CONTENTS General operation data Certification of medical devices in compliance with Regulation (EU) 2017/745 CLASS I Introduction Signals and warnings Important notice on central aspirator certified as medical device Installation and starting Routine maintenance Extraordinary maintenance Instructions for adjusting some parameters in the menus of Smart aspirators Instructions for configuration of WIFI (wireless) communication Alarm description Important warnings...
IEC 417-5008 Closed (connection to power supply mains) IEC 417-5007 Manufactured by CATTANI S.p.A. - PARMA - ITALIA Sound pressure level detected according to ISO 3746-1979 (E). Parameters: r= 1.5 - background noise: 34 dB(A) - Bruel & Kjaer type 2232 instrument.
Page 5
IEC 417-5008 Closed (connection to power supply mains) IEC 417-5007 Manufactured by CATTANI S.p.A. - PARMA - ITALIA Sound pressure level detected according to ISO 3746-1979 (E). Parameters: r= 1.5 - background noise: 34 dB(A) - Bruel & Kjaer type 2232 instrument.
Page 6
IEC 417-5008 Closed (connection to power supply mains) IEC 417-5007 Manufactured by CATTANI S.p.A. - PARMA - ITALIA Sound pressure level detected according to ISO 3746-1979 (E). Parameters: r= 1.5 - background noise: 34 dB(A) - Bruel & Kjaer type 2232 instrument.
Page 7
IEC 417-5008 Closed (connection to power supply mains) IEC 417-5007 Manufactured by CATTANI S.p.A. - PARMA - ITALIA Sound pressure level detected according to ISO 3746-1979 (E). Parameters: r= 1.5 - background noise: 34 dB(A) - Bruel & Kjaer type 2232 instrument.
Page 8
IEC 417-5008 Closed (connection to power supply mains) IEC 417-5007 Manufactured by CATTANI S.p.A. - PARMA - ITALIA Sound pressure level detected according to ISO 3746-1979 (E). Parameters: r= 1.5 - background noise: 34 dB(A) - Bruel & Kjaer type 2232 instrument.
Page 9
EN 62353 standard "Medical Electrical Equipment" Smart, Micro Smart Cube, Turbo Smart, Turbo - Periodic inspections and tests to be carried out Smart TS, Turbo Smart Cube, Turbo Smart 2V after repair work on medical electrical equipment Technicians authorised to carry out repair work...
Updated manuals are available on our website www. This presentation illustrates the procedures for installation and starting of Smart aspirators and cattani.it . We recommend you refer to these especially provides information on the hazards and the precautions for the latest safety updates.
IMPORTANT NOTICE ON CENTRAL ASPIRATOR CERTIFIED AS MEDICAL DEVICE IMPORTANT NOTICE ON CENTRAL ASPIRATOR CERTIFIED AS MEDICAL DEVICE A centralized aspirator does not provide a direct medical devices, without a proper distribution the air flow to the dental unit, but to the distribution airflow cannot be guaranteed, thus failing the very net that must ensure the correct distribution to the aim of the certification of the central aspirator as a...
For outdoor installation of Turbo Smart and Micro Smart we can supply a box featuring: double insulating roof, antifreeze and ventilation systems (both fitted with fixed thermostat for automatic temperature control).
Page 13
Use protections and danger signs, in order to prevent accidental contacts and the risk of electric shocks or the event (unlikely but possible) of fire, explosion and contaminating air or liquid leakage. Use only indoor and outdoor boxes designed and produced by the manufacturer of the system.
Page 14
(2) (fig. A and B, page 40). (*) 50 mm diameter for Turbo Smart, Turbo Smart TS, Turbo Smart Cube and Turbo Smart 2V, 30 mm diameter for Micro- Smart and 40 mm diameter for Micro Smart Cube. (**) 50 mm diameter for Turbo...
Page 15
Even when the Smart aspirators are installed on a lower level than the dental units, the suction piping must go down from the surgeries to the level of the aspirator, with a few metres of piping laid horizontally on the same level as the machine and then move up with a flexible tube to the centrifugal separator (fig.
