ATAGO PAL-DASHI Meister Instruction Manual

ATAGO PAL-DASHI Meister Instruction Manual

Digital hand-held pocket dashi brix-salt mete

Advertisement

Quick Links

R
4943-E01
Instruction Manual
Sample Stage/
Sensor
Digital Hand-held "Pocket" DASHI
Brix-Salt Meter
PAL-DASHI Meister
LCD
Cat. No. 4943
How to Get Extended Warranty
出汁巻 Key
To turn off the display,
1 year
2 years
press and hold for
more than 2 seconds.
The warranty period will be
Good Key
extended from 1 year to 2
years when you register your
information.
QR code
ATAGO Logger NFC can also be
downloaded at the same time.
Battery Compartment
Trouble scanning the code? Access this
link.
https://www.atago.net/ur/index.php?l=en
Main unit...1
m Please remove the tape in the battery compartment before first use.
ATAGO instruments are rigorously inspected to ensure each unit meets the highest standards of quality assurance.
※Recommended on a daily basis.
Salt: Zero-setting
Std display
3.
1.
2.
Clean the sensor
Press the
出汁巻 Key
Measurement : not 0.00%
(with nothing on
the sensor).
Press the 水 Key
Std display
Measurement : 0.00%
Std
- - -
Brix
The instrument is
zero-set correctly.
- - -
Salt
Brix
:
The sum of all the extracts (umami) dissolved in the soup stock.
Salt (Std scale):
Default mode for measurement and checking with reference solution.
Soup stocks with less than 6% Brix can be measured by Std scale.
Salt (USr scale):
Sanility scale to be created by the user.
15 types of user scales can be registered.
Creating user scales allows you to measure Salt concentration in undiluted liquids
such as sauces with a Brix of 6% or higher.
User scales can measure up to 28.50%.
Low Sodium Index:
Ratio of how much salt is contained to the total concentration of all soluble solids in the sample.
Low Sodium Index is calculated by Salt ÷ Brix.
N Recommended that you note the sample and scale number
used to create the user scales.Use the table below if you like.
Preparation Select a Scale Number
Select a scale to be used for the measurement. Std scale or user scale from 1 to 15 selectable.
1.
3.
2.
Press and hold the
Press the Good Key
Select a Scale Number.
Good Key for 5 seconds
to confirm the selection.
Use the 水 Key to
(while the unit is
move the number
powered on).
upward.
Use the 出汁巻 Key
to move the number
downward.
U 2
std
5s
U 1
sEt
SET display
Scale Number display
Parts
Remove the tape from the battery
compartment and close the cover.
mThe cover must be closed properly to
maintain water resistance. Keep the
batteries in during storage.
tape
水 Key
Closed
Lanyard Hole
Contents
Calibration Report...1
AAA batteries...2
Preparation Zero-Setting
Brix: Zero-setting
4.
5.
6.
Apply tap
Press
water.
the 出汁巻
Measurement: not 0.0 Brix
20.4
Key.
-Zero-setting is required.-
0 0 0
Salt
At least 0.6mL
Press the 水 Key.
Flash
Light up
Measurement: 0.0 Brix
The instrument is
20.4
L L L
Brix
zero-set correctly.
Press the 出汁巻
0.0
Salt
Key.
About Scale
 
