SonoSite, Inc., in various jurisdictions. FUJIFILM is a registered trademark of FUJIFILM Corporation in various jurisdictions. Value from Innovation is a trademark of FUJIFILM Holdings America Corporation. DICOM is a registered trademark of the National Electrical Manufacturers Association. All other trademarks are the property of their respective owners.
Table of Contents About the mounting solution ..................1 Inside the box ......................... 2 Mounting solution installation ..................3 Safety and ergonomics ....................8 General ........................... 8 Electrical safety ......................9 Ergonomic safety ......................9...
About the mounting solution The Sonosite PX, when equipped with the mounting solution, can be attached to a mounting system available at the installation site. CAUTION Read all warnings, cautions, and instructions. Failure to follow these instructions can result in serious injury or death.
Inside the box 4 mm Mounting assembly Cover Stand head #1 Phillips screwdriver AC power cable 5-inch plunger plate with three 10-32 x 5/8 inch screws Power supply 4 mm bit Inside the box...
5-inch plunger plate to the customer mounting system (also 18-8 stainless steel). 2. Slide the base of the FUJIFILM Sonosite mounting assembly into the 5-inch plunger plate attached to the mounting system. There are plastic set screws underneath the tray that may need to be loosened before the adapter can slide in.
Page 7
4. Insert the stand head pedestal into the mounting assembly stub. 5. Use the bit driver with the 4 mm bit to tighten the four screws on the stand head. 6. Use the #1 Phillips bit to Install the cover with two screws. Mounting solution installation...
Page 8
7. Connect the column power cable to the stand head by pushing it into place and screwing it into the connector. Tug gently to make sure it is secure. 8. Secure the power supply. Mounting fixtures may be available from the manufacturer of the mounting system.
Page 9
12. Make sure that the power status light on the rear of the stand head is on. 13. Tighten the connection between the system and the TTC by pushing the connector back into place, rotating the latch in the opposite direction, then pressing it up. 14.
Page 10
15. After connecting a transducer, use the cord hooks on the front corners of the stand head to manage the transducer cables. Mounting solution installation...
NOTE The Sonosite PX mounting solution, including the mounting assembly, the base tray and the screws used to attach the tray to the customer mounting plate, were tested to meet the Tensile Safety Factor requirements per IEC 60601-1:2012 (Edition 3.1), sub-clause 9.8.2, Table 21 (tensile safety factor...
Connection of accessories and peripherals not recommended by FUJIFILM Sonosite could result in electrical shock. Contact FUJIFILM Sonosite or your local representative for a list of accessories and peripherals available from or recommended by FUJIFILM Sonosite.
Page 13
23.34 in. 593 mm At full tilt 12.3 in. 312 mm Minimum clearance from wall for swivel application 12.3 in. 312 mm 27.5 - 37.5 in. 700 - 950 mm Recommended mounting 14.28 in. height from floor 363 mm At minimum tilt Safety and ergonomics...