Water I.D. PoolLab 1.0 User Manual

Photometer
Hide thumbs Also See for PoolLab 1.0:

Advertisement

Quick Links

User Manual
Gebruikershandleiding
Brugervejledning
Användarmanual
Руководство
пользователя
Quick Start
Guide included!

Advertisement

loading
Need help?

Need help?

Do you have a question about the PoolLab 1.0 and is the answer not in the manual?

Questions and answers

Summary of Contents for Water I.D. PoolLab 1.0

  • Page 1 User Manual Gebruikershandleiding Brugervejledning Användarmanual Руководство пользователя Quick Start Guide included!
  • Page 2 PoolLab 1.0 Website ® Water-i.d. ® Website...
  • Page 3 Content | Inhoud | Indhold | Innehåll | Содержание Delivery Content | Inhoud | Levering af indhold | Leveransinnehåll | Содержание доставки Batteries | Batterijen | Batterier | Batterier | Батареи Switch on/off | In-/uitschakelen | Tænd/sluk | Slå på/av | Включение/выключение Notes | Opmerkingen | Noter | Anteckningar | Примечания...
  • Page 4 Content | Inhoud | Indhold | Innehåll | Содержание OR / UR Error codes | Foutcodes | Fejlkoder | Felkoder | Коды ошибок Troubleshooting | Problemen oplossen | Fejlfinding | Felsökning | Устранение неполадок Changing Cuvette/Calibration | Van cuvet wisselen/Kalibratie | Udskiftning af kuvette/kalibrering | Byte av kyvett/kalibrering | Смена...
  • Page 5 Delivery Content | Inhoud | Levering af indhold | Leveransinnehåll | Содержание доставки 01 x PoolLab 1.0 ® 01 x Light shield 03 x AAA Batteries 01 x Crushing | Stirring Rods 01 x 10 ml syringe 01 x User Manual 20 x Phenol Red Photometer tablets 20 x DPD N°...
  • Page 6 Batteries | Batterijen | Batterier | Batterier | Батареи Change | Verander | Forandring | Förändring | Изменить No rechargeable batteries! | Geen oplaadbare batterijen! | Ingen genopladelige batterier! | Inga uppladdningsbara batterier! | Никаких перезаряжаемых батареек! 3 x AAA...
  • Page 7 Switch on/off | In-/uitschakelen | Tænd/sluk | Slå på/av | Включение/выключение Seite 7 00:02 The On/Off button can also be used to skip countdown during measurement (not recommended). De Aan/Uit knop kan ook gebruikt worden om het aftellen over te slaan tijdens het meten (niet aanbevolen). On/Off-knappen kan også...
  • Page 8 Notes | Opmerkingen | Noter | Anteckningar | Примечания dness Total Hardness ardness Total Hardness Total Hardness Total Hardness Total Hardness Total Hardness Total Hardness Total Hardness ardness Total Hardness Hardness Total Hardness Total Hardness Total Hardness Total Hardness Total Hardness Total Hardness Total Hardness Total Hardness Total Hardness Total Hardness Hardness Total Hardness Total Hardness Total Hardness Total Hardness...
  • Page 9 Important | Belangrijk | Vigtigt | Viktigt | Важно Detergent It is important to clean the device after each measurement to get rid of any reagent residues! Please ensure that the cuvette has been cleaned before each measurement (e.g. under clear water/or simply rinsing the cuvette in the pool is sufficient as long as no residues remain).
  • Page 10 Do not leave the device in the sun! Laat het apparaat niet in de zon liggen! Efterlad ikke enheden i solen! Lämna inte apparaten i solen! Не оставляйте устройство на солнце! The PoolLab 1.0 is also suitable for saltwater pools/salt ® electrolysis pools! De PoolLab 1.0 is ook geschikt voor zoutwaterzwembaden/ ®...
