Endress+Hauser Solitrend MMP60 Operating Instructions Manual
Endress+Hauser Solitrend MMP60 Operating Instructions Manual

Endress+Hauser Solitrend MMP60 Operating Instructions Manual

Material moisture measurement
Hide thumbs Also See for Solitrend MMP60:

Advertisement

Quick Links

BA01946M/00/EN/04.25-00
71698260
2025-02-28
Products
Operating Instructions
Solitrend MMP60
Material moisture measurement
Solutions
Services

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Solitrend MMP60 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Endress+Hauser Solitrend MMP60

  • Page 1 Products Solutions Services BA01946M/00/EN/04.25-00 71698260 2025-02-28 Operating Instructions Solitrend MMP60 Material moisture measurement...
  • Page 2 Solitrend MMP60 Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3: Table Of Contents

    Solitrend MMP60 Table of contents Table of contents About this document ....4 Repair ......25 Purpose of this document .
  • Page 4: About This Document

    About this document Solitrend MMP60 About this document Purpose of this document These Operating Instructions contain all the information that is required in the various phases of the life cycle of the device: from product identification, incoming acceptance and storage, to mounting, connection, operation and commissioning through to troubleshooting, maintenance and disposal.
  • Page 5: Documentation

    Solitrend MMP60 About this document Documentation The following document types are available in the Downloads area of the Endress+Hauser website (www.endress.com/downloads): For an overview of the scope of the associated Technical Documentation, refer to the following: • Device Viewer (www.endress.com/deviceviewer): Enter the serial number from the nameplate •...
  • Page 6: Basic Safety Instructions

    Basic safety instructions Solitrend MMP60 Basic safety instructions Requirements for personnel The personnel for installation, commissioning, diagnostics and maintenance must fulfill the following requirements: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task. ‣...
  • Page 7: Workplace Safety

    Solitrend MMP60 Basic safety instructions Workplace safety For work on and with the device: ‣ Wear the required personal protective equipment according to federal/national regulations. Operational safety Risk of injury! ‣ Operate the device only if it is in proper technical condition, free from errors and faults.
  • Page 8: Product Description

    Product description Solitrend MMP60 Product description Device for measuring low-density bulk solids with conductivity values up to 1 mS/cm. Product design A0040364  1 Product design Wave guide Measuring cell; TECAPEEK Housing 3.1.1 ATEX version A0053310  2 Rectangular sensor, ATEX version...
  • Page 9: Incoming Acceptance And Product

    ‣ Enter the serial number from the nameplate into the Endress+Hauser Operations App or scan the 2-D matrix code on the nameplate.  All of the information on the measuring device and on the scope of the technical documentation pertaining to the device is displayed.
  • Page 10: Mounting

    Mounting Solitrend MMP60 Mounting Mounting requirements • The device must be installed at a point in the process in such a way that ensures a constant bulk density, as the bulk density directly affects the calculation of the water content. Where necessary, a bypass should be created or structural measures may be needed at the place of installation to ensure that the material flow, and therefore the bulk density, over the measurement surface is constant.
  • Page 11: Post-Mounting Check

    Solitrend MMP60 Mounting 5.2.1 ATEX electronics housing A0053050  4 Dimension of ATEX electronics housing. Unit of measurement mm (in) 5.2.2 Mounting the housing with the remote electronics module The housing with the remote electronics module can be mounted with two screws (M5).
  • Page 12: Electrical Connection

    Electrical connection Solitrend MMP60 Electrical connection Connecting requirements 6.1.1 Cable specification Connecting cables with a pre-assembled 10-pin socket on the device side are available in different standard lengths: • 4 m (13 ft) • 10 m (32 ft) • 25 m (82 ft) Shielded cable UNITRONIC PUR CP, twisted pairs 6 ×...
  • Page 13 Solitrend MMP60 Electrical connection 6.2.2 ATEX rectangular sensors ATEX version Rectangular sensor • Cable length between sensor and ATEX electronics housing 5 m (16 ft) • Cable firmly connected to the sensor with plug to the ATEX electronics housing ATEX electronics housing •...
  • Page 14 Electrical connection Solitrend MMP60 6.2.3 Connection to a remote display (optional) 1 234 A0040962  8 Connection to a remote display USB (Mini B type), USB-IMP-Bridge, firmware update (only for service purposes) Socket for supply voltage and bus interface Connector for supply voltage and bus interface (included in the delivery for "remote display")
  • Page 15: Post-Connection Check

