Advertisement
Use
Always pay attention to information that is marked by the following warnings with warning pictograms. Implement the specified precautionary measures.
Representation | Explanation |
| indicates potential minor injuries |
| indicates circumstances that may lead to damage to the products |
Symbols and writing standards
| Note: Basic or further information. |
1. ... 2. ... | Action: more steps, the sequence must be followed. |
>... | Action: a step or an optional step. |
-... | Result of an action. |
[OK] | Control keys of the instrument or buttons of the program interface. |
> Only operate the product properly, for its intended purpose and within the parameters specified in the technical data. Do not use any force.
> Do not operate the instrument if there are signs of damage at the housing, mains unit or feed lines.
> Do not perform contact measurements on non-insulated, live parts.
> Do not store the product together with solvents. Do not use any desiccants.
> Carry out only the maintenance and repair work on this instrument that is described in the documentation. Follow the prescribed steps exactly. Use only original spare parts from Testo.
> Any further or additional work must only be carried out by authorised personnel. Testo will otherwise refuse to accept responsibility for the proper functioning of the measuring instrument after repair and for the validity of certifications.
> Only use the device in closed, dry rooms and protect it from rain and moisture.
> Temperatures given on probes/sensors relate only to the measuring range of the sensors. Do not expose handles and feed lines to any temperatures in excess of 70°C unless they are expressly permitted for higher temperatures.
> Dangers may also arise from the systems being measured or the measuring environment: Note the safety regulations valid in your area when performing the measurements.
Use distilled water, or alternatively mild solvents to clean the flue gas analyzer.
Do not store any objects that have come into contact with solvents and/or degreasers in the case. Evaporating or leaking solvents and/or degreasers may cause damage to the instrument and to the sensors.
The use of strong or harsh alcohol or brake cleaner can result in damage to the instrument.
Acid in the sensors.
May cause chemical burns.
Powder in the sensor filters.
May cause irritation to the skin, eyes or respiratory tract.
> Improper use of rechargeable batteries can lead to destruction or injuries by means of current surges, fire or escaping chemicals. The following instructions must be observed to avoid such hazards:
> Dispose of faulty rechargeable batteries/spent batteries in accordance with the valid legal specifications.
> At the end of its useful life, send the product to the separate collection for electric and electronic devices (observe local regulations) or return the product to Testo for disposal.
The testo 310 is a handheld measuring instrument for the professional flue gas analysis of combustion plants:
The testo 310 must not be used:
Measurement parameter | Measurement range | Resolution | Accuracy | Response time t90 @ 22°C |
O2 | 0.0 to 21.0 Vol.% | 0.1 vol.% | ±0.2 vol.% | 30s |
CO | 0 to 4000 ppm | 1 ppm | ±20 ppm (0...400 ppm) ±5% v. Mw. (401...2000 ppm) ±10% v. Mw. (2001...4000 ppm) | 60s |
COamb | 0 to 4000 ppm | 1 ppm | ±20 ppm (0...400 ppm) ±5% v. Mw. (401...2000 ppm) ±10% v. Mw. (2001...4000 ppm) | 60s |
Draught | -20.00 to 20.00 hPa | 0.01 hPa | ± 0.03hPa (-3.00 to 3.00 hPa) ±1.5% of meas. val. (rest of range) | - |
ΔP | -40.0 to 40.0 hPa | 0.1 hPa | ± 0.5hPa | - |
Flue gas temperature | 0.0 to 400.0°C | 0.1°C | ± 1°C (0.0 to 100.0°C) ± 1.5% of meas. val. (>100°C) | <50s |
Ambient temperature | -20.0 to 100.0°C | 0.1°C | ± 1°C | <50s |
Flue gas analyser
Feature | Values |
Storage and transport temperature | -20 to 50°C |
Operating temperature | -5 to 45°C |
Power supply | Rechargeable battery: 1500mAh Mains unit: 5 V/1 A |
Protection class | IP40 |
Weight incl. probe | approx. 700g |
Dimensions | 201 x 83 x 44 mm |
Battery charge time | approx. 8 h |
Rechargeable battery life | > 8h (pump on, 20°C ambient temperature) |
EU Directive | 2014/30/EC |
Button | Functions |
[![]() | Switch measuring instrument on / off |
[OK] Example | Function key (orange, 3x), relevant function is shown on the display |
[▲] | Increase value, select parameter |
[▼] | Decrease value, select parameter |
[esc] | Back, cancel print process |
[![]() | Switch display light on/off |
[![]() | Transmit data to the Testo protocol printer. |
Icon | Measurement |
![]() | Flue gas (icon visible when the instrument is switched off) |
![]() | Ambient CO (icon visible when the instrument is switched off) |
![]() | Draught (icon visible when the instrument is switched off) |
![]() | Differential pressure (icon visible when the instrument is switched off) |
Icon | Meaning |
![]() | Measuring gas pump (icon visible when the instrument is switched off) The inner segments light up alternately when the measuring gas pump is running. |
| Error Flashes when an error occurs, an error code is also displayed. |
![]() | Print Lights up during data transmission to the report printer |
![]() | Open configuration menu |
![]() | Fuel type / fuel number Depending on the set fuel, one of the icons (solid fuel, oil, gas) and the associated fuel number lights up. |
![]() | Battery capacity. Indication of the remaining capacity of the rechargeable battery by fill level of the battery icon:
|
Icon | Possible assignment | |
![]() | Left function key: Select measurement type | |
OK Start Stop | Middle function key: Confirm input Start measurement Stop measurement | |
Set → | Right function key: Open configuration menu Switch to the next parameter: |
Magnetic field
May be harmful to those with pacemakers.