  • Page 11 Tablet Mode → Liquid Mode Tabletmodus → Vloeistofmodus Tablet-tilstand → flydende tilstand Läget för surfplatta → Läget för vätska Режим планшета → Режим жидкости pH | fCl | tCl | cCl | Br | ClO Scan the QR-code to watch our instruction video...
  • Page 12 Tablet Mode: TABLET TEST TEST MODE Liquid Mode: LIQUID TEST TEST MODE...
  • Page 13 From firmware version 72 onwards, you have the option of measuring the following parameters with tablets as well as with liquid reagents: pH, chlorine, chlorine dioxide, ozone and bromine. You can choose between two measurement settings in the unit: Tablets and Liquid reagents. You can switch between the modes by pressing and releasing the ZERO &...
  • Page 14 Från och med firmwareversion 72 har du möjlighet att mäta följande parametrar både med tabletter och med flytande reagenser: pH, klor, klordioxid, ozon och brom. Du kan välja mellan två mätinställningar i enheten: Tabletter och flytande reagens. Du kan växla mellan lägena genom att trycka in och släppa ZERO &...
  • Page 15 QUICK START GUIDE Snelstartgids Hurtig start-guide Guide för snabb start Краткое руководство пользователя Scan the QR-code to watch our instruction video...
  • Page 16 3 x AAA 00:02 START: Take 10 ml Water Sample ZERO TEST...
  • Page 17 Remove Lightshield TEST TEST Reagent(s) for your test (refer to chapter in manual) Put on Lightshield TEST TEST Await Countdown Shortcut for your Test (Refer TEST result in PPM to chapter in manual)
  • Page 18 Empty & Clean For Next Test: Take 10 ml If Device was If Device was not switched off, start switched off, start from step 4 from step 7 Repeat Repeat steps steps 4-13 7-13 for further for further Parameters Parameters End Of Quick Start Guide | Einde Snelstartgids | Slut på...
  • Page 19 ZERO START: Take 10 ml Water Sample TEST ZERO...
  • Page 20 Only one Time Per Test Batch | Slechts één keer per testbatch | Kun én gang pr. testbatch | Endast en gång per testparti | Только один раз для каждой тестовой партии The „ZERO“ step (page 19) is only necessary once after switching on.
  • Page 21 Active Oxygen (MPS) Actieve zuurstof (MPS) Aktiv ilt (MPS) Aktivt syre (MPS) Активный кислород (MPS) 0.0 – 30.0 ppm (mg/l) DPD N°4 Photometer* *not part of Standard Equipment Take 10 ml Water Sample...
  • Page 22 1 x DPD N°4 ZERO! (p. 19) Photometer* Completely Dissolved Residue TEST ZERO TEST TEST 12.2 ppm...
  • Page 23 Alkalinity Alkaliteit Alkalinitet Alkalinitet Щелочность 0 – 200 ppm (mg/l) CaCO Alkalinity - M Photometer Take 10 ml Water Sample...
  • Page 24 1 x Alkalinity-M ZERO! (p. 19) Photometer Completely Dissolved Residue TEST ZERO TEST 96 ppm TEST CaCO3...
  • Page 25 Bromine Broom Brom Brom Бром Tablet Mode: Liquid Mode: 0.00 – 18.00 ppm (mg/l) 0.00 – 9.00 ppm (mg/l) DPD N°1 Photometer Tablet + DPD 1B Liquid* Glycine* Glycine* *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 26 Only if your water sample does contain Chlorine next to Bromine (both disinfectants used), the following procedure „A“ needs to be followed and Glycine* reagent needs to be used. Otherwise (only Bromine present), please follow procedure „B“. Alleen als uw watermonster naast broom ook chloor bevat (beide gebruikte desinfectiemiddelen), moet de volgende procedure “A”...
