tntertek
Total Quality. Assured.
SCHEDULE:
TYPE EXAMINATION CERTIFICATE NUMBER ITS18ATEX203388X
requirements of IEC 60079-15. Test procedures are as follows:
Voltage:
500Vrms
Duration:
60s
Permitted leakage current:
5mA
Alternatively, the equipment can be subjected to the following parameters.
Voltage:
600Vrms
Duration:
0.1s
Permitted leakage current:
5mA
The manufacturer is required to record and maintain all results obtained.
14.
Essential Health and Safety Requirements (EHSRs)
The relevant Essential Health and Safety Requirements (EHSRs) affected by this variation have
been identified and assessed in lntertek Report: 103291356DAL-002-EHSR Dated: 6/12/2018.
15.
Drawings and Documents
Title:
Drawing No.:
Rev. Level:
Date:
ATEX Technical File
68445
A
11/12/19
16.
Details of Certificate changes Issue 1
Addition of integral cable version of equipment for Ex ic version
This Certificate is for the exclusive use of lotertek's client and is provided pursuant to the agreement between Intertek and its Client. Intertek's responsibility and liability are limited tothe terms and
conditions of the agreement. Intertek assumes no liability to any party, other than to the Client in accordance with the agreement, for any loss, expense or damage occasioned by the use of this
Certificate. Only the Client is authorized to permit copying or distribution of this Certificate and then only in its entirety. Any use of the Intertek name or one of its marks for the sale or
advertisement of the tested material, product or service must first be approved in writing by Intertek.
Intertek Testing & Certification Limited, Cleeve Road, Leatherhead, Surrey, KT22 iSA
Registered No 3272281 Registered Office: Academy Place, 1.9 Brook Street, Brentwood, Essex, CM14 SNO.
© 2019 INTERTEK
Page 3 of 3
[Document Control No. CT-R-UK-Cert-39291
Need help?
Do you have a question about the IMI SENSORS ICP EX604 Series and is the answer not in the manual?
Questions and answers