Install the transmitter in an area where vibration, electromagnetic, and radio frequency interference are minimized or absent. 1.2 Mounting The 1058 supports panel, wall, and pipe mounting. Refer to Mounting configuration for drawings showing each of these mounting configurations. Mounting configuration Figure 1-1: Panel mount, front view www.Emerson.com...
Page 4
Quick Start Guide August 2024 Figure 1-2: Panel mount, side view A. Panel mount gasket B. Panel supplied by customer; maximum thickness: 0.375 in. (9.52 C. Four mounting brackets and screws provided with the instrument www.Emerson.com...
Page 5
August 2024 Quick Start Guide Figure 1-3: Panel mount, bottom view A. Conduit openings www.Emerson.com...
Page 6
Quick Start Guide August 2024 Figure 1-4: Panel cut-out A. Maximum www.Emerson.com...
Page 7
August 2024 Quick Start Guide Figure 1-5: Wall mount, front view A. QTY 4 Ø5/16 wall anchor screws Figure 1-6: Wall mount, side view www.Emerson.com...
Page 8
Quick Start Guide August 2024 Figure 1-7: Wall mount, bottom view A. Front panel B. 2-in. (51 mm) pipe mount bracket C. Two sets U-bolts for 2-in. (51 mm) pipe in kit, PN 23820-00 www.Emerson.com...
PG13.5 cable glands. A ½-in. NPT thread is recommended. To maintain ingress protection, use Type 4X or IP66 rated cable glands for all cables entering the device and block unused openings with Type 4X or IP66 conduit plugs. www.Emerson.com...
Page 10
Digital sensors with M12 connectors Digital sensors with M12 type connectors are connected to the transmitter by plugging the M12 male connector on the sensor cable into the M12 female connector installed in one of the transmitter conduit openings. www.Emerson.com...
1. Wire the relay leads on each of the independent relays to the terminal on the main board using the lead markings (+/positive, -/negative) on the board. Emerson provides male mating connectors with each unit.
Page 12
The device has four alarm relay outputs. To use the relay outputs, wire the relay leads on each of the independent relays to the correct position on the power supply board using the printed lead markings (NO/Normally Open, NC/Normally Closed, or Com/Common) on the board. www.Emerson.com...
The grounding plate is connected to the earth terminal of the 85-265 VAC power supply device version. The green colored screws on the grounding plate are intended for connection to some sensors to minimize radio frequency interference. The green screws are not intended to be used for safety purposes. www.Emerson.com...
Page 14
Quick Start Guide August 2024 Figure 1-11: Power Wiring for 24 VDC Power Supply PN 24365-030 A. To main board (ribbon cable) B. Rosemount 1058 DC power supply board PN 24365-030 www.Emerson.com...
Page 15
August 2024 Quick Start Guide Figure 1-12: Power Wiring for 85-264 VAC Power Supply PN D0000340-01 A. Earth ground B. Neutral C. Line power D. To main board (ribbon cable) E. Rosemount 1058 AC power supply board PN D0000340-01 www.Emerson.com...
Page 16
Once all wiring connections are secured, close the front cover of the device, and install the four cover screws. Then apply power to the device. When the device is powered up for the first time, Quick Start screens appear and guide you through the initial configuration of the device. www.Emerson.com...
Quick Start Guide 3 Product certifications For Rosemount 1058 Dual Channel Transmitter product certifications, see the Rosemount 1058 Dual Channel Transmitter Quick Start Guide. 3.1 European Directive Information A copy of the Declaration of Conformity can be found at the end of the Quick Start Guide.
Page 18
NON-INCEDNIVE Class I, Division 2, Groups ABCD Markings: T4 T -10 °C To +60 °C NIFW Per DWG A1058-1300 Special Conditions for Safe Use (X): 1. Sensors having exposed electrodes in the process will be used in a non-flammable liquid only. www.Emerson.com...
Quick Start Guide August 2024 3.7 China RoHS www.Emerson.com...
Page 21
August 2024 Quick Start Guide www.Emerson.com...
Page 22
Quick Start Guide August 2024 www.Emerson.com...
Page 23
August 2024 Quick Start Guide www.Emerson.com...
Page 24
2024 Emerson. All rights reserved. Emerson Terms and Conditions of Sale are available upon request. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners.
Need help?
Do you have a question about the Rosemount 1058 and is the answer not in the manual?
Questions and answers