Page 16
41-42-43-44). Turbo Smart, Turbo Smart TS, Turbo must be increased. Turbo Smart, Turbo Smart TS, Smart Cube and Turbo Smart 2V units in parallel Turbo Smart Cube and Turbo Smart 2V are always can offer better performance and energy savings...
Page 17
STARTING, FINAL TEST AND STAFF INSTRUCTION Once the aspirator has been installed and connected, simply press the on/off switch and suction will start when one of the dental units has started working. To check if the Smart aspirator is working correctly, it is advisable to carry out the dynamic tests (shown in fig.
Page 18
“Hydrocyclone ISO 18” amalgam separator for Turbo Smart, Turbo Smart TS, Turbo Smart Cube and Turbo Smart 2V and with "Hydrocyclone ISO 6" for Micro Smart and Micro Smart Cube. The amalgam separator comes with a separate instruction and maintenance manual.
ROUTINE MAINTENANCE ROUTINE MAINTENANCE Routine maintenance must be carried out by suitably instructed surgery staff. • We recommend you pay special attention to all the danger signs and wear goggles, gloves and disposable overalls for protection. EVERY DAY • Check the display for alarms and in case of danger •...
EXTRAORDINARY MAINTENANCE EXTRAORDINARY MAINTENANCE Extraordinary maintenance procedures must be carried out by a qualified technician using original spare parts: • We recommend you pay special attention to all the danger signs and wear goggles, gloves and disposable overalls for protection. •...
Page 21
RECOMMENDED CHECKS EVERY 18-24 MONTHS • Check the efficiency of the centrifugal separator (11) and of the recirculation valve (14). RECOMMENDED CHECKS EVERY 10,000 HOURS • Rubber parts: O rings, gaskets or seals must be replaced every time they are removed. •...
Page 22
OF THE PARAMETERS IN THE MENUS OF SMART ASPIRATORS MAIN MENUS When the control unit is switched on, the graphic display will show the Cattani logo for 10 seconds, then the main menu will appear. MAIN MENU “A1” This menu shows parameters such as: operation frequency, suction activation time, temperature, presence/absence of amalgam container and system software revision.
Page 23
MENU DOWNLOAD MENU “A7” You can check if there are any downloads available to update the firmware. CONTROL MENU “A2” Counters – Odom . A2 Power Cycles 00000 Number of times switched on with master switch Uptime [h] 00000 Total hours switched ON (with motors not running) Work Cycles 00000 Number of times activated by suction control Work Time [h] 00000 Actual hours of operation (motors running)
Page 24
To access this menu, enter the password “0000456000”. You can change switch-off delay time and other technical parameters. FACTORY SETTINGS RESTRICTED ACCESS This menu is reserved for exclusive use by Cattani. WI FI SETTINGS You can access this menu without a password to change WiFi parameter settings.
Page 25
Password and the cursor will disappear. Press to go back to the “Cattani S.p.A.” menu. You will now be able to change the parameters of the “User Parameters” menu. Repeat the same procedure to enter the password 0000456000 in order to make changes in the “System Parameters Setup”...
Page 26
Centrifugal separator current System temperature Maximum recorded temperature (to reset use code 19404 in menu drive commands) Maximum recorded temperature Bus voltage Maximum recorded bus voltage Ripple bus voltage cod: ed. 04-2024...
Page 27
0 or 1 or 2 or 3 or 4 or 5. Press Enter to confirm the change. Generated code - This code is generated by Cattani. Each appliance has a unique code. Activation code - If used cod: ed. 04-2024...
Page 28
“SYSTEM PARAMETERS SETUP” MENU To access this menu, enter the Access Password 0000456000 (see instructions on page 24). You can now change some of the settings. To change the parameters in this menu: Use the arrows to scroll the menus and highlight the parameters you want to change.
Page 29
Blower current limit for medium vacuum level (only Micro Smart and Micro Smart Cube) Surgery vacuum level (only Micro Smart and Micro Smart Cube) Maximum blower frequency for surgery vacuum level (only Micro Smart and Micro Smart Cube) Blower current limit for surgery vacuum level (only Micro Smart and Micro Smart Cube) Switch off delay (max.
Page 30
Options used Enables or disables reading of amalgam sensor. 0- without amalgam separator 1- with amalgam separator Inverter controls. Reserved for factory settings cod: ed. 04-2024...