    Solitrend MMP60 Electrical connection 6.2.4 Connection example of 10-pin socket 6 7 8 A0037418  9 Connection example, cable with 10-pin socket (on device side) and wire end ferrules at the cable end Transmitter PLC/distribution box Remote display (optional) power supply...
  • Page 16: Operation Options

    Operation options Solitrend MMP60 Operation options RS485 IMP-Bus A0046920  10 Overview Device Remote display Computer LED display PLC or water dosing computer Endress+Hauser...
  • Page 17: Commissioning

    Solitrend MMP60 Commissioning Commissioning Analog outputs for measured value output The measured values are output as a current signal via the analog output. The device can be set to 0 to 20 mA or 4 to 20 mA. The current output can also be set inversely to 20 to 0 mA or 20 to 4 mA for special controllers and applications.
  • Page 18: Operating Mode

    Commissioning Solitrend MMP60 Moisture range The moisture range and the temperature range at outputs 1 and 2 can be individually configured. • Moisture range in % • Maximum: e.g. 20 % • Minimum: 0 % • Temperature range in °C •...
  • Page 19: Calibration Curve Set A For General Bulk Solids Applications

    Solitrend MMP60 Commissioning • CK mode (Cyclic with Boost Filter) For complex applications in mixers and dryers • CC mode (Cyclic Cumulated) With automatic totalization of moisture quantity measurements in one batch process if no PLC controller is used • CH mode (Cyclic Hold) Standard operating mode for applications in the construction industry.
  • Page 20: Settings

    Commissioning Solitrend MMP60 H [%] t [ps] A0037432  12 Calibration curve set A (Cal.8, Cal.9, Cal.12, Cal.13, Cal.14, Cal.15) Gravimetric moisture; % Radar transit time; picoseconds Cal.8, Lignite Cal.9, Basic calibration Cal.12, Sewage sludge Cal.13, Cereals (linear) Cal.14, Air/water 0 to 100 % Cal.15, Raw data calibration (...
  • Page 21 Solitrend MMP60 Commissioning material. This method delivers a characteristic value depending on the mineral concentration. Depending on the moisture content, the conductivity measuring range of the devices here is up to 1 mS/cm. 8.5.2 Material temperature measurement The device contains an integrated temperature sensor, which measures the housing temperature.
  • Page 22: Diagnostics And Troubleshooting

    Diagnostics and troubleshooting Solitrend MMP60 Diagnostics and troubleshooting The device is normally precalibrated with calibration set B and Cal.14 (air/water 0 to 100 %) upon delivery. Fine-tuning to achieve an accuracy of ±0.1 % in relation to the laboratory value can be performed via a PLC or the remote display (optional).
  • Page 23 Solitrend MMP60 Diagnostics and troubleshooting Incorrect data processing In the event of inaccurate data processing, check the moisture value displayed in the PLC. 1. Connect the device to the remote display 2. Compare the moisture value shown in the PLC with the moisture value shown on the display 3.
  • Page 24: Maintenance

    Maintenance Solitrend MMP60 Maintenance No special maintenance work is required. 10.1 Exterior cleaning When cleaning the exterior, make sure that the cleaning agent used does not corrode the surface of the measuring cells and housing. Endress+Hauser...
  • Page 25: Repair

    Repair 11.1 General information 11.1.1 Repair concept Under the Endress+Hauser repair concept for the device, the device can be repaired by the Endress+Hauser Service Department. For more information, please contact the Service Department at Endress+Hauser. 11.2 Return The requirements for safe device return vary depending on the device type and national legislation.
  • Page 26: Technical Data

    Technical data Solitrend MMP60 Technical data 12.1 Input Measured variable • Channel 1 Material moisture in % (variable setting) • Channel 2 Conductivity 0 to 1 mS/cm or temperature 0 to 100 °C (32 to 212 °F), this also applies for the high-temperature version.
  • Page 27: Performance Characteristics

    Solitrend MMP60 Technical data 12.3 Performance characteristics Reference operating The following reference conditions apply to the performance characteristics: conditions • Ambient temperature: 24 °C (75 °F) ±5 °C (±9 °F) • Ideal installation conditions: • Constant bulk density • Sufficient material volume flow through the measuring field •...
  • Page 28: Process

    Technical data Solitrend MMP60 Operating height Up to 2 000 m (6 600 ft) above sea level Degree of protection IP65 12.5 Process Process temperature range • Standard, 0 to 70 °C (32 to 158 °F) • High-temperature version (electronics module located remotely in separate housing), 0 to 100 °C (32 to 212 °F)
  • Page 32 *71698260* 71698260 www.addresses.endress.com...

Table of Contents