Repair

13.1

General information

13.1.1
Repair concept
Under the Endress+Hauser repair concept, devices have a modular design and repairs are
carried out by Endress+Hauser Service or by properly trained customers.
Spare parts are grouped into logical kits with the associated replacement instructions.
For more information on service and spare parts contact Endress+Hauser Service.
13.1.2
Repair of Ex-certified devices
WARNING
L
Incorrect repair can compromise electrical safety!
Explosion Hazard!
Repairs to Ex-certified devices must be carried out by Endress+Hauser Service or by
specialist personnel according to national regulations.
Relevant standards and national regulations on hazardous areas, safety instructions
and certificates must be observed.
Use only original Endress+Hauser spare parts.
Please note the device designation on the nameplate. Only identical parts may be used
as replacements.
Carry out repairs according to the instructions.
Only the Endress+Hauser service team is permitted to modify a certified device and
convert it to another certified version.
13.2

Spare parts

• Some replaceable device components are identified by a spare part nameplate. This
contains information about the spare part.
• All the spare parts for the measuring device, along with the order code, are listed in the
Device Viewer (www.endress.com/deviceviewer) and can be ordered. If available, users
can also download the associated Installation Instructions.
Device serial number:
• Located on the device and spare part nameplate.
• Can be read out via the device software.
13.3

Replacement

CAUTION
L
Data upload/download is not permitted if the device is used for safety-related
applications.
After an entire device or an electronics module has been replaced, the parameters can
be downloaded to the device again via the communication interface. For this, the data
must have been uploaded to the PC beforehand using the "FieldCare/DeviceCare"
software.