  • Page 27 With Chlorine | Met chloor | Med klorin | Med klorin | С хлором Glycine* Completely Dissolved TEST TEST Residue Tablet or Liquid? (p. 11) Completely Dissolved DPD N°1 TEST TEST Photometer Residue...
  • Page 28 Without Chlorine | Zonder chloor | Uden klor | Utan klor | Без хлора Tablet or Liquid? (p. 11) Completely Dissolved TEST TEST DPD N°1 Residue Photometer 8a / 6 TEST 6.00 ppm...
  • Page 29 Calcium Hardness Calciumhardheid Calcium-hårdhed Kalcium Hårdhet Твердость кальция 0 – 500 ppm (mg/l) CaCO3 POL20CH1* POL20CH2* *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 30 POL20CaH1** POL20CaH2** *Shake before use! CaH1 CaH2 10 x 10 x TEST TEST TEST...
  • Page 31 68 ppm CaCO 3...
  • Page 32 Hardness Conversion | Conversie van hardheid | Omregning af hårdhed | Omvandling av hårdhet | Преобразование твердости CaCO °dH* °e* °f* mval S 4,3 mg/l mmol/l (KH) (CH) (DC) 1 mg/l 0.01 0.056 0.07 0.02 CaCO 1 mmol/l 10.0 S 4,3...
  • Page 33 Chlorine Chloor Klor Klor Хлор Tablet Mode: Liquid Mode: 0.00 – 8.00 ppm (mg/l) 0.00 – 4.00 ppm (mg/l) DPD N°1 Photometer Tablet + DPD DPD N°3 Photometer Tablet DPD 3C Liquid* *not part of Standard Equipment Take 10 ml Water Sample...
  • Page 34 ZERO! (p. 19) Tablet or Liquid? (p. 11) ZERO TEST DPD N°1 Photometer TEST TEST...
  • Page 35 Tablet or Liquid? (p. 11) 1.27 ppm 1.27 ppm DPD N°3 Photometer 1.27 ppm 1.27 ppm...
  • Page 36 1.65 ppm Total Chlorine is measured directly after free Chlorine without emptying the cuvette. The DPD N°3 tablet is added to the sample water which already contains the DPD N°1 tablet (dissolved). Combined Chlorine is calculated as Total Chlorine minus free Chlorine. The free chlorine measurement must be taken within 1 minute after dissolving the tablet.
  • Page 37 Total klor måles direkte efter frit klor uden at tømme kuvetten. DPD N°3-tabletten tilsættes til prøvevandet, som allerede indeholder DPD N°1-tabletten (opløst). Kombineret klor beregnes som total klor minus frit klor. Målingen af frit klor skal foretages inden for 1 minut efter opløsning af tabletten.
  • Page 38 Chlorine Dioxide Chloordioxide Klordioxid Klordioxid Диоксид хлора Tablet Mode: Liquid Mode: 0.00 – 15.00 ppm (mg/l) 0.00 – 7.60 ppm (mg/l) DPD N°1 Photometer Tablet DPD 1A + DPD Liquid* Glycine* Glycine* *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 39 Only if your water sample does contain Chlorine next to Chlorine Dioxide (both disinfectants used), the following procedure „A“ needs to be followed and Glycine* reagent needs to be used. Otherwise (only Chlorine Dioxide present), please follow procedure „B“. Alleen als je watermonster naast Chloordioxide ook Chloor bevat (beide gebruikte desinfectiemiddelen), moet de volgende procedure “A”...
  • Page 40 With Chlorine | Met chloor | Med klorin | Med klorin | С хлором Glycine* Completely Dissolved TEST TEST Residue Tablet or Liquid? (p. 11) Completely Dissolved DPD N°1 TEST TEST Photometer Residue...