> Keep a minimum distance of 10 cm between pacemaker and instrument.
ATTENTION
Magnetic field
Damage to other devices!
> Keep a safe distance away from products which could be damaged by the effects of magnetism (e.g. monitors, computers or credit cards).
The instrument can be configured for 5 area versions. Countryspecific calculation formulas, associated measurement parameters and fuels are activated with this setting. Similarly, the setting influences the date and time format. The area version is set in the instrument configuration menu / commissioning, see Instrument configuration menu / commissioning.
Area version (ArEA ) | Countries (recom mendation) | Parameters | Fuels |
1 | USA, HU, IN, KR | Line 1: Line 2: | nAt GAS – Natural gas ProP GAS – Propane FUEL OIL2 – Fuel oil 2 Bioh 5 – Biomass 5% Uood 20 – Wood 20% |
2 | GB, RU, DK, AU, JP, CN | Line 1: Line 2: | nAt GAS – Natural gas LPG GAS – LPG ProP GAS – Propane But GAS – Butane LI OIL – EL fuel oil CEro SEnE – Kerosene HE OIL – Heavy fuel oil Uood PELL - Wood pellets |
3 | NL, SE, TR, RO | Line1: Line 2: | nAt Hb – Natural gas Hb |
4 | DE, AT, CH, CZ, FR, ES, BE, PL, PT, AR, BR | Line1: T, CO, uCO, Line 2: | nAt GAS - Natural gas ProP GAS - Propane but GAS - Butane CoO GAS - Coke oven gas Toun GAS - Town gas GAS OIL – Gasoleo A LI OIL - EL fuel oil HE OIL - Heavy fuel oil Uood 15 - Wood 15% |
5 | IT | Line1:: Line 2: | nAt GAS – Natural gas |
Display | Measurement parameter |
![]() | Ambient temperature |
T | Flue gas temperature |
CO | Carbon monoxide |
O2 | Oxygen |
CO![]() | Ambient carbon monoxide |
qA | Flue gas loss without due consideration of the calorific value range |
Eff.net | Net efficiency (without due consideration of the calorific value range) |
Eff.gross / Eff* | Gross efficiency (with due consideration of the calorific value range) |
Eff. | Efficiency |
λ | Air ratio |
P2 | Differential pressure |
CO2 | Carbon dioxide |
P1 | Flue draught |
uCO | Carbon monoxide undiluted |
rat | Ratio |
EXA | Air surplus |
UnI | only when UnI ON Line 1: Line 2: |
ET | Heat of condensation |
The rechargeable battery is permanently installed and can only be changed by a Testo service centre.
The measuring instrument is supplied with a partially charged rechargeable battery.
> Charge the rechargeable battery fully before using the measuring instrument.
The rechargeable battery can only be charged at an ambient temperature of ±0 to +35°C. If the battery has been discharged completely, the charging time at room temperature is approx. 5-6 hrs.
Charging in the measuring instrument
- The charging process will start. The charge status will be shown on the display. The charging process will stop automatically when the battery is fully charged.
Battery care
> Do not fully exhaust the rechargeable battery.
> Only store instrument with battery charged and at low temperatures, but not below 0°C (best storage conditions with a charge level of 50-75% = 2 segments, at an ambient temperature of 10-20°C, recharge completely before use).
