2.2.2    Seating Limitations
AVIVA RX with Lifter
110°
 
Limitation
Tilting >110° inhibited
A
Lifting inhibited
B
There are no other seating limitations for AVIVA RX without lifter and 
TDX SP2 (with / without lifter) despite of the Max. Back Angle Limit
1585725-M
Cause for Limitation
If lifter is raised
 
>50 mm
l
If backrest angle is
 
>110°
l
Max. Back Angle Limit
The maximum back angle limits how far the backrest can be reclined 
using either the tilt or recline actuators. The seating systems are 
typically preset at the factory to the maximum allowable angle and do 
not require any further adjustment unless the maximum angle needs 
to be decreased (see hazard statement below).
Max. Back Angle Limit =
168° (TDX SP2)
170° (AVIVA RX)
NOTICE!
— When establishing the maximum back limit, always 
consider the size and location of any personal gear that 
may be carried on the wheelchair, as it could cause 
interference between the backrest and the power 
wheelchair base when fully tilted / reclined, and damage 
the actuator and / or wheelchair.