Instructions for use
17-10-07-643-002h_2024-11S

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 878 dr and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Seca 878 dr

  • Page 1 878 dr Instructions for use 17-10-07-643-002h_2024-11S...
  • Page 2: Table Of Contents

    Disinfecting........... 15 Sterilizing ............15 7 Function check ............. 16 8 Troubleshooting............ 16 9 Servicing ..............17 10 Technical data............17 10.1 General technical data........17 10.2 Weight measurement ........18 11 Optional accessories..........18 2 • seca 878 dr...
  • Page 3: Contraindications ......................................... 3 13 Warranty

    English DESCRIPTION OF DEVICE Intended use The electronic flat scale supports physicians in decision-making regarding diagnosis or therapy, based on the patient's weight or weight changes. User qualification The device may only be used by persons with sufficient expertise. Patient target group The device is intended for persons of any age who can stand independently and who do not exceed the maximum weight capacity of the device.
  • Page 4: Safety Information

    ► Technical modifications may not be made to the device. The device does not contain any user-serviceable parts. Only have servicing and repairs performed by an authorized seca service partner. You can find your local service partner at www.seca.com. ► Use only seca original accessories and spare parts, otherwise seca will not grant any warranty.
  • Page 5: Preventing Electric Shock

    ► Keep high-frequency equipment such as cellphones at a minimum distance of approx. 1 meter to prevent faulty measurements. ► The actual transmission output of HF devices may require minimum distances of more than 1 meter. For details, go to www.seca.com. Preventing electric shock WARNING! Electric shock ►...
  • Page 6: Preventing Damage To Device

    (e.g. in seca software or in an EMR system), ensure that the measured values are plausible. ► If measured values have been sent to seca software or to an EMR system, ensure before continuing to use them that the measured values are plausible and assigned to the correct patient.
  • Page 7: Handling Batteries And Rechargeable Batteries

    This prevents acid from leaking into the device. ► If harmful substances have penetrated the device, do not continue using it. Have the device checked by an authorized seca service partner and repaired if necessary. Safety information • 7...
  • Page 8: Overview

    Takes 6 Mignon batteries (Type AA/1.5 V) Spirit level Indicates whether the device is horizontal Symbols in the display Item Symbol Meaning function active Tare HOLD Hold function active Non-verifiable function active Weight in kilograms Batteries are low 8 • seca 878 dr...
  • Page 9: Labels

    (product identification number as per Directive (EU) 2017/745) Article number Serial number Value in m/s² (model-dependent) • Gravitational acceleration on earth • Depends on the intended location seca product identification number ProdID Approval Type Type designation of design approval Mat.No. Variant number...
  • Page 10 • : Operate device with direct current seca adapter only • : Use compatible seca power supply units only Do not dispose of device in household waste Markings on the packaging Symbol Meaning Protect from moisture...
  • Page 11: Starting Up Device

    English STARTING UP DEVICE Scope of delivery Item Component Pcs. Personal scale Interchangeable design panel Battery − "Changing design panel" flyer − Instructions for use Starting up device • 11...
  • Page 12: Establishing Power Supply

    Conventional power supply units may deliver a higher voltage than is indicated on them. The measuring device may overheat, catch fire, melt or short-circuit. ► Use only original seca power supply units with a controlled 12 V output volt- age.
  • Page 13: Operation

    ► Set the ON/OFF switch to position “I”. ON/OFF switch ð appears in the display. SECA ð All the elements of the display are shown briefly. ð The scale is operational when appears in the display. Switching off the device ►...
  • Page 14: Weighing

    5. Ask the adult to get onto the scale holding the child. ð The child's weight is displayed. ð The symbol and the messages are displayed. HOLD 6. Read off the measuring result. 14 • seca 878 dr...
  • Page 15: Hygiene Treatment

    English 7. Ask the adult to get off the scale with the child. 8. You have the following options for deactivating the 2 in 1 function: ► Keep the hold/2 in 1 key depressed until the symbol and the messages are no longer displayed.
  • Page 16: Function Check

    ► Follow the section entitled “Servicing” in this document. TROUBLESHOOTING If faults occur when operating the device, first attempt to remedy them yourself using the following table. If the fault persists, contact seca Service or an authorized service partner. Fault...
  • Page 17: Servicing

    CAUTION! Faulty measurements as a result of poor servicing ► Have servicing and repairs carried out exclusively by seca Service or an au- thorized service partner. ► You can find your local service partner at www.seca.com.
  • Page 18: Weight Measurement

    • Plug-in power supply, Euro: 230 V~ / 50 Hz / 12 V= / 150 mA 68 32 10 252 • Switch-mode power supply unit: 100-240 V~ / 50-60 68 32 10 266 Hz / 12 V= / 0.5 A 18 • seca 878 dr...
  • Page 19: 12 Disposal

    The warranty will be voided if the device is opened by persons not expressly authorized by seca to do so. In the event of a warranty issue, please contact your local seca office or the dealer from whom you ordered the product.
  • Page 20 Medical Measuring Systems and Scales since 1840 seca gmbh & co. kg Hammer Steindamm 3  ̶   2 5 22089 Hamburg . Germany T +49 40 20 00 00 0 F +49 40 20 00 00 50 E info@seca.com seca operates worldwide with headquarters...