconnect SPI/UART pins between USB2ANY and PGA900 The PGA300 and PGA305 are prepared for OWI communication and do not require additional device configuration. SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 8 J16: Open ASICGND IRETURN Figure 5. Additional Loop Current in 4-mA to 20-mA Mode When Using SPI/I2C/SWD as Digital Interface PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 9 J16: Open ASICGND IRETURN Figure 6. Additional Loop Current in 4-mA to 20-mA Mode When Using OWI as Digital Interface SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 10 5 ± 7.5 V for Activation INA138 16.2 NŸ 1 NŸ Offset Voltage Current Compensating Voltage Figure 7. OWI Block Diagram PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 11 PGA300 data sheet (SLDS204). Figure 8. OWI Activation Pulse for the PGA900 Generated by the PGAxxxEVM-034 in Voltage Mode SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 12 320 bps. Figure 9. OWI Data at 320 bps; Oscilloscope Probe is Connected at TP20 in the PGAxxxEVM-034 PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 13 Figure 10 shows the proper connection for the XDS200 emulator. Figure 10. XDS200 Emulator Connection to the PGAxxxEVM-034 SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 14 12, one enclosed and one open. The functionality is the same for both. Figure 11. USB2ANY Connection to the PGAxxxEVM-034 Figure 12. USB2ANY PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 15 A USB cable is included to connect the device to the PC. No external supply is necessary. Figure 13. USB2ANY Pinout SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 16 NOTE: J15 must be closed during OTP programming. This is only applicable for the PGA900. The PGA300 has no OTP functionality. PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 17 OPEN FOR 4-20 mA LOOP J16 - CLOSED FOR VOLTAGE OUTPUT SPI/I2C COMMUNICATION 0.1µF 0.1µF ASIC_GND IRETURN PGA300ARHHR ASIC_GND Figure 15. PGAxxxEVM-034 Main Schematic SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 18 Open J4, J5, J9 and J10 0.1µF 560pF ASIC_GND COMP COMP ASIC_GND 10.0 IRETURN Figure 16. Input and Output Schematic PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 19 J19 - USBGND TO IRETURN: OWI in 4-20 mA LOOP J19 - USBGND TO ASIC_GND: SPI/I2C/SWD in 4-20 mA LOOP USBGND ASIC_GND IRETURN Figure 17. USB2ANY Schematic SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 20 2.00k IRETURN TLC352ID OWI_TX 2.00k 1.00k V3P0 8.06k 3.0k IRETURN IRETURN IRETURN Figure 18. OWI Activation Pulse and Data Schematic PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 21 IRETURN SCL SDA V3P3 0.1µF 1000pF IRETURN SHDN 0.1µF IRETURN V3P3 IRETURN AD5280BRUZ50 IRETURN Figure 19. OWI Power Amplifier Schematic SLDU011B – May 2015 – Revised August 2018 PGAxxxEVM-034 User Guide Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 22 BYPASS 1µF 0.1µF 0.01µF 1µF TPS79133DBV TPS76950QDBVRG4Q1 10µF IRETURN IRETURN IRETURN IRETURN IRETURN IRETURN IRETURN Figure 20. Power Supplies Schematic PGAxxxEVM-034 User Guide SLDU011B – May 2015 – Revised August 2018 Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 23 • Updated designators for text in DAC Output Configurations in Figure 16 ..................... • Updated designator for J19 in Figure 17 SLDU011B – May 2015 – Revised August 2018 Revision History Submit Documentation Feedback Copyright © 2015–2018, Texas Instruments Incorporated...
  • Page 24 STANDARD TERMS FOR EVALUATION MODULES Delivery: TI delivers TI evaluation boards, kits, or modules, including any accompanying demonstration software, components, and/or documentation which may be provided together or separately (collectively, an “EVM” or “EVMs”) to the User (“User”) in accordance with the terms set forth herein.
  • Page 25 FCC Interference Statement for Class B EVM devices NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
  • Page 26 【無線電波を送信する製品の開発キットをお使いになる際の注意事項】 開発キットの中には技術基準適合証明を受けて いないものがあります。 技術適合証明を受けていないもののご使用に際しては、電波法遵守のため、以下のいずれかの 措置を取っていただく必要がありますのでご注意ください。 1. 電波法施行規則第6条第1項第1号に基づく平成18年3月28日総務省告示第173号で定められた電波暗室等の試験設備でご使用 いただく。 2. 実験局の免許を取得後ご使用いただく。 3. 技術基準適合証明を取得後ご使用いただく。 なお、本製品は、上記の「ご使用にあたっての注意」を譲渡先、移転先に通知しない限り、譲渡、移転できないものとします。 上記を遵守頂けない場合は、電波法の罰則が適用される可能性があることをご留意ください。 日本テキサス・イ ンスツルメンツ株式会社 東京都新宿区西新宿6丁目24番1号 西新宿三井ビル 3.3.3 Notice for EVMs for Power Line Communication: Please see http://www.tij.co.jp/lsds/ti_ja/general/eStore/notice_02.page 電力線搬送波通信についての開発キットをお使いになる際の注意事項については、次のところをご覧ください。http:/ /www.tij.co.jp/lsds/ti_ja/general/eStore/notice_02.page 3.4 European Union 3.4.1 For EVMs subject to EU Directive 2014/30/EU (Electromagnetic Compatibility Directive): This is a class A product intended for use in environments other than domestic environments that are connected to a low-voltage power-supply network that supplies buildings used for domestic purposes.
  • Page 27 Notwithstanding the foregoing, any judgment may be enforced in any United States or foreign court, and TI may seek injunctive relief in any United States or foreign court. Mailing Address: Texas Instruments, Post Office Box 655303, Dallas, Texas 75265 Copyright © 2018, Texas Instruments Incorporated...
  • Page 28 IMPORTANT NOTICE FOR TI DESIGN INFORMATION AND RESOURCES Texas Instruments Incorporated (‘TI”) technical, application or other design advice, services or information, including, but not limited to, reference designs and materials relating to evaluation modules, (collectively, “TI Resources”) are intended to assist designers who are developing applications that incorporate TI products;...
  • Page 29 Mouser Electronics Authorized Distributor Click to View Pricing, Inventory, Delivery & Lifecycle Information: Texas Instruments PGA305EVM-034...

This manual is also suitable for:

Pga900evm-034Pga300evm-034Pga305evm-034