This prescription device is for use under the direction and supervision of a clinician. THIS GUIDE IS A SUPPLEMENTAL DOCUMENT FOR CLINICIAN USE ONLY. CAUTION Use only DeRoyal brand canisters and accessories ® with this PROSPERA PRO-II device to enable continuous and ®...
Up and Down buttons. Filter run time elapsed; replacement of the internal filter by an authorized DeRoyal service technician is required. X = Represents Sensitivity of “System Closed” alarms Alarm display settings...
LED will illuminate when connected properly. 3. When connected properly, the plug icon will appear on the display screen. The PROSPERA device should be charged daily with the DeRoyal ® provided power supply. When in use, place the device upright on a secure surface, avoiding areas where device is prone to falling or being damaged.
Page 4
Before Starting Suction, Confirm 1. Pressure (mmHg) settings on home screen match prescribed pressure (mmHg) 2. Sensitivity settings are set to 5/L as shown 3. DeRoyal canister is securely ® connected to device 4. Tubing is applied properly and connected to canister 5.
Page 5
Canister Exchange Stop suction prior to initiating any canister changes. Confirm canisters and accessories are within arms reach prior to beginning any canister change procedures. Inspect all packaging for damage or defects. Replace the canister if the packaging is damaged. CONNECTING THE DISPOSABLE EXUDATE CANISTER OF THE PRO-II DEVICE ®...
Continuous Operation In the continuous operating mode, the PRO-II device maintains ® continuous pressure. The pressure value can be set from -20 mmHg to -200 mmHg in increments of 5 mmHg. 7. Press the button to start suction. The following screen is displayed.
PRO-II Canisters are for use with the Prospera PRO-II only. PRO-II canisters are prescription medical devices for use under the direction of a licensed and trained clinician. Read and understand all accompanying documentation including all indications, contraindications, warnings, cautions, precautions and instructions for use completely and carefully before use.