> The rechargeable battery life depends on the storage, operating and ambient conditions. The available useful life of the rechargeable battery reduces more and more with frequent use. If the useful life is significantly shortened, the rechargeable battery should be replaced.
- The measuring instrument is powered by the mains unit.
- If the measuring instrument is switched off, the charging process starts automatically. Switching the measuring instrument on has the effect of stopping the battery charging and the measuring instrument is then powered via the mains unit.
The instrument has two different configuration menus. The menu that is opened depends on the instrument status when called up.
When the instrument is switched on initially, the configuration menu is opened automatically.
Following initial commissioning, the configuration menu can also be opened again by pressing the right function key [set] during the instrument's initialisation phase (a duration of approx. 4 seconds). The values for area version, measuring units, time and date can be set.
Making settings
Use [esc] to revert to the previous parameter at any time.
Display / parameter | Explanation |
ArEA (area version) | Selecting the area version activates different calculation formulas and associated measurement parameters, see Area versions. > Select area version code: [▲] and [▼]. > Switch to the next parameter: [OK]. |
UnI | Only when area version 5 is selected Measurement procedure according to UnI norm: activate [On] / deactivate[OFF]. |
Unit of pressure | >Select the unit: [▲] and [▼]. >Switch to the next parameter: [OK]. |
Unit of temperature | >Select the unit: [▲] and [▼]. >Switch to the next parameter: [OK]. |
Setting the time | >Set values: [▲] and [▼]. >Switch selection between hour, minute (tens) and minute (single units): [→]. >Switch to the next parameter: [OK]. |
Setting the date | >Set values: [▲] and [▼]. >Switch selection between year (tens), year (single units), month, day (tens) and day (single units): [→]. >Exit the configuration menu: [OK]. |
This configuration menu enables you to make important settings relating to a measurement. Fuels and measuring units can be set.
Press right function key ([set]) after the instrument's initialisation phase.
Use [esc] to revert to the previous parameter at any time.
Display / parameter | Explanation |
Fuel | Selecting the area version activates different calculation formulas and measurement parameters, see Area versions. > Select the fuel: [▲] and [▼]. > Switch to the next parameter: [→]. > Exit the configuration menu: [OK]. |
UnI | Only when area version 5 is selected Measurement procedure according to UnI norm: activate [On] / deactivate[OFF]. |
Unit of pressure | > Select the unit: [▲] and [▼]. > Switch to the next parameter: [OK]. |
Unit of temperature | > Select the unit: [▲] and [▼]. > Exit the configuration menu: [OK]. |
Gas sensors
If flue gas measurement () or ambient CO measurement (
) is configured, the gas sensors are zeroed when the instrument is switched on (zeroing phase).
The flue gas probe must be in the open air during the zeroing phase!
Pressure sensor
If draught measurement ()or differential pressure measurement (
) is configured, the pressure sensor is zeroed when the instrument is switched on (zeroing phase).
Measuring the combustion air temperature
During the zeroing phase, the temperature is measured via the thermocouple of the flue gas probe. This temperature is permanently accepted by the instrument once the zeroing phase has been completed. All dependent parameters are calculated using this value. However, ensure that the flue gas probe is near the intake duct of the burner during the zeroing phase.
Checking the thermocouple
The thermocouple of the flue gas probe must not lie against the probe cage.
> Check before use. Bend the thermocouple back if necessary.
Aligning the flue gas probe
The flue gas must be able to flow freely past the thermocouple.
> Align the probe by turning it as required.
The tip of the probe must be in the core current of the flue gas flow.
> Align the flue gas probe in the flue gas duct so that the tip is in the core current (area of the highest flue gas temperature).
To carry out a flue gas measurement, the fuel must be set correctly, see Measurements configuration menu.
Select measurement type
> Select → [OK].
Carry out the measurement (area version 1 – 4, area version 5 with setting UnI OFF)
Carry out measurement (area version 5 with setting UnI ON)
To calculate a mean value, a series of measurements is carried out with 3 measuring phases (UnI 1 – UnI 3), each lasting 2 min. and 2 sec.
Cigarette smoke influences the measurement by more than 50 ppm. The breath of a smoker influences the measurement by about 5 ppm. The probe must be in the open air (CO-free) during the zeroing phase!
Select measurement type
> Select → [OK].
Carry out the measurement
Do not measure for longer than 5 min, as a drift of the pressure sensor means that the readings could be outside the tolerance limits.
Select measurement type
> Select→ [OK].