Page 31
Available in models Notes Turbo Smart from serial number WP5xxxxx (produced from October (new keypad with external, 2015) removable display) Turbo Smart TS Micro Smart (produced from October from serial number WM3xxxxx 2015) Turbo Smart 2V (produced from September 2017) Turbo Smart Cube cod: ed.
Page 32
Available in models Note Micro Smart Cube Turbo Smart, Turbo Smart TS, Turbo Smart 2V, Turbo Smart Cube, Micro Smart and Micro Smart Cube can be controlled by a professional using SmartApp. SETTING UP THE CONNECTION AND COMMUNICATION IN THE ASPIRATOR When power is switched on, the display will show this picture.
Page 33
Once completed, it will show all the WI-FI networks available in the area. Select the network using the UP and DOWN network and then confirm. Once you have selected the WI-FI network, the display will go back to the main menu A1 in a couple of seconds. Now press enter.
Page 34
In Menu M5-S5, you can enter the password for the WI-FI router, using the arrows. Using the arrows in the M5 S6 Menu you can enter the final part of the password if it is too long. Check that LCD Options is set to 1. If you want the DHCP to be fixed set to 8.
ALARM DESCRIPTION Alarm codes Description Solution Microcontroller memory alarm Contact the technician Microcontroller memory alarm Contact the technician Find out where the short circuit is Short circuit due to one of two motors located and eliminate it Contact the technician (board probably Short circuit before motor control damaged) Condensers not charged...
• The warranty and manufacturer's liability shall • Updated manuals are available on our website be invalidated if the appliances are treated with www.cattani.it. We recommend you refer to these unsuitable products or products other than those especially for the latest safety updates.
WASTE DISPOSAL INFORMATION FOR PROFESSIONAL USERS • Pursuant to art. 13 Legislative Decree no.151 25 of negative effects on the environment and on July, "Implementation of directive 2011/65 EU ROHS human health and promotes reuse and/or recycling and 2003/108/CE, concerning reduction of the use of the materials with which the appliance is made.
Page 38
ELECTROMAGNETIC COMPATIBILITY CONFORMITY LEVELS PER STANDARD EN 60601-1-2:2015 ELECTROMAGNETIC COMPATIBILITY CONFORMITY LEVELS PER STANDARD EN 60601-1-2:2015 Warnings: Although compliant with standard EN 60601-1-2, the medical device can interfere with other nearby equi- pment. The device must never be used in the proximity or be stacked on top of other equipment. Install the device away from other equipment emitting high frequency (short waves, microwaves, electro surgery unit, cellular telephones).
Page 39
ELECTROMAGNETIC COMPATIBILITY CONFORMITY LEVELS PER STANDARD EN 60601-1-2:2015 cod: ed. 04-2024...
Page 40
INSTALLATION LAYOUT INSTALLATION ON SAME LEVEL Fig. A Fig. A to aspirator all’aspiratore INSTALLATION ON A LOWER LEVEL Fig. B Fig. B to aspirator all’aspiratore Min 1000 mm Min 1000 mm cod: ed. 04-2024...
Page 46
DIAGNOSTIC TESTS ON TURBO SMART, TURBO SMART TS, TURBO SMART CUBE, TURBO SMART 2V, MICRO SMART AND MICRO SMART CUBE To check that the Smart aspirator is working in operation and with open suction inlet, i.e. not correctly, you can carry out a few dynamic tests, as connected to piping.
Page 47
I Centrifuge Frequency Head reached I Total Model separator Condition reached [Hz] [mbar] frequency [A] Turbo Smart Suction inlet Cube 80 - 95 190 - 210 5,2 - 5,8 closed Version A Turbo Smart Suction inlet Cube 85 -100 190 - 210...
Page 64
We reduce costs: less maintenance and lower energy costs mean that we are always the most economical on a cost-benefit analysis. We reduce environmental impact: we save 50% on raw materials, and allow you to save between 30% and 50% on electrical consumption. Via Natta 6/A 43122 Parma – Italy T +39 0521 607 604 info@cattani.it WWW.CATTANI.IT...