Advertisement

INSTRUCTIONS FOR USE
Carendo
04.CC.01_27EN • 02/2019

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Carendo and is the answer not in the manual?

Questions and answers

Summary of Contents for Arjo Carendo

  • Page 1 INSTRUCTIONS FOR USE Carendo 04.CC.01_27EN • 02/2019...
  • Page 2 Mandatory to read the Instructions for Use Design Policy and Copyright ® and ™ are trademarks belonging to the Arjo group of companies. © Arjo 2019. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.
  • Page 3: Table Of Contents

    Contents Foreword ...................4 Intended Use ................5 Safety Instructions ..............6 Preparations ................7 Parts Designation ..............8 Product Description / Functions ..........10 Reclined Function ..............10 Lifting Function – High/Low ..........10 Care Raiser .................10 Leg Support .................10 Driving Handles ..............11 Manoeuvre Handles ............11 Castors ................11 Safety Belt ................11 Hand Control ...............12 Back Pillow (Accessory) ............12...
  • Page 4: Foreword

    Foreword Definitions in this IFU Thank You for Purchasing Arjo Equipment. Your Carendo shower and toilet chair is part of a series WARNING of quality products designed especially for hospitals, nursing homes and other health care use. Means: Safety warning. Failure to understand and obey Please contact us if you have any questions about the this warning may result in injury to you or to others.
  • Page 5: Intended Use

    The equipment is intended for indoor use. maintenance in this IFU. The Carendo must be used by appropriately trained Depending on level of wear some equipment parts such caregivers with the adequate knowledge of the care...
  • Page 6: Safety Instructions

    (ISO 10993). the body and use designated grab supports during any movement. NOTE Do not lift the Carendo shower and toilet chair using the armrests; use the chassis. WARNING Only one function movement (e.g reclined To avoid the device from tipping and the function) can be done at a time.
  • Page 7: Preparations

    Actions Before the First Use (4 steps) (9 steps) 1 Check that all parts of the Carendo shower and 1 Visually check the package for damage. If the toilet chair are in place. Compare to section Parts product looks damaged make a complaint to the Designation on page 8.
  • Page 8: Parts Designation

    Parts Designation 1 Chassis 2 Care Raiser 3 Castors 4 Lifting frame 5 Backrest 6 Neck pillow 7 Transfer handles 8 Manoeuvring handles 9 Storage net 10 Safety belt 11 Seat 12 Leg support 13 Armrests 14 Hand control 15 Battery 16 Battery charger 17 Battery holder 18 Seat cushion...
  • Page 10: Product Description / Functions

    Product Description / Functions Reclined Function Fig. 1 A function which makes hygienic care of the patient easier. From a sitting position, a reclined position can be achieved by means of turning gears which makes the backrest move in backwards and upright directions. Corresponding to this movement the leg rest is moving up and down.
  • Page 11: Driving Handles

    (See Fig. 2) Castors Fig. 3 The Carendo shower and toilet chair is equipped with lockable castors. The castors will brake by pressing the castor lock tab down with your foot (A). (See Fig. 3) The castors are released by pressing up the centre tab (B).
  • Page 12: Hand Control

    Hand Control Fig. 1 Connect the connector of the hand control on the right side of the backrest. (See Fig. 1) The hand control has a hook on the back, making it easy to attach Fig. 2 it to various parts of the chair. (See Fig. 2) Hand control Fig.
  • Page 13: Battery Charger

    Battery Charger Fig. 1 The battery holder should be attached to a suitable wall and the charger connected to a properly grounded wall socket. (See Fig. 1) NOTE The battery charger is designed for use in a dry location and should not be installed in bathrooms, etc. See NEA 8000 Battery Charger IFU.
  • Page 14: Emergency Lowering

    Emergency Lowering Fig. 1 In case of electrical malfunction, the chair can be lowered by pushing the green button. (See Fig. 1) NOTE This function should only be used in emergency. Emergency Stop Fig. 2 In any case of emergency, all functions can be stopped by pushing the red button (A).
  • Page 15: Bedpan And Bedpan Holder

    Bedpan and Bedpan Holder Fig. 1 WARNING To avoid entrapment or pinching of genitals, make sure there is enough clearance during movement over the tub edge, toilet, bedpan or other furniture. A bedpan holder can be mounted under the seat. (See Fig. 1-2) See separate instruction Bedpan Holder.
  • Page 16: Independent Patients

