Page 1
English Instructions for use MK-dent Air Motors Version: 2025-01-06-RevG...
Page 2
Unpacking ................................9 Assembly ................................9 Preparing the connection ............................9 6.4.1 Connection to the treatment unit....................... 10 6.4.2 Connection to the instruments ........................10 Startup and operation ............................11 6.5.1 Check for correct operation ........................11 MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 3
Technical data ......................... 19 Explanation of symbols ....................20 Guarantee and warranty ....................21 15.1 Guarantee ................................ 21 15.2 Legal warranty ..............................21 15.3 Conditions ................................ 21 15.4 Utilization ................................. 22 15.5 Exclusion of liability ............................22 MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 4
1 About these instructions for use These instructions for use apply to all MK-dent air motors (see section 4 Product description). All products listed here are only intended for dental treatment in the field of dentistry. Any kind of misuse or modification of the product is not permitted and may lead to hazards.
Page 5
Secondary users are the practice staff who are responsible for transport, processing and cleaning. User environment Application takes place in dental practices or dental facilities. Users wear protective gloves. The peripherals used fulfil the requirements of ISO 7494. MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 6
Never disassemble or modify this product. This instrument must not be repaired by the user in the event of a malfunction. Contact your dealer or the MK-dent repair service. Please do not use the product if the performance of the product deteriorates or in case of malfunctions.
Page 7
If anything is missing, contact your dealer immediately. Required materials that are not included NOTE NOTE The scope of delivery does not include a dental handpiece. The scope of delivery does not include MK-dent Premium Service Oil LU1011. Configurations Not applicable MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 8
The use of unauthorized accessories or unauthorized modifications to the product can lead to injuries or instrument malfunctions. CAUTION → Only use spare parts and components that have been approved by MK-dent for combination with the product. AM0002, AM0004 Replacement motor without light...
Page 9
When a handpiece is removed, water and oil remain on the motor connection. Remove the water and oil completely. Make sure that no water or oil remains in your connection hose. If present, remove with a dry cloth. MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 10
Attach the hose connection to the air motor and tighten the thread. 6.4.2 Connection to the instruments Lightly oil the O-rings on the motor with MK-dent Premium Service Oil LU1011. Push the instrument onto the air motor until it clicks into place.
Page 11
As the hose supplied is longer than necessary, cut it to the appropriate length before use. 6.5.5 Water regulation (AM1116) The amount of water in the spray can be adjusted by turning the regulating ring at the rear of the air motor. MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 12
Always stop working whenever operating noises become louder or irregular, there are stronger vibrations or in any case of overheating or visual damage. In CAUTION these cases, contact your dealer or the MK-dent repair service. Disconnecting after use WARNING The product must not be dropped.
Page 13
→ Flush the instrument immediately when you finish working on the patient. Do not place the instrument in solutions or similar. NOTE NOTE Clean the connection hose and the inside of the instrument connector according to 6.4 Preparation for use before reconnecting the instrument. MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 14
Blow the motor with medical compressed air from the outside and from the inside until no more drops of water are visible. Drinking water quality according to the EU Drinking Water Regulation (total bacterial count max. 100 CfU/ml) MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 15
Observe the exposure time specified by the disinfectant manufacturer. Follow the disinfectant’s instructions for use. Automatic processing Not applicable WARNING Automatic reprocessing is not to be used for air motors. WARNING Sterilization Not applicable WARNING Sterilization in an autoclave is not to be used for air motors. WARNING MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 16
Inappropriate maintenance or service can lead to malfunctions and/or deterioration. WARNING → Always conduct proper maintenance and servicing. CAUTION For safe use of the product, perform checks and maintenance of the instrument CAUTION according to these instructions for use. MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 17
Process this instrument according to section 8 Processing according to ISO 17664-1 and send it to your dealer or directly to MK-dent if it is defective. Please contact MK-dent customer service to arrange for shipping or collection for factory repair. Questions about noise, maintenance, processing and functionality can often be clarified in a conversation, making it unnecessary to send in the product.
Page 18
Do not use lubricants such as grease to oil the O-rings. WARNING Remove the worn O-rings and replace them with new ones. Carefully oil the new O-rings on the motor with MK-dent Premium Service Oil LU1011. 12.2 Replacing the light source WARNING Risk of burns due to hot lamp.
Page 20
Medical device (CE conformity assessment procedure) Catalog number Serial number Manufacturer Date of manufacture Unique product identification – UDI (Unique Device Identifier) – indicates a barcode as containing Unique Device Identification HIBC Code (Health Industry Bar Code) MK-dent-Instructions-for-use-airmotors-EN-RevG Version: 2025-01-06-RevG...
Page 21
(ISO 14457 Dentistry - Handpieces and motors). In addition, we offer you comprehensive warranty services to ensure your satisfaction. 15.1 Guarantee MK-dent grants the end customer a guarantee of perfect function and freedom from defects in material and workmanship from the date of purchase for a period of 24 months. Exclusions: •...
Page 22
15.4 Utilization In the event of a warranty claim, please contact MK-dent customer service directly. In the event of a warranty claim, please contact your contractual partner. Contact MK-dent customer service Phone: +49 4532 40049-0 Email: service@mk-dent.com Please have the following information ready: •...