Failure to properly clean, disinfect or maintain the system as per the manufacturer’s recommendations may affect the WARRANTY. Cattani have products available that are fully compatible with the system’s components and deliver the ultimate care. Care Product details are listed on the inside back cover of this User Maintenance Guide.
Page 4
SMART User Maintenance Guide Daily Maintenance Fill the Pulse Cleaner with 2.5 litres Each line on the inside of the Pulse of warm water. Cleaner represents 1 litre, so fill it up to halfway between the second and third line. Slightly unscrew the red cap (i.e. 1 - 2 Place the Puli-Jet Gentle on a level and steady surface.
Page 5
If too much solution has entered the Unscrew and remove the red cap. dispensing chamber, it can be poured Pour the solution from the dispensing back into the main bottle by tipping cup into the full Pulse Cleaner. the bottle towards the small tube on Take care to not allow more solution the side.
Page 6
SMART User Maintenance Guide Daily Maintenance continued • Wipe the outside of the filter basket FILTER/SOLIDS COLLECTOR ON and the inside of the filter housing DENTAL CHAIR OR REAR DELIVERY SYSTEMS with a paper towel, especially around the rubber o-ring seal, to clean away This process should be completed each day, after the disinfection process.
Weekly Maintenance • Apply a few drops of Lubri-Jet FILTER/SOLIDS COLLECTOR ON CATTANI SUCTION UNIT lubricant to the rubber o-ring, work it around the entire o-ring and then • This process should be completed reinstall the filter into its housing. once per week at the end of the working day, after the disinfection •...
C-211974 or delays in daily practice and loss of Suction inlet filter o-ring (Green Viton) income due to stopping for unexpected (Turbo SMART & Turbo SMART Cube) repairs. M-040727 Routine servicing by a technician Lubri-Jet Silicone Lubricant Drops...
Page 10
Whilst this process can be performed by clinic staff, it is highly recommended Cattani for further advice. to employ your equipment service In this case, the rib in the back of the technician.
Page 11
Micro SMART / Cube 1. Turn the unit off unit with the switch on the control panel. 2. Turn the handle underneath the amalgam container clockwise until the container reaches the base. 3. Gently slide container out of the front of the holder.
Page 12
SMART User Maintenance Guide Turbo SMART / Cube 1. Turn the unit off with the switch on the control panel. 2. Open the drain tap (the tap is open when the lever is in line with the tube) and loosen the breather valve (do not remove).
Page 13
Maxi SMART 1. Turn the unit off unit with the switch on the control panel. 2. Open the drain tap (the tap is open when the lever is in line with the tube) and loosen the breather valve (do not remove). Wait for 15 seconds. 3.
Page 14
EPA certified apparatus. Please don’t place this item in the rubbish bin and/or down the sink. CMA Ecocycle Aust 1300 32 62 92 – your amalgam recycling specialist. 0800 732 922 Contact your equipment service technician or Cattani if you require any further information.