  • Page 41 Without Chlorine | Zonder chloor | Uden klor | Utan klor | Без хлора Tablet or Liquid? (p. 11) Completely Dissolved TEST TEST DPD N°1 Residue Photometer 8a / 6 4.11 ppm TEST CLO2...
  • Page 42 Cyanuric Acid Cyanuurzuur Cyanursyre Cyanursyra Циануровая кислота 0 – 160 ppm (mg/l) CYA-Test Photometer* *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 43 1 x CYA-Test Photometer* Completely Dissolved TEST TEST Residue TEST 42 ppm...
  • Page 44 (LR) Hydrogen Peroxide Waterstofperoxide Hydrogenperoxid Väteperoxid Перекись водорода 0.00 – 2.90 ppm (mg/l) Hydr. Peroxide LR Photometer* *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 45 1 x Hydr. Peroxide LR Photometer* Completely Dissolved TEST TEST Residue 1.4 ppm TEST H202...
  • Page 46 (HR) Hydrogen Peroxide Waterstofperoxide Hydrogenperoxid Väteperoxid Перекись водорода 0 – 200 ppm (mg/l) Hydr. Peroxide HR Photometer* | Acidifying *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 47 1 x Hydr. Peroxide HR Photometer* Completely Dissolved Residue TEST TEST 1 x Acidifying Completely Dissolved Residue TEST TEST 1.4 ppm TEST H202...
  • Page 48 Ozone Ozon Озон Tablet Mode: Liquid Mode: 0.00 – 5.40 ppm (mg/l) 0.00 – 2.70 ppm (mg/l) DPD N°1 Photometer Tablet + DPD DPD N°3 Photometer Tablet DPD 3C Liquid* Glycine* Glycine* *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 49 Only if your water sample does contain Ozone next to Chlorine (both disinfectants used), the following procedure „B“ needs to be followed and Glycine* reagent needs to be used. Otherwise (only Ozone present), please follow procedure „A“. Alleen als je watermonster Ozon bevat naast Chloor (beide gebruikte desinfectiemiddelen), moet de volgende procedure “B”...
  • Page 50 Without Chlorine | Zonder chloor | Uden klor | Utan klor | Без хлора Tablet or Liquid? (p. 11) Completely Dissolved DPD N°1 + TEST TEST DPD N°3 Residue Photometer 0.3 ppm TEST...
  • Page 51 With Chlorine | Met chloor | Med klorin | Med klorin | С хлором Tablet or Liquid? (p. 11) Completely Dissolved DPD N°1 + 0.3 ppm 0.3 ppm DPD N°3 Residue Photometer 0.3 ppm STEP 2...
  • Page 52 STEP 2 Take 10 ml Water Sample Completely Glycine* Dissolved Residue STEP 2 STEP 2...
  • Page 53 Tablet or Liquid? (p. 11) Completely Dissolved DPD N°1 + STEP 2 STEP 2 DPD N°3 Residue Photometer 0.6 ppm STEP 2...
  • Page 54 Tablet Mode: Liquid Mode: 6.50 – 8.40 6.50 - 8.40 Phenol Red Photometer Phenol Red Liquid* *not part of Standard Equipment Take 10 ml Water Sample...
  • Page 55 ZERO! (p. 19) Phenol ZERO Red Photometer TEST Phenol Red* Completely Dissolved TEST TEST Residue...
  • Page 56 7.10 The Total Alkalinity value has to be minimum 50 mg/l to obtain a correct pH value. De totale alkaliteitswaarde moet minimaal 50 mg/l zijn om een correcte pH-waarde te verkrijgen. Den samlede alkalinitetsværdi skal være mindst 50 mg/l for at opnå...
  • Page 57 PHMB 5 – 60 ppm (mg/l) PHMB Photometer* *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 58 1 x PHMB Photometer* Completely Dissolved TEST TEST Residue TEST 42 ppm phmb...
  • Page 59 It is imperative that you clean the objects used for the measure-ment and come into contact with the sample water containing the reagent (cuvette, lid, stirring rod) thoroughly with a brush, water and then with distilled water, otherwise the measuring equipment may turn blue over time. Alkalinity values (M) ≠...