    1 Apply the castor brakes on the wheelchair. 2 Lower the Carendo shower and toilet chair so that its seat is at the same level as the wheelchair seat, or as close as possible.
  • Page 17: Dependent Patients

    Follow the separate operating instructions regarding the passive lift. 1 Apply the brakes on the Carendo shower and toilet chair. 2 Place the passive lift with the patient as close as possible to the Carendo shower and toilet chair. Position the patient in the sling directly over the seat of the Carendo shower and toilet chair.
  • Page 18 Carendo shower and toilet chair. (See Fig. 2) 3 Apply the brakes on the active lift. 4 Lower the patient till he/she reaches the seat of the Carendo Fig. 3 shower and toilet chair. Unhook sling and remove, if present.
  • Page 19: Transfer To The Shower

    Transfer to the Shower (18 steps) Fig. 1 1 Place the Carendo shower and toilet chair with the patient in the shower room. 2 Apply the castor brakes. 3 Raise the chair in a sitting position to a comfortable height for the caregiver.
  • Page 20 16 Put on the incontinence pads, if used, and pull up the trousers completely. (See Fig. 7) 17 Adjust the Carendo shower and toilet chair back to a sitting position and lower the chair. Secure the patient with the safety belt, if needed.
  • Page 21: Transfer To The Toilet

    Transfer to the Toilet (13 steps) The Carendo shower and toilet chair can be used regardless of whether the toilet is attached to the wall or to the floor. See the measurements below on matching toilet chairs: min 5.5 in 16.9 in...
  • Page 22 11 Put on the incontinence pads, if used, and pull up the Fig. 5 trousers completely. (See Fig. 5) 12 Adjust the Carendo shower and toilet chair back to a sitting position and lower the chair. WARNING To avoid entrapment, make sure to keep the patient’s hair, arms and feet close to the body and...
  • Page 23: Home Care Use

    Home Care Use Using the Carendo For Distributors and Renters Before using the Carendo shower and toilet chair in When transporting the Carendo shower and toilet home environment make sure that: chair to the customer with a van: (See Fig. 1-3) •...
  • Page 24 Fig. 1 Fig. 2 36" (900 mm) 36" (900 mm) 87" (2200 mm) 88" (2200 mm) Fig. 3 1000 39" 37" 36" 33" 31" 30" 1000 39" 37" 36" 33" 31" 30"...
  • Page 25: Cleaning And Disinfection Instructions

    Arjo Clean If you have any questions regarding disinfecting the • Arjo General Purpose Disinfectant Carendo shower and toilet chair, or wish to order • Arjo Disinfectant Cleanser IV disinfectant, contact Arjo Customer Service. See section Parts and Accessories on page 42.
  • Page 26 Always Follow these 13 Steps for Disinfection Cleaning and Disinfection (step 8-13 of 13) If an Arjo shower panel will be used, see respective 8 Spray disinfectant on: IFU. • the Carendo shower and toilet chair. Removing visual residue • all removed parts...
  • Page 27: Battery Instructions

    To avoid bodily injury, do NOT crush, puncture, open, dismantle or otherwise • If the Carendo shower and toilet chair is used mechanically interfere with the battery. frequently, it is recommended to use two batteries, one in operation and one as a stand-by in the •...
  • Page 28: Care And Preventive Maintenance

    Care and Preventive Maintenance Caregiver Obligations The Carendo shower and toilet chair is subject to wear and tear, and the following actions must be performed when specified to make sure that the product remains within its original manufacturing specification. WARNING To avoid malfunction resulting in injury, make sure to conduct regular inspections and follow the recommended maintenance schedule.
  • Page 29 IFU. Every day Fig. 1 • Disinfect. The Carendo shower and toilet chair has to be disinfected immediately after every use. Arjo disinfectants and shampoos should be used in recommended concentrations. See section Cleaning and Disinfection Instructions on page 25.
  • Page 30 Every sixth months Fig. 7 Lubricate the locking pins. Use grease Wekem WS267 (can be purchased from Arjo). (See Fig. 7) Yearly The Carendo shower and toilet chair has to be serviced according to the table QUALIFIED PERSONNEL Action/Check on page 31.
  • Page 31 WARNING To avoid injury and/or unsafe product, the maintenance activities must be carried out at the correct frequency by qualified personnel using correct tools, parts and knowledge of procedure. Qualified personnel must have documented training in maintenance of this device. QUALIFIED PERSONNEL Action/Check Every YEAR Lubricate bearings...
  • Page 32: Troubleshooting

