CALIBRATION
NOTES
13.
COMPARISON
VOLTAGE
a.
Adjust Ve Cal, R310:
+11V,
40.05%
Set
the
differential
voltmeter
to
read
11 volts.
Connect
the
plug
(+)
lead
to
the
VC OUTPUT
and
the
common
lead
to
the
gnd
post.
Adjust
R310
Ve
Cal
for
11
volts.
b.
Adjust
tracking,
R328:
+1
volt,
t0.05%
Remove
the
differential
voltmeter
plus
(+)
lead
from
the
Ve
OUTPUT.
Set
the
differential
voltmeter
for
a l volt
reading.
Set
the
COMPARI-
SON
VOLTAGE
(Vc)
knob
to
0.
Leave
the
dial
at
10.0.
Reconnect
the
differential
voltmeter
plus
(+)
lead
to
the Ve
OUTPUT
and
adjust
R323
for
1 volt.
@.
Check
accuracy:
+0.1% + (0.05%
of Ve RANGE)
Set
the
COMPARISON
VOLTAGE
(VC)
knob
to 0,
and
the
dial
to
0.00.
Set
the
differential
voltmeter
to read
0 volt.
Using
the
table below,
check
the
accuracy
of
the
COMPARISON
VOLTAGE.
voltmeter
maximum
Ve knob
Vc
dial
reading
deviation
0
0.00
0.0V
-
-005V
1
0.00
1.0V
-006V
2
0.00
2.0V
.007V
3
0.00
3.0V
.008V
4
0.00
4.0V
.009V
5
0.00
5.0V
-O010V
6
0.00
6.0V
.O11V
7
~0.00
7.0V
.O12V
8
0.00
8.0V
.013V
9
0.00
9.0V
.
.O14V
10
~ 0.00
-10.0V
.015V
10
1.00
10.1V
.0151V
10
2.00
10.2V
.0152V
10
3.00
10.3V
.0153V
10
4.00
10.4V
.O154V
10
5.00
10.5V
-O155V
10
6.00
10.6V
.0156V
10
7.00
10.7V
.0157V
10
8.00
10.8V
.0158V
10
9.00
10.9V
—.0159V
10
10.00
11.0V
.016
V
March
1969
W