Introduction / About This Manual
This manual is designed to assist with routine service procedures for the V.A.C.ULTA™ Therapy Unit. Following
these procedures ensures that the V.A.C.ULTA™ Therapy Unit is properly cleaned, fully functional and ready for
patient use.
These procedures include:
ㆍ
Cleaning and disinfection
ㆍ
Inspection for damaged and / or missing parts
e
Verifying unit function
。
Preparing the unit for transport to patient
V.A.C.ULTA™ Therapy Unit Checklist
Before starting any cleaning or service procedure detailed in this manual, make photocopies of the
V.A.C.ULTA™ Therapy Unit Checklist located in the back of this manual (see page 32). After the inspection and
service are complete, retain the completed form for each unit as a permanent record.
All steps in these procedures must be followed in the order presented to provide proper
functionality and reliability of the V.A.C.ULTA™ Therapy Unit.
Each V.A.C.ULTA™ Therapy Unit must be cleaned, disinfected, inspected and charged
between each patient use. For further questions about the required frequency of service
procedures, contact KCI (see page 31).
Opening the therapy unit to gain access to the internal components may void any
warranty.