Contents Important Notice Contents System Overview Safety Precautions System Preparation Intended Use Patient Set-Up Operation – Control Handset Operation – Modes of Operation Operation – Quick Twist CPR Operation – Transport Mode Operation – Deflation / Storage Care and Cleaning Troubleshooting Technical Specifications Warranty Statement...
System overview The ProCair Prime Advanced is an alternating A full-size, movable foam sheet can be mattress replacement system for the positioned at the bottom of the mattress, prevention and treatment of skin breakdown working in conjunction with the cell in cell air and pressure injuries in patients of high to very cells to provide bottoming-out protection.
Page 5
The system consists of the following components: Mattress replacement with integrated control unit, control handset, power cord and CPR release User Guide Carry bag CPR valve Head end It is recommended that all packing materials and the User Guide be kept in the carry bag, for Left ease of storage and/or transport.
This includes, Novis Healthcare recommends using the but is not limited to, the appropriate use cord retention loops on the mattress base of side rails to prevent falls.
Component failures may cause the unit Should the power cord be found to be worn to operate erratically, possibly producing or damaged, contact Novis Healthcare for a potential hazards to patient and carers. replacement. Explosion Hazard...
20°C. missing components should be reported to This helps ensure optimal performance and Novis Healthcare as soon as possible. prevents potential damage caused by sudden temperature changes from the storage temperature range of 5°C to 60°C.
Page 9
System preparation Check the internal quick release control unit air connectors are securely connected. Open the top cover by unzipping the right side of the mattress (zipper located at foot end), check the two connectors are secure by pushing the male and female ends together, there should be no movement.
Page 10
System preparation On the control handset, press and hold the Power button for a minimum of three seconds. The Power indicator will glow green to indicate the system is operational and automatically inflating. While inflating, the mode/pressure indicator will alternate between “oo” and “OO”, indicating mattress is being inflated to maximum pressure in preparation for use.
Page 11
System preparation Universal Therapy System™ Mattress Configuration Dynamic Mode Hybrid Mode The ProCair Prime Advanced is initially Hybrid mode allows greater comfort for the configured as a full dynamic mattress, the patient by adding a softly padded foam sheet patient is placed in direct contact with the air over the air cells to allow immersion of the cells for optimum pressure offloading during patient’s body over the mattress surface, while...
Instable spinal cord injury professional. Cervical traction Novis Healthcare accepts no liability for any use, change or assembly of the product other Intended care setting than that stated in this User Guide. Refer to our Warranty Statement for more details.
Patient set-up Once initial inflation is complete, a patient may be placed on the system. ‘Bottoming Out’ test Once the mattress is fully inflated, bedding can be replaced. Check system is in alternation mode by Secure sheets loosely enough to ensure ensuring the mode/pressure indicator is they do not interfere with cell alternation.
Operation Control handset Battery status Green LED lit The battery is fully charged Green LED The battery is charging flashing Amber LED The battery power is low flashing LED is off The battery is malfunctioning Mode/pressure indicator Normally displays the current mode of operation (AF, St, AL, FL).
Modes of operation Mode selection In Alternating Mode, alternate mattress cells inflate and deflate following a fixed cycle time of Press button to cycle between 14 minutes, with the exception of static head cells. modes, as indicated by the mode/ pressure indicator: Alternating mode is used for normal therapeutic Auto Firm (Max Inflate)
Operation Quick Twist CPR Transport function Rapid deflation of the mattress may be required To prepare for patient transport, for emergency treatment (or to pack up the unit). press the Mode button and select Static mode, wait 14 minutes to The Quick Twist CPR valve is located at the top ensure all cells are fully inflated.
Deflation and storage Power button Press the power button for a minimum of three seconds to switch off the control unit. Switch off mains power and unplug the power cord from the mains outlet. Open Turn the Quick Twist CPR to OPEN to release air and deflate all cells.
Care and cleaning To prevent cross-contamination, the Switch off and disconnect the control unit mattress should be examined and from mains power supply before cleaning. disinfected between patient use, and a liner Failure to do so could result in equipment or bed sheet is recommended for added damage or electric shock.
Page 19
Control Unit /Handset cleaning Top Cover Removal Disconnect control unit from mains power Raise the waterfall skirt and locate the before cleaning. Gently wipe down the external zippers at the foot end of the mattress. case with a soft cloth. Soak the cloth in warm water containing mild PH neutral detergent, and wring any excess water before gently wiping all external...
Troubleshooting Alert Codes An audible alert, accompanied by an Alert Code display, indicate the control unit or mattress is experiencing a fault. The Alert Code display will continue to flash until the fault is cleared. The audible alert can be silenced for 20 minutes by pressing the Alert Mute button.
Page 21
Please ensure the battery has sufficient power while using disturbance or erratically unstable mains battery power/backup. power supply. Restart system by powering OFF, and then powering ON again. If the problem persists, move patient to another mattress and contact Novis Healthcare. User Guide ProCair Prime Advanced...
Technical Specifications Prime Advanced Single King Single System code APMPC-R03A APMPC-R03AK Capacity 30 - 200 kg System No of cells 21, including 3 static head cells and 4 narrow foot cells Compliance IEC60601-1, IEC60601-1-2 and IEC60601-1-11 ARTG 289458 Part number APMPC-CU07 Control system Digital micro controller...
Preventative maintenance needs to be conducted every 12 months. Please Clip the core onto the power cable of the contact Novis at 1300 738 885 or a certified pump, approximately 70 cm from the pump, maintenance technician. to prevent Radiated Emission (RE) from...
Technical specifications The Procair Prime Universal Therapy System M31 Control Unit is intended for use in the electromagnetic environment (for home and professional healthcare) specified below. The customer or the user of the Procair Prime Universal Therapy System M31 Control Unit should assure that it is used in such an environment.
Page 25
Electromagnetic environment –guidance (for Immunity test IEC 60601 test level Compliance level home and professional healthcare environment) 3 Vrms: 3 Vrms: Portable and mobile RF communications equipment should be used no closer to any part 0,15 MHz – 80 MHz 0,15 MHz –...
Technical specifications Manufacturer’s declaration – electromagnetic immunity Test specifications for Enclosure Port Immunity to RF wireless communications equipment Compliance level (V/m) (for home and Test Immunity frequency Maximum Distance test level professional (MHz) Band (MHz) Service Modulation power (W) (V/m) healthcare) Pulse modulation 380 –390...