System Diagram
4-Probe Expansion Board
Multi-Probe
Measuring Probe (2 m) CA-P02
Measuring Probe (5 m) CA-P05
High luminance Measuring Probe (2 m) CA-PH02
High luminance Measuring Probe (5 m) CA-PH05
(Optional)
Measuring Probe Holder
CA-A11 (Standard)
Specifications
Receptor
Display range
Luminance
Chromaticity
Luminance
Measurement range
Accuracy (for white)*1
Repeatability *1
Chromaticity
Measurement range
Accuracy *1
(temperature : 23°C±2°C,
relative humidity : (40±10)%)
Repeatability *1
Measurement speed
xyLv
*3
Display
Digital
Analog
SYNC mode
Object under measurement
Memory channel
Analyzer function
Interface
Multi-point Measurement
Operating temperature/humidity range
Storage temperature/humidity range
Input voltage range
Size
Mainbody, Probe
Weight
Mainbody, Probe
*1 The chromaticity and luminance are measured under Konica Minolta's condition (standard CRT (6500 K) is used).
*2 The luminance for monochrome is measured when the reading of luminance for white is 40.00 cd/m
*3 Measuring probe connected to probe connector P1 only,used USB (used RS-232C Baud rate: 38400 bps)
*4 At the CA-200 mode
� Select the probe among the four types.
� Specifications are subject to change without notice.
SAFETY PRECAUTIONS
For correct use and for your safety, be sure to read
the instruction manual before using the instrument.
Always connect the instrument to the
specified power supply voltage.
Improper connection may cause a fire
or electric shock.
©2003 KONICA MINOLTA SENSING, INC.
PC (Commercially available)
PC-AT compatible
CA-B04
(Optional)
PC Software for Color Analyzer
CA-SDK (Standard)
CA-100Plus
Measuring Probe (2 m) CA-P02
Measuring Probe (5 m) CA-P05
High luminance Measuring Probe (2 m) CA-PH02
AC Power Cord (Standard)
High luminance Measuring Probe (5 m) CA-PH05
CA-100Plus(Measuring Probe 2 m or 5 m)
Detector : Silicon photo cell
0.01 to 1000 cd/m
2
Displayed in 4 or 3-digit value (Can be chosen)
0.05 to 1000 cd/m
2
±2 %±1 digit of reading (temperature : 23°C±2°C, relative humidity : (40±10)%)
0.2 %+1 digit (2 σ)
0.05 to 1000 cd/m
2
0.05 to 1000 cd/m
2
0.05 to 0.19 cd/m
±0.006 (for white)
2
±0.004 (for white)
0.20 to 0.49 cd/m
2
±0.003 (for white)
0.50 to 1000 cd/m
2
±0.002 (for white), ±0.004 (for monochrome)*2
40.00 cd/m
2
0.006 (2 σ)
0.05 to 0.19 cd/m
2
0.20 to 0.49 cd/m
0.002 (2 σ)
2
0.001 (2 σ)
0.50 to 1000 cd/m
2
0.05 to 0.99 cd/m
5 measurements/sec. (4.5 measurements / sec.)*4
2
1.00 to 1000 cd/m
20 measurements/sec. (17 measurements / sec.)*4
2
xyLv, T∆uvLv, RGB, XYZ, u'v'Lv
∆x, ∆y, ∆Lv, R/G, B/G, ∆G, ∆R, B/R, G/R
NTSC, PAL, EXT, UNIV, INT
Vertical syncronizing frequency : 40 to 200 Hz
100 channels
Standard function
USB (1.1 conformity) , RS-232C (38,400 bps or below)
Max. 5 points (Use 4-Probe Expansion Board CA-B04)
Temperature : 10 to 28°C; relative humidity 70 % or less with no condensation
Luminance change : ±2 % ±1 digit of reading for white
Chromaticity change : ±0.002 for white, ±0.006 for monochrome
from reading of Konica Minolta's standard CRT*1, 40.00 cd/m
0 to 28°C : relative humidity 70 % or less with no condensation
28 to 40°C : relative humidity 40 % or less with no condensation
100-240 V ~ , 50-60 Hz 50 VA
340 × 127 × 216 mm (W × H × D), φ45 × 142 mm
3.58 kg, 285 g
Dimensions
340
CA-100Plus(High luminance Measuring Probe 2 m or 5 m)
0.01 to 2000 cd/m
2
0.05 to 2000 cd/m
2
0.05 to 2000 cd/m
2
0.05 to 2000 cd/m
2
0.05 to 0.09 cd/m
2
0.10 to 0.39 cd/m
2
0.40 to 0.99 cd/m
2
1.00 to 2000 cd/m
2
40.00 cd/m
2
0.05 to 0.09 cd/m
2
0.10 to 0.39 cd/m
2
0.40 to 0.99 cd/m
2
1.00 to 2000 cd/m
2
0.05 to 1.99 cd/m
2
2.00 to 2000 cd/m
2
, with 23°C 40 %
2
.
2
Certificate No : YKA 0937154
Certificate No : JQA-E-80027
Registration Date : March 3, 1995
Registration Date : March 12, 1997
9242-4885-12
(Units : mm)
φ
60
φ
45
30.2
φ
44
0.2 %+1 digit (2 σ)
±0.008 (for white)
±0.006 (for white)
±0.004 (for white)
±0.003 (for white)
±0.002 (for white), ±0.004 (for monochrome)*2
0.009 (2 σ)
0.006 (2 σ)
0.002 (2 σ)
0.001 (2 σ)
5 measurements/sec. (4.5 measurements / sec.)*4
20 measurements/sec. (17 measurements / sec.)*4