Performing the measurement
- The flue gas probe must be outside the flue.
|
Dangerous mixture of gases > Before measurement close the gas path with the sealing plug, as described below! > Make sure there are no leaks between the sampling point and the measuring instrument. > Do not smoke or use naked flames during measurement. |
Do not measure for longer than 5min, as a drift of the pressure sensor means that the readings could be outside the tolerance limits.
Select measurement type
> Select → [OK].
- PLUG appears.
> Close the gas path with the sealing plug. Please see the description below.
Prepare for measurement
Hot probe shaft! Risk of burns! > Allow the probe shaft to cool down after a measurement, before touching it! > Only attach the silicone hose to the probe shaft once it has cooled down! |
Carry out the measurement
- The silicone hose must be clear (pressureless, no kinks).
After the measurement
> If the housing of the measuring instrument is dirty, clean it with a damp cloth.
Use distilled water, or alternatively mild solvents to clean the flue gas analyzer.
Leaking solvents and degreasers!
Damage to the instrument and sensors!
The following substances can damage the instrument or the sensors:
Do not store cleaning cloths, solvents and degreasers, such as isopropanol, in the case.
Strong or harsh alcohol or brake cleaner! Damage to the instrument!
-Do not use any strong or harsh alcohol or brake cleaner.
> In case of contamination, clean the probe shaft and the handle of the flue gas probe with a damp cloth. Do not use any aggressive cleaning agents or solvents! Mild household cleaning agents and soap suds may be used.
Any cleaning of contamination within the probe shaft may only be carried out by Testo Customer Service.
The fill level of the condensate trap can be monitored via the markings on the condensate trap.
|
Skin irritation due to condensate! > Avoid skin contact. > Make sure that the condensate does not run over the housing. |
|
Damage to the sensors and the flue gas pump due to condensate entering the gas path! > Do not empty the condensate container while the flue gas pump is in operation. |
The condensate outlet must be completely closed, other wise measuring errors could occur if external air gets in.
Checking the particle filter:
> Check the particle filter of the flue gas probe for contamination at regular intervals: check visually by looking through the window of the filter chamber.
Replace the filter if there are signs of contamination.
Replacing the particle filter:
The filter chamber may contain condensate.
Question | Possible causes / solution |
Rechargeable battery low | > Switch to mains operation. |
Measuring instrument switches off automatically or cannot be switched on | Batteries / rechargeable batteries empty. > Charge rechargeable battery or switch to mains operation. |
When switching off, the instrument rinses the gas path for a long time and does not shut down. | The gas path is closed by the sealing plug. > Remove the sealing plug and insert the filter. |
Error message: E01 | Instrument temperature outside the valid range > Bring the instrument into the valid operating temperature range. |
Error message: E04 | O2 sensor is worn out (30% of the calibration signal) > Replace sensor and calibrate. |
Error message: E05 | O2 sensor faulty (O2 sensor 25% above the calibration signal) > Replace sensor and calibrate. |
Error message: E06 | O2 sensor close to wear limit or sensor not in fresh air during the zeroing phase > Carry out zeroing phase again or suggest sensor replacement. |
Error message: E07 | CO sensor worn out > Replace sensor and calibrate. |
Error message: E08 | CO measuring value outside the measuring range (>4000 ppm) > Purge CO sensor with fresh air |
Error message: E12 | CO baseline increased (> 50 ppm above original zero point) > Carry out zeroing phase again or replace sensor. |
Error message: E13 | CO baseline unstable (drift > 20 ppm) > Carry out zeroing phase again or replace sensor. |
Error message: E14 | Checksum error > Carry out calibration |
Is it possible to print out company data? | Company data should be displayed in the header of the print-out. > Input/reading in of the company data by Testo Service. |
For more information, please contact your local dealer or the Testo Customer Service. For contact details, see the back of this document or the website: www.testo.com/service-contact
Printer
Description | Item no. |
Protocol printer | 0554 3100 |
Spare thermal paper for printer (6 rolls) | 0554 0568 |
Accessory for flue gas probe
Description | Item no. |
Particle filter, 10 pcs. | 0554 0040 |
Other accessories
Description | Item no. |
Mains unit 5V 1A with mini-USB connecting cable | 0554 1105 |
Instrument cleaner (100 ml) | 0554 1207 |
For other accessories and spare parts, please refer to the product catalogues and brochures or look up on the internet at www.testo.com
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Need help?
Do you have a question about the 310 and is the answer not in the manual?
Questions and answers