e.g. Dashi
 
Std
 U 8
 U 1
 U 9
 U 2
 U 10
 U 3
 U 11
 U 4
 U 12
 U 5
 U 13
 U 14
 U 6
 U 7
 U 15
Preparation
The instrument is equipped with two types of measurement modes.Factory default is Mode HH.Select the mode according to your preference.
Mode HH...Allows you to quickly measure hot Dashi. Resolution is 0.1%. 
Mode TT...Allows you to perform measurements with less variation in repeated measurements. Resolution 0.1% or 0.01% can be selected.
1.
2.Press 水 Key to
Press the 出汁巻 and
Good Keys
select the modes.
simultaneously
-H-
r 0.1
-H- : Mode HH
-t- : Mode TT
Open
Brix and Salt measurements simultaneously. Screen transitions differ between Mode HH and Mode TT.
For Mode HH
1.
2.
Clean the
Apply the
sensor.
sample.
At least 0.6mL
Brix: Zero-setting
20.4
0 0 0
Brix
Flash
Light up
For Mode TT
Proceed to
20.4
Brix
0.0
measurement.
1.
2.
Apply the
Clean the
0 0
sensor.
sample.
Salt
0.6mL
以上
3.
End of measurements
The same measurement values
continue for 3 times
for both Brix and Salt.
1.
2.
Wipe off the
Clean with a
sample.
mild soap, and rinse
well with water.
Mild
soap
N Do not use alcohol.
About Measurement Mode - ModeHH and ModeTT -
Preparation Mode Settings
3. Press the 出汁巻
4. -1 For Mode HH
Key to confirm the
selection.
Setting completed.
-t-
4. -2 For Mode TT
r 0.1
-t-
r 0.1
Press 水 Key to select the resolution.
The upper row blinks
twice when Mode TT
-t-
is selected.
r0.01
How To
Measure
3.
Measurements
4.
Low Sodium index
Std display
Std
20.4
Brix
5.3
Brix
- - -
Press the
1.3
- - -
Salt
Salt
Good Key
---Flash
Results
U
display
20.4
Press the
0.0 6
出汁巻 Key
Ratio
U 1
20.4
Brix
- - -
Brix
25.3
- - -
Low Sodium Index
6.0
Salt
Salt
3. Start
measurements
"-" flashes until the measured value stabilizes.
Std
Std
Std
Std
Brix
-
Brix
-
Brix
-
display
-
-
-
Salt
Salt
Salt
Press the
出汁巻 Key
U 1
U 1
U 1
U
Brix
Brix
-
Brix
-
-
display
-
-
-
Salt
Salt
Salt
4.
Low Sodium index
Flashing stops and
end of measurements.
20.4
Brix
5. . 2
or
1.31
Press the
Salt
出汁巻 Key
Results
Low Sodium index
Cleaning
3.
Dry the area with
Wash under running water (<50℃).
tissues thoroughly.
Do not submerge
in water.
Rub strongly with
dry tissue paper
-H-
Blink twice
r 0.1
5.Press 出汁巻 Key.
5.-2
 Setting completed.
 
resolution
r 0.1
: 
0.1%
Blink
 
-t-
resolution
r0.01 : 
0.01%
twice
r0.01
Once the measured value has stabilized,
the measured value will blink.
Continue measuring continuously.
20.4
measured value,
5. . 28
Brix
Flash the
"Brix" and "Salt"
1.31
Salt
NMeasurement ends and the power turns off
automatically 5 minutes after pressing
the 出汁巻 Key.
Press the
Good Key
20.4
0.240
Ratio
Error Messages
・Brix:Sample not enough.
L L L
・Temperature:Below the range.
・Brix:Zero-setting with other than water.
A A A
・Salt:Calibration with other than reference
solution.
H H H
・Above the measurement range.
n n n
・Brix and Salt(USr):Too Bright. Shade the
sample stage with your hand.
o o o
・Low sodium index error.
E r r
・Creating user scales error.
Lo
・The battery is low.

Advertisement

loading
Need help?

Need help?

Do you have a question about the PAL-DASHI Meister and is the answer not in the manual?

Questions and answers

Summary of Contents for ATAGO PAL-DASHI Meister

  • Page 1 Brix and Salt measurements simultaneously. Screen transitions differ between Mode HH and Mode TT. m Please remove the tape in the battery compartment before first use. For Mode HH ATAGO instruments are rigorously inspected to ensure each unit meets the highest standards of quality assurance. Clean the Apply the...
  • Page 2 USB NFC Reader/writer to the PAL unit. When there is any doubt regarding accuracy of measurement results, N The reference solution is available from ATAGO. N Data history can be read out while PAL is powered off. adjust the reference value according to the following procedure.
  • Page 3              ・Before use, carefully read the instruction manual and fully understand the function and operation for each part of the instrument. ・ATAGO is not liable for any loss and damage caused by the measurement and use of this instrument.  ・If this instrument is used to measure highly acidic samples, the sensor section and sample stage may be damaged, resulting in inaccurate measurements.