  • Page 60 Total Hardness Totale hardheid Total hårdhed Total hårdhet Общая жесткость 0 – 500 ppm (mg/l) CaCO POL20TH1* POL10TH2* *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 61 POL20TH1** POL10TH2** *Shake before using! 10 x TEST TEST 57 ppm TEST CACO3...
  • Page 62 Urea Ureum Urea Urea Мочевина 0.00 – 2.50 ppm (mg/l) Dechlor* | PL Urea | PL Urea Ammonia N°1* | Ammonia N° *not part of Standard Equipment ZERO! (p. 19) Take 10 ml Water Sample ZERO...
  • Page 63 Dechlor* TEST TEST Completely Dissolved TEST Residue TEST PL Urea *Shake before using! UREA 1...
  • Page 64 PL Urea *Shake before using! UREA 2 TEST Ammonia N°1...
  • Page 65 Ammonia N°1* Ammonia N°1 Ammonia N°2 Ammonia N°2...
  • Page 66 1.5 ppm TEST Ch4n2o If the sample contains free chlorine, a „Dechlor“ tablet has to be added to the vial, before adding PL Urea 1 and PL Urea 2. Ammonia N°1 only dissolves entirely after Ammonia N°2 was added. Ammonia and chloramines will be detected together.
  • Page 67 Hvis prøven indeholder frit klor, skal der tilsættes en »Dechlor«-tablet til hætteglasset, før PL Urea 1 og PL Urea 2 tilsættes. Ammoniak nr. 1 opløses først helt, når ammoniak nr. 2 er tilsat. Ammoniak og kloraminer vil blive detekteret sammen. Det viste resultat viser summen af begge. Prøvens temperatur skal være mellem 20 °C og 30 °C.
  • Page 68 OR-UR (Dilution) | OR-UR (Verdunning) | OR-UR (fortynding) | OR-UR (utspädning) | OR-UR (разбавление) OR = Overrange / UR = Underrange. Test result is outside the range of the method. OR results can be brought into measurement range by dilution. Use syringe to take only 5 ml (or 1 ml) sample water plus 5 ml (9 ml) distilled water.
  • Page 69 Error Codes | Foutcodes | Fejlkoder | Felkoder | Коды ошибок BAT! Change Batteries | Batterijen vervangen | Skift batterier | Byt batterier | Замените батарейки ERR02 (Too dark) Clean Measurement chamber or Dilute Sample | (Te donker) Meetkamer reinigen of monster verdunnen | (For mørkt) Rengør målekammeret eller fortynd prøven | (För mörkt) Rengör mätkammaren eller späd ut provet | (Слишком...
  • Page 70 Troubleshooting | Problemen oplossen | Fejlfinding | Felsökning | Устранение неполадок 1) 01.01.1970: The date on the PoolLab 1.0 is set to 01.01.1970 ® when delivered, after each battery change and after each update. Please reconnect to the LabCOM app so that the smartphone date is adopted again.
  • Page 71 1) 01.01.1970: Datumet på PoolLab 1.0 är inställt på ® 01.01.1970 vid leverans, efter varje batteribyte och efter varje uppdatering. Vänligen återanslut till LabCOM-appen så att datumet på din smartphone ändras igen. 2) Idealiska värden: Kontakta leverantören av din poolkemi för att be om idealvärden för din pool.
  • Page 72 Changing Cuvette/Calibration | Van cuvet wisselen/Kalibratie | Udskiftning af kuvette/kalibrering | Byte av kyvett/kalibrering | Смена кюветы/калибровка 90° 90°...
  • Page 73 Accessories | Accessoires | Tilbehør | Tillbehör | Аксессуары Reagents | Reagentia | Reagenser | Reagenser | Реактивы POL01-Nf 20/20/10/10/10 Phenol Red / DPD N° 1 / DPD N° 3 / –Test / Alkalinity–M Photometer TbsPph50 50 x Phenol Red Photometer TbsPD150 50 x DPD N°...
