CONTENT 1. Introduction ....................... 1-1 Getting Started ....................1-1 Major Components ..................1-1 Mechanisms of Motion..................1-3 Configurations ....................1-3 Glossary of Symbols ..................1-5 Safety Warnings ..................... 1-6 2. Description....................2-1 Control Unit....................2-1 Operator Pendant.................... 2-2 3. Installation and Setup................... 3-1 Unpacking the Unit ..................
1 Introduction 1. INTRODUCTION Getting Started Thank you for purchasing a Burleigh PCS-4100 series motorized micromanipulator. This manual is intended to provide in-depth information on the operation and use of the Burleigh PCS-4100. Burleigh PCS-4100 series patch-clamp micromanipulators (Figure 1-1) from EXFO Life Sciences &...
Page 6
Burleigh PCS-4100 Series Operating Manual 1 Introduction Figure 1-2: Burleigh PCS-4100 micromanipulator (head stage not included) Operator Pendant The operator pendant (Figure 1-3) is a remote control for positioning the motorized vertical axis. It provides the ability to control both the speed and direction of travel as well as to set and return to two memorized locations.
The other is mounted to the moving portion of the vertical axis and allows for an angle adjustment of the pipette holder. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Page 8
Burleigh PCS-4100 Series Operating Manual 1 Introduction Figure 1-6: Rotary pivot stage with adjustable stop mechanism Life Sciences & Industrial Division (11026-M-00 Rev. C)
Do not proceed unless you understand and meet the required conditions. Chassis Ground Motor Direction Switch Input/ Output Control Signals Fuses- User Replaceable Safety Ground EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Burleigh PCS-4100 Series Operating Manual 1 Introduction Safety Warnings WARNING Should this equipment be used in a manner not specified by EXFO Life Sciences & Industrial Division, the protection provided by the equipment may be impaired. Always follow the operating instructions as indicated! WARNING Refer servicing to qualified personnel.
The Move indicator will be lit only when motor is being commanded to move. CAUTION Never connect or disconnect the cables when the unit is on. This could cause an arc and damage to the control unit. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Operator Pendant Motor A and Motor B Operator Pendants A and B plug into their respective connections on the back of the unit. Operator Pendant Speed Selector Switch Rotate to select between LOW, MEDIUM and MAX ranges. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Page 13
The setting is lost upon power off. To move the vertical stage to the value stored in memory, press and quickly release (hold less than two seconds). EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Figure 3-1: Location of first fine-adjust screw 2. Bring the tip of the adjustment screw into contact with the thrust washer (Figure 3-2). bring into contact Figure 3-2: Location of second fine-adjust screw EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Page 16
BOTH motor direction switches (see Figure 2-1). Store the bracket and the four screws in the tool kit. Figure 3-6: Location of remaining screws EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Page 17
Figure 3-9: Location of locking pins. 10. Turn off the control unit. Unplug the AC line cord, the operator pendant and the motor to prepare for mounting on the microscope. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
IMPORTANT Be sure there is enough slack in the cables to prevent tension on the stages when the stages move or the manipulator rotates. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
HEKA EPC-9 headstage. Additional holes may be added for other headstages. A separate mounting plate is available for the mounting of HEKA EPC10. Contact EXFO Life Sciences & Industrial Division for more information 1.
1. Use the operator control to adjust the vertical axes to the center of travel. 2. Use the manual adjusting knob to center the manual stages. 3. Affix the micromanipulator to the mounting surface. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
With the headstage pivot stage facing you, the pin should be resting below the brass block. Tighten the visible set screw with the .050” Allen wrench. The other set screw is covered by the brass block. This is normal. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Loose mechanical components also can cause problems. For more information refer to the EXFO Application Note: Pipette Drift Sources and Solutions for Patch Clamp Recording Experiments. Warm up If it is required to maintain stability of 1 µm per hour, be sure that the...
3. Drive the vertical stage to the highest position possible to spread the lubricant along the screw and to lubricate the internal mechanism of the motor. Fig 5-1: Lubricating the motor lead screw EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
To replace the fuse(s): 1. Turn off the unit and disconnect from the power source. 2. Remove the fuse holder by prying up as shown in Figure 5-3. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Problem: Excess motion of pipette tip during manual adjustment Typical symptoms Excess or erratic motion of pipette tip during manual adjustment. Pipette sensitive to vibration. "Losing" cells after making a seal. Also refer to Drift later in this section. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Motor may be running but end of travel has been reached and motor is in Stall condition. End of travel may have been reached and motor is stuck. Incorrect connection of the motor cables. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Page 29
(Max) positions. Verify that the orange Move light becomes lighted when Up or Down is pressed. Listen closely to the motor to hear any noise. If the motor is running, a noise should be heard. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
7 " x 8.5" x 7.5" 12" x 10" x 8.5" Weight 5 lbs. 6 lbs. Power Requirements Voltage and Frequency 100 to 240 V 50/60 Hz, 1.0A Fuse T4A, 250V, 0.25" x 1.25" [5x20mm] EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference and (2) this device must accept any interference received, including interference that may cause undesired operation. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
Das oben gezeigte, durchkreuzte Mülltonnensymbol soll Sie einladen, diese Systeme zu benutzen. • Sollten Sie weitere Informationen über Systeme zur Sammlung, Wiederverwendung und Wiederverwertung haben, so setzen Sie sich bitte mit ihrer zuständigen Müllentsorgung in Verbindung. EXFO Life Sciences & Industrial Division (11026-M-00 Rev. C)
EXFO Life Sciences & Industrial Division personnel. The liability of EXFO Life Sciences & Industrial Division (except as to title) arising out of supplying of said product, or its use, whether under the foregoing warranty, a claim of negligence, or otherwise, shall not in any case exceed the cost of correcting defects in the products as herein provided.