Important Details
The MK-dent Air Motor (AM0012/ AM0014)
- is only intended for dental treatment of patients.
- is according to relevant national regulations as a medical device.
These provisions indicate, that
- applicable accident prevention regulations and health and safety regulations must be observed.
- only a competent user is allowed to use the product according to this instruction.
- only faultless work equipment is to be used.
- any contamination from the product must be avoided.
- patients, users and third parties are to be protected from danger.
Meaning of Symbols
REF
Item No.
Connection to the Treatment Unit
Before first use and after each use, sterilize the Air Motor. Always use dry, cleaned and decontaminated compressed air.
Premature wear from contaminated and humid compressed air can shorten the lifetime of the product.
Checking the Compressed Air with a Manometer
Install the manometer between tubing and Air Motor. The following values should be displayed:
Forced Air (F.A./ T.R.)
Return Air (R.A./ R.L.)
Water (W.)
Spray Air (S.A./ Spr.L.)
Connecting to the INTRAmatic-connection of Handpieces
In order to avoid damage to the driver, do not attach or release the Air Motor while the handpiece is in movement.
Technical Facts
MK-dent Air Motor (AM0012/ AM0014)
Input speed: 5.000 - 20.000 rpm
Air input: 2,2 - 2,8 bar
Air use: up to 60Nl/ min.
Connection: 2-/ 4-hole standard connection, INTRAmatic connector
SN
Serial No.
Caution!
Place the connection of the turbing on the Air Motor and
turn the thread until stop.
=
2,2 - 2,8 bar
<
1,1 bar
=
0,8 - 2,5 bar
=
1,0 - 4,0 bar
Place the handpiece on the Air Motor until it audibly locks in place.
Sterilizable in an
Suitable for spray- and
autoclave up to max.
wipe disinfection
135°C/ 275°F
Instruction manual
AM0012, AM0014
0123
0123
CE-Sign

Advertisement

loading
Need help?

Need help?

Do you have a question about the AM0012 and is the answer not in the manual?

Questions and answers

Summary of Contents for mk-dent AM0012

  • Page 1 Instruction manual AM0012, AM0014 0123 Important Details The MK-dent Air Motor (AM0012/ AM0014) - is only intended for dental treatment of patients. - is according to relevant national regulations as a medical device. These provisions indicate, that - applicable accident prevention regulations and health and safety regulations must be observed.
  • Page 2 Turn the control ring to change the speed and the rotation direction. Disinfektion The MK-dent Air Motor can be disinfected externally. Only use DGHM listed chemical disinfectant for spray and wipe disinfection. Don‘t place the Air Motor in disinfectant baths. Results are mulfunctions.
  • Page 3 Other claims of any kind, in particular damage compensation are excluded. In the event of default, gross negligence or intent, this shall only apply unless there are compelling legal regulations. MK-dent is not liable for defects and their consequences that have arisen whose arising may be due to natural wear, improper handling, improper cleaning, or maintenance, non-compliance with operating or manual instructions, calcination or corrosion, contaminated air and water supply or chemical or electrical influences that are unusual or not permitted according to MK-dent’s instruction for use or to other manufacturer‘s instructions.

This manual is also suitable for:

Am0014