  • Page 74 App and Software | App en software | App og software | App och programvara | Приложения и программное обеспечение Bluetooth ON Bluetooth OFF www.labcom.cloud Windows /MacOS ® ®...
  • Page 75 Technical data | Technische gegevens | Tekniske data | Tekniska data | Технические данные www.poollab.org MSDS msds.water-id.com Cloud labcom.cloud LED: 530 nm / 570 nm / 620 nm 3 x AAA (1.5 V, Lr03) 5 min. 5 - 45°C IP 68 (1 h / 1.2 m) Developed in Germany Produced in PRC...
  • Page 76 Tolerances | Toleranties | Tolerancer | Toleranser | Допуски Active Oxygen (MPS) | Actieve zuurstof (MPS) | Active Oxygen (MPS) | Aktivsauerstoff (MPS) Aktiv ilt (MPS) | Aktivt syre (MPS) | Oxígeno Activo (MPS) | Oxygène Actif (MPS) Активный кислород (MPS) Ossigeno Attivo (MPS) Range 0.0 - 5.0...
  • Page 77 Bromine | Broom | Brom | Brom | Бром Range 0.0 - 2.5 0.2 ppm (mg/l) 2.5 - 6.5 0.6 ppm (mg/l) 6.5 - 11.0 1.7 ppm (mg/l) 11.0 - 13.5 2.3 ppm (mg/l) 13.5 - 18.0 3.0 ppm (mg/l) Calcium Hardness | Calciumhardheid | Calcium-hårdhed | Kalcium Hårdhet | Твердость...
  • Page 78 Cyanuric Acid | Cyanuurzuur | Cyanursyre | Cyanursyra | Циануровая кислота Range 0 - 15 1 ppm (mg/l) 15 - 50 5 ppm (mg/l) 50 - 120 13 ppm (mg/l) 120 - 160 19 ppm (mg/l) Chlorine Dioxide | Chloordioxide | Klordioxid | Klordioxid | Диоксид...
  • Page 79 Hydrogen Peroxide | Waterstofperoxide | Hydrogenperoxid | Väteperoxid | Перекись водорода – (HR) Range 0 - 50 5 ppm (mg/l) 50 - 110 6 ppm (mg/l) 110 - 170 11 ppm (mg/l) 170 - 200 13 ppm (mg/l) Ozone | Ozon | Ozon | Ozon | Озон Range 0.00 - 1.00 0.07 ppm (mg/l)
  • Page 80 PHMB Range 5 - 60 5 ppm (mg/l) Total Hardness | Totale hardheid | Total hårdhed | Total hårdhet | Общая жесткость Range 0 - 30 3 ppm (mg/l) 30 - 60 5 ppm (mg/l) 60 - 100 10 ppm (mg/l) 100 - 200 17 ppm (mg/l) 200 - 300...
  • Page 81 The manufacturer of this device, Water-i.d. GmbH, ® Daimlerstr. 20, D-76344 Eggenstein will dispose of your PoolLab 1.0 Photometer ® free of charge (not including costs of sending the device to us). Send your Pool- for disposal -freight prepaid - to the address shown above.
  • Page 82 UKCA Complaince Statement UK Conformity Assessed We, Water-i.d. GmbH Germany, hereby certify our responsibility, that the following product: PrimeLab 2.0 Photometer, is tested to and conforms with the essential test suites included in the following standards, which are in force within the UK: Standards Legislation Number EN 61000-3-2: 2014;...
  • Page 83 FCC Part 15 Compliance Statement IC Licence-Exempt RSS Compliance Statement This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
  • Page 84 Changes or modifications not expressly approved by Water-i.d. GmbH could void the user‘s authority to operate the equipment. FCC ID: 2ALRR-POOLLAB10 22610- POOLLAB10 Model: POOLLAB 1.0...
  • Page 85 Certificate Of Compliance We hereby certify that the device PoolLab 1.0 ® With it‘s serial number as stated below, has passed intensive visual and technical checks as part of our QM documentation. We confirm the device got factory-calibrated. Water–i.d. GmbH (Germany) ®...
  • Page 86 Notes...
  • Page 87 Notes...