Instrument Options—608
OPTION 21
Option 21 Performance Check Summary
Characteristic
Performance Requirement
Test Method
Procedure Title
HORIZONTAL (X) AMPLIFIER
Polarity (-X INPUT)
Positive signal applied de-
flects beam to the left; nega-
tive signal deflects beam to
Satisfactory operation is
substantiated by other tests.
Common-Mode Rejection
DC to 100 kHz
At least 100:1 cmr ratio for
(1X attenuation.)
A 100-kHz, 10 V (p-p) sine
wave is applied to both the
+X and -X INPUTS. Length
Check Horizontal Common-
Mode Rejection.
100 kHz to
1 MHz
At least 50:1 cmr ratio for
(1X attenuation.)
A 1-MHz, 10 V (p-p) sine
wave is applied to both the
+X and -X INPUTS. Length of
horizontal trace is checked.
) AMPLIFIER
Polarity (-Y INPUT)
flects beam down; negative
substantiated by other tests.
Common-Mode Rejection
DC to 100 kHz
(1X attenuation.)
A 100-kHz, 10 V (p-p) sine
wave is applied to both the
+¥ and -Y INPUTS. Length of
vertical trace is checked.
Check Vertical Common-
Mode Rejection.
100 kHz to
1 MHz
At least 50:1 cmr ratio for
(1X attenuation.)
A 1-MHz, 10 V (p-p) sine
wave is applied to both the
+Y and -Y INPUTS. Length of
Common-Mode Rejection
DC to 100 kHz
A +10 ws ramp is applied to
10-V (p-p) sine wave is ap-
plied to both the +Z and -Z
INPUTS. A test oscilloscope
Check Z-Axis Amplifier
Common-Mode Rejection.
100 kHz to 1 MHz
A +10 ys ramp is applied to
the +X INPUT. A 1-MHz, 10 V
(p-p) sine wave is applied to
both the +Z and -Z INPUTS. A
test oscilloscope is connect-
ed to TP599 and the ampli-
tude of the Z-Axis output
signal is checked.