If you have any questions, please contact Novis Healthcare on 1300 738 885. Class II Protection against Electric Shock Type BF Applied part Novis Healthcare has a policy of continuous product improvement and reserves the right to Alternating Current amend specifi cations presented in this guide.
Operation - Pull Cap CPR Release Operation - Defl ation / Storage Care and Cleaning Troubleshooting Technical Specifi cations EMI/EMC Statement and Manufacturer’s Declaration Warranty Statement SOFT GOODS CONTROL UNIT 3 YEAR WARRANTY 3 YEAR WARRANTY ProCair Trio User Guide...
System Overview The ProCair Trio is an alternating mattress The system consists of the following components: replacement system for the prevention and Mattress replacement with umbilical air treatment of skin breakdown and pressure injuries hoses and CPR release in patients of high to very high risk. It is designed ...
Institutions pressure care Intended Care Setting guidelines. Re-positioning of the patient is always Intended care settings for the ProCair Trio mattress recommended when using an Alternating replacement systems are Home healthcare and Pressure Air Mattress (APAM). Professional healthcare.
Page 6
The device should be positioned where it can be easily disconnected from main power The ProCair Trio control unit can be fi tted to the foot outlet. board of most hospital or aged care beds. Do not obstruct the mains plug or position the...
Only the control unit and mattress combination attaching it to an electrical outlet by the head as indicated by Novis Healthcare should be of the bed. used, otherwise the correct function of the product cannot be guaranteed.
Page 8
Keep the cord away from heated surfaces. Should the power cord be found Ensure there is no moisture in or near the to be worn or damaged, contact Novis Healthcare power inlet, control handset and power cord for a replacement.
Carefully unpack the system and inspect each item for any damage that may have occurred during transit and handling. Any damage or missing components should be reported to Novis Healthcare as soon as possible. Confi rm there are no sharp objects in the immediate area which may risk damage to the mattress replacement.
Page 10
System Preparation Check that the foot cell internal quick release air hose connectors are securely connected. Open the top cover by unzipping the CPR-side of the mattress (zipper located at foot end), check each foot cell connector is secure by pushing the air hose connectors together (there should be no movement).
Page 11
The system automatically defaults to Alternating Mode at start up, with AutoCair feature (automatic weight detection) always in operation to select an appropriate pressure for patient support according to mattress load, unless overridden manually. ProCair Trio User Guide...
You may need to unzip the cover to feel the they do not interfere with cell alternation. cells for infl ation. Place the patient on to the ProCair Trio With the patient lying supine, unzip one mattress. side of the top cover just past sacral region The system will automatically set an (lower spine).
(page 21) for detail, alarms pressure pressure fi rmer pressure include: range pressure range LOW PRESSURE - possible loss of pressure ProCair Trio User Guide...
Page 14
When periodical servicing is required, the amber temporarily. Alarm will reactivate in 20 minutes if Standby light will fl ash. Please contact Novis the issues has not been resolved, or immediately if Healthcare for system maintenance.
Page 15
BED EXIT ALARM light will fl ash. The alarm may be silenced by pressing the mute button, when sensor pad detects patient presence or when the alarm is turned off. ProCair Trio User Guide...
The battery unit is a consumable part and has a a reclined position (or below a 5° angle), Fowler usable life of up to 3 years and should be replaced Boost mode will automatically disengage. by an authorised Novis repair agent.
Max Infl ate button and wait 15 To deactivate the alarm, press BED EXIT button minutes to ensure all cells are fully again, the ON light will extinguish to indicate the infl ated. alarm is off. ProCair Trio User Guide...
LOCATED AT MATTRESS AUTO HEAD END - ALTERNATE FOWLER TWIST TO OPEN STATIC Switch off mains power and unplug the power POWER ProCair Trio cord from the mains outlet. STANDBY FLASHING NORMAL FLASHING: CHARGING BATT. SERVICE DUE IN USE <30% FLASHING: LOW BATT Open CPR release to release air and defl...
Cleaning and Infection Control Control Unit Cleaning Disconnect control unit from mains power before It is recommended that the ProCair Trio cleaning. Gently wipe down the external case with system is cleaned every two weeks if in a soft cloth.
Page 20
To ensure optimum performance and minimise potential mattress is running for approx. 12 months (8760 safety risk, Novis or its authorised service agent should hrs).reactivate if the fault is not rectifi ed or if a perform Preventative Maintenance Service check new fault is detected.
ALTERNATION FAILURE If issue persists, contact Novis Healthcare. A service may be Air cells have failed to alternate required. Ensure air hose connector is securely fastened and reconnect if SENSOR DISCONNECTED loose.
Page 22
For faster mattress reinfl ation, press MAX INFLATE and wait until the LP alarm code stops fl ashing. Press the ALT button to resume alternation. If the problem persists, move patient to an alternate product and contact Novis Healthcare.
Technical Specifi cations MODEL ProCair Trio ProCair Trio King Single SYSTEM CODE APMPC-R05 APMPC-R05K CAPACITY 40 - 250 kg SYSTEM 20, including 3 static head cells and 7 disconnectable heel cells. NO OF CELLS All alternating cells include a static lower chamber (cell-in-cell)
Page 24
(landfi lls) that are not appropriate according the misuse the ProCair Trio system or any of its legislation. Please be environmentally responsible components. and contact your local authority on available options to recycle this product at its end of life.
The ProCair Trio control unit is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the ProCair Trio control unit should ensure that it is used in such an environment. ProCair Trio User Guide...
Page 26
The ProCair Trio control unit is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the ProCair Trio control unit should ensure that it is used in such an environment. Immunity test...
Page 27
fi xed RF transmitters, an electromagnetic site survey should be considered. If the measured fi eld strength in the location in which the ProCair Trio control unit is used exceeds the applicable RF compliance level above, the ProCair Trio control unit should be observed to verify normal operation.
Page 28
EMI/EMC Statement and Manufacturer’s Declaration MANUFACTURER’S DECLARATION-ELECTROMAGNETIC IMMUNITY The M19-8 is intended for use in the electromagnetic environment (for home and professional healthcare) specifi ed below. The customer or the user of the M19-8 should assure that is used in such and environment. Electromagnetic environment- guidance Immunity test...
Page 29
D Caution: If abnormal behavior is observed due to EM disturbances, please relocate the device accordingly. Caution: Please do not use any other cables or accessories not approved by the manufacturer in this manual to avoid negative infl uence on electromagnetic compatibility ProCair Trio User Guide...
To obtain Novis will undertake at its option, to repair or replace, free of compensation, you will need to provided documentary charge, each product or part thereof on the condition that: evidence of the loss or damage suffered and documentary ...