Advertisement

Quick Links

INSTRUCTIONS FOR USE
Ninjo
6001313502_7EN • 2024-12

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Ninjo FD1600 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Arjo Ninjo FD1600

  • Page 1 INSTRUCTIONS FOR USE Ninjo 6001313502_7EN • 2024-12...
  • Page 2 © Arjo 2024. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice. The content of this publication may not be copied either whole or in part without the consent of Arjo.
  • Page 3: Table Of Contents

    Contents Foreword...............................5 Safety precautions............................7 Important..............................7 Isolator switch..............................7 In an emergency............................7 Presentation..............................8 Intended use..............................8 Machine models............................8 Control panel............................... 8 Components............................. 8 Functions..............................9 Display options............................9 Displayed while a program is running.......................9 Display after program end........................9 Cleaning and disinfection principles......................
  • Page 4 Using chemical agents..........................20 Arjo Liquids..............................20 Positioning containers..........................20 Switching containers..........................20 Water quality...............................21 Requirements............................21 Main factors...............................21 Recommendation............................21 Local standard............................21 Local standard, typical specification......................21 Environmental product declaration......................22 Comments..............................23 Distribution..............................23 Use................................23 Load pallets............................... 23 End of life disposal............................ 23 Symbol explanation...........................
  • Page 5: Foreword

    Customer contact information Unauthorised modifications on any Arjo product can affect its safety and performance. Arjo will For questions regarding this product or other not be held responsible for any accidents or Arjo products and services, contact Arjo, an Arjo incidents resulting from such modifications to its authorised representative or visit www.arjo.com.
  • Page 6 Definitions in this IFU WARNING Safety warning. Failure to understand and obey this warning may result in injury or machine damage. WARNING Hot surface. Risk of injury. WARNING Electricity. Risk of injury. NOTE Note means: This is important information for the correct use of this system or product.
  • Page 7: Safety Precautions

    Leakage in the system, for example a worn Instructions for use and accompanied door seal, must be rectified without delay. documents before using the product. • Spare parts must only be obtained from Arjo's sales companies. • The machine is designed for connection to water and electricity.
  • Page 8: Presentation

    To meet the requirements of EN ISO 15883, the items must be placed in the proper holder, recommended by Arjo. The customer is responsible for ensuring that an Installation Qualification, Operating Qualification and Performance Qualification according to ISO 15883 are carried out before the product is put into use.
  • Page 9: Functions

    Functions If an alternative with more than one parameter is selected, the parameters will be displayed FUNCTION alternately. The A value always remains at 0 Yellow flashing indicator light: The until the thermal disinfection begins. When the process can be aborted. temperature reaches 85 °C, the A value will Yellow indicator light: The machine is...
  • Page 10: Operating Instructions

    Operating instructions WARNING To prevent contamination, use personal protective equipment, such as gloves and protective glasses, when handling soiled waste containers and loading the goods into the flusher- disinfector. Program overview The machine's standard programs are described here. There may be other programs for the program buttons.
  • Page 11: Examples Of Goods And Program Selection

    Examples of goods and program WARNING selection Example of goods for economy program (A): To prevent cross-contamination, use a clean hand to touch the door handles and control panel. Hold dirty goods with the other hand. 1. Open the door. 2.
  • Page 12: Cooling (Option)

    1. When the program is complete: • The green light at lights up. • The door unlocks automatically. • Cooling of goods may continue after completed program (option) and is thus terminated when the door is opened. 2. Open the door and remove the goods. Cooling (option) The goods to be cleaned and disinfected are cooled using the following methods:...
  • Page 13: Care And Preventive Maintenance

    Arjo recommends that the stated maintenance operations be performed at intervals according to the service table. It is recommended to use Arjo Liquids Flusher Rinse to prevent hard water scale from building up on the goods and in the flusher-disinfector.
  • Page 14: Service Table

    Service table WARNING The machine is connected to the electricity supply and some components are live. WARNING Maintenance may only be performed by qualified service personnel. NOTE The service table is only shown for information purposes. The service table below shows recommended maintenance intervals. In addition to system testing, we recommend that appropriate cleaning tests and temperature validation are performed according to EN ISO 15883.
  • Page 15 MEASURE EVERY YEAR / EVERY OTHER TIME* 7500 CYCLES YEAR / 15000 CY- (MINUTES) CLES Check for leakage in the nozzle attachments and the hose connec- ● tions to the chamber. Check for proper operation of the ● temperature sensor. Check that the door seal and chamber seal are tight.
  • Page 16 MEASURE EVERY YEAR / EVERY OTHER TIME* 7500 CYCLES YEAR / 15000 CY- (MINUTES) CLES Check the hoses between the dos- 5-10 ing pump and the detergent hold- ● er. Replace if necessary. Check the dosage amount. Adjust 10-30 ● if necessary.
  • Page 17: Alarm