    Troubleshooting NOTE If the product does not work as intended, immediately contact your local Arjo representative for support. PROBLEM ACTION It is only possible to lower the lift. The lifting device of this equipment is equipped with a a safety nut.
  • Page 33: Technical Specifications

    Technical Specifications Carendo, Lift and Shower Chair Model BIB2000-01 Safe Working Load (SWL) (Maximum total load) 136 kg (300 lbs) Maximum weight of lift 64 kg (141 lbs) Maximum total weight (lift + patient) 200 kg (441 lbs) Chassis, maximum width 640 mm (25 1/4”)
  • Page 34 This requirement applies in the European Union. Do not dispose these products as unsorted municipal waste. You can return your device and accessories to Arjo or you can contact your local authorities for advice on disposal.
  • Page 35: Dimensions

    Dimensions Measurements in mm and in inches (") Sitting, lowest position Sitting, highest position Reclined, lowest position Reclined, highest position Care Raise, lowest position Care Raise, highest position...
  • Page 36 Measurements in mm and in inches (") Ø1530 (60 1/4”) Ø960 ( 5/8”) Turning radius, sitting Turning radius, reclined...
  • Page 37: Labels

    Labels Label explanation NEA0100-083 Product no. of the battery Attention Important user information for 24V DC 2.5 Ah Label correct use. Mandatory to read the Instructions Approval States approvals for the current for Use Label product. Battery Label States safety and environmental information for the battery.
  • Page 38 Attention Symbol Battery Label Approval Label Identification Label Emergency Lowering Label Data Label...
  • Page 39: List Of Standards And Certificates

    List of Standards and Certificates STANDARDS/ DESCRIPTION CERTIFICATE EN/IEC 60601-1:2005 Medical Electrical Equipment - Part 1: General Requirements for AMD1:2012 Basic Safety and Essential Performance ANSI/AAMI ES60601-1 Medical Electrical Equipment - Part 1: General Requirements for (2005) AMD 1 (2012) Basic Safety and Essential Performance CAN/CSA-C22.2 No.
  • Page 40: Electromagnetic Compatibility

    (electromagnetic interference) from external sources. Some procedures can help reduce electromagnetic interferences: • Use only Arjo cables and spare parts to avoid increased emissions or decreased immunity which can compromise the correct functioning of the equipment. • Ensure that other devices in patient-monitoring and/or life-support areas comply to accepted emissions standards.
  • Page 41 Guidance and manufacturer’s declaration – electromagnetic immunity Electromagnetic Immunity test IEC 60601-1-2 test level Compliance level environment – guidance Electrostatic ±2kV, ±4kV, ±8kV, ±15kV air ±2kV, ±4kV, ±8kV, ±15kV air Floors should be wood, discharge (ESD) concrete or ceramic tile. If floors are covered with synthetic material the relative EN 61000-4-2...
  • Page 42: Parts And Accessories

    Parts and Accessories For disinfectant contact Back Pillow NEA0001-031 Neck Pillow 8350897-031 the Arjo local representative Bed pan CDA1500-07 Safety belt S8662104-03 Battery NEA0100 Holder NEA1101-90 Charger NEA8000...
  • Page 43 AUSTRALIA FRANCE POLSKA Arjo Australia Pty Ltd Arjo SAS Arjo Polska Sp. z o.o. 78, Forsyth Street 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 O’Connor CS 70133 PL-62-052 KOMORNIKI (Pozna ) AU-6163 Western Australia FR-59436 RONCQ CEDEX...
  • Page 44 At Arjo, we are committed to improving the everyday lives of people affected by reduced mobility and age-related health challenges. With products and solutions that ensure ergonomic patient handling, personal hygiene, disinfection, diagnostics, and the effective prevention of pressure ulcers and venous thromboembolism, we help professionals across care environments to continually raise...

Table of Contents

Save PDF