    Alarm Alarm indication WARNING If the disinfection process is cancelled, the goods are not clean. The goods must be processed again. If the red lamp at lights up, the disinfection has been aborted because of low temperature or some other error. The display shows an error code, see Troubleshooting on page 18.
  • Page 18: Troubleshooting

    Troubleshooting Types of codes NOTE Qualified service personnel should only be called in the event of error code F. In the case of error code U or H, the fault can be corrected by the user. • Ux is an information code – it does not need to be acknowledged. •...
  • Page 19 CODE DESCRIPTION MEASURE Could not fill/empty tank. Contact qualified service personnel. Leakage Contact qualified service personnel. Program error Contact qualified service personnel. TROUBLESHOOTING...
  • Page 20: Chemical Agents

    Follow the instructions on the container for additional instructions on using the chemical agent. Switching containers The machine has been proven and tested with Arjo’s chemical agents. We recommend that you WARNING use Arjo chemical agents. • Chemical agents recommended in the...
  • Page 21: Water Quality

    Recommendation • Arjo recommends that water used in the pre- rinsing, cleaning and disinfection and final rinsing phases should be of drinking quality in accordance with the guidelines.
  • Page 22: Environmental Product Declaration

    Environmental product declaration PLASTIC STAINLESS STEEL FD1600 FD1610 FD1615 UB Total weight kg / % by 90 / 100 100/ 100 75 / 100 weight Mechanical parts Stainless steel 41 / 45.5 60 / 60 35 / 46.6 Steel 1 / 1.1 1 / 1 1 / 1.3 Cast iron...
  • Page 23: Comments

    Users must make sure that the packaging material is sorted and recycled. The flusher-disinfector affects the environment during its entire life cycle. Arjo considers that electricity and water consumption have the greatest environmental impact. Based on 7000 process per year, the machine has a yearly...
  • Page 24: Symbol Explanation

    Symbol explanation SYMBOL EXPLANATION General warning Warning, hot surface Warning, electricity Wear protective gloves. Wear eye protection. Refer to instruction manual/ booklet - Instructions for use should be read. Name and address of the manufacturer Manufacturing date Use by date Catalogue number Serial number Indicates that the product is a Medical Device according to EU Medical...
  • Page 25 SYMBOL EXPLANATION WaterMark certification label with license number WMK26397 WaterMark WMTS104 LicWMK26397 SAI Global LAU-LBP13042 Water Supply (Water Fittings) Regulations and Scottish Water Byelaws DVGW CERT Conformity Certificate KIWA safety mark indicating compliance to EN 1717 and evaluation to KIWA BRL-K14011 Intertek Warnock Hersey Certification Mark Electrical test label.
  • Page 26 UK marking indicating conformity with UK Medical Devices Regulations 2002 (SI 2002 No 618, as amended) Figures indicate UK Approved Body supervision. UK Responsible Person & UK Importer: Arjo (UK) Ltd, ArjoHuntleigh House, Houghton Regis. LU5 5XF For Northern Ireland (NI) CE marking will still apply until further amendment to applica- ble regulations.
  • Page 27: Electromagnetic Compatibility

    Electromagnetic compatibility Intended environment: industrial electromagnetic environment. GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSION Emission test Compliance Guidance RF emissions CISPR 11 Group 1 This equipment uses RF energy only for its internal functions. Therefore its RF emissions are very low and RF emissions CISPR 11 Class A are not likely to cause any interference in nearby elec-...
  • Page 28 Intentionally left blank...
  • Page 29 Intentionally left blank...
  • Page 30 Intentionally left blank...
  • Page 31 AUSTRALIA FRANCE POLSKA Arjo Australia Arjo SAS Arjo Polska Sp. z o.o. Building B, Level 3 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 11 Talavera Road CS 70133 PL-62-052 KOMORNIKI (Poznań) Macquarie Park, NSW, 2113, Australia FR-59436 RONCQ CEDEX...
  • Page 32 At Arjo, we believe that empowering movement within healthcare environments is essential to quality care. Our products and solutions are designed to promote a safe and dignified experience through patient handling, medical beds, personal hygiene, disinfection, diagnostics, and the prevention of pressure injuries and venous thromboembolism. With over 6500 people worldwide and 65 years caring for patients and healthcare professionals, we are committed to driving healthier outcomes for people facing mobility challenges.

This manual is also suitable for:

Ninjo fd1610Ninjo fd1615 ub

Table of Contents