6.2.2 Disinfection..........................20 6.2.3 Visual Guide for Cleaning and Disinfection ................21 7 Operation of C.MOR HD and C.MOR HD³ ....................22 Symbols for Control Elements ......................22 Switch the CCU on and off ........................23 Configure Setup Menu......................... 23 7.3.1...
Page 3
Haag-Streit AG Table of Contents General ..............................35 Maintenance Intervals ......................... 35 10 Disposable Materials and Spare Parts ...................... 36 11 Technical Data ............................. 37 12 Disposal ............................... 39 13 Appendix ..............................40 13.1 Specifications on Electromagnetic Compatibility ................. 40 Glossary ...............................
You can obtain these from the respective components' type plates. You will find the contact details for your local sales representative in the notes pages of these in- structions. A current list of local sales representatives worldwide can be found on the Haag-Streit website (https://www.haag-streit.com/haag-streit-group/contact/haag-streit-distributors/distributors).
Page 5
Haag-Streit AG Manufacturer’s Details and Contact Information | 1 Item Name 12 Product Name 1500 4000089 02010_web...
2 Information for Safety Residual Risks The equipment from Haag-Streit has been designed and produced according to state of the art standards and best practices at the time of manufacture. However, some residual risks remain and are indicated on the equipment and in the form of safety messages (WARNING, CAUTION, NO- TICE) in this document.
Page 7
Regularly check the functionality and quality of the magnified vision. c) Do not use the system in case of any obvious defects, error states or functional restrictions. Call the Haag-Streit Service. CAUTION Risk of unexpected moving parts! Parts that move unexpectedly can cause injuries.
3.2 Important Notes Obligatory Reporting of Serious Incidents Any serious incident occurring in relation to the product must be reported to Haag-Streit and the competent authority of the member state in which the user and/or the patient is established. First Installation and Commissioning The first installation and commissioning of the medical device must be done in accordance with the instructions in the service manual for the medical device.
Page 9
Modifications and maintenance must only be performed by persons explicitly authorized by Haag- Streit. For maintenance only original parts from Haag-Streit as well as approved parts from third- party suppliers have to be installed. After maintenance or technical modifications, the device must be readjusted following the service manual of the surgical microscope system.
4 | Overview Haag-Streit AG 4 Overview 4.1 C.MOR HD and C.MOR HD³ 4.1.1 Component Overview Item Name Note Camera Control Unit (CCU) For setting up and operating the camera. Hollow Connector 5.5/2.1 For connecting the power supply unit to the CCU.
Haag-Streit AG Overview | 4 4.1.2 Camera Head Illustration 1: C.MOR HD Illustration 2: C.MOR HD³ Item Name Note Type Plate on Top of the Reference number (REF), serial number (SN), Health In- Camera Head dustry Bar Code (HIBC) of the camera head...
4 | Overview Haag-Streit AG 4.1.3 CCU Front View Item Name Note Control Elements For setting the camera functions, see Symbols for Control Elements [} 22]. Menu Navigation Controls Buttons for selecting menu items and determining settings, see Symbols for Control Elements [} 22].
Haag-Streit AG Overview | 4 4.1.4 C.MOR HD CCU Rear View Item Name Note Output for connecting a VBS compatible monitor via BNC connector. Transmits the image in analog SD quality. Output for connecting an S-Video compatible monitor via Hosiden connector. Transmits the image in analog SD qual- ity.
4 | Overview Haag-Streit AG 4.1.5 C.MOR HD³ CCU Rear View Item Name Note Output for connecting a VBS compatible monitor via BNC connector. Transmits the image in analog SD quality. Output for connecting an S-Video compatible monitor via Hosiden connector. Transmits the image in analog SD qual- ity.
Haag-Streit AG Installation to the Surgical Microscope | 5 5 Installation to the Surgical Microscope WARNING Risk that accessories fall into the surgical area! Accessories that can fall into the surgical area can injure the patient. a) Ensure careful attachment of accessories.
5 | Installation to the Surgical Microscope Haag-Streit AG 4. Connect the female plug of the camera cable to the CCU connection of the camera head. Screw the connector tight. 5. Connect the male plug of the camera cable to the “Camera” input of the CCU. Screw the con- nector tight.
Haag-Streit AG Installation to the Surgical Microscope | 5 3. Connect the hollow plug of the power supply unit to the “Power” connector of the CCU. 4. Insert the mains plug of the power supply unit into the mains power socket.
6 | Before each Operation Haag-Streit AG 6 Before each Operation 6.1 Initial Preparation 1. Check that all electrical and mechanical connections are firmly seated and free of defects. 2. Switch off the CCU with the power button. 3. Pull the mains plug of the power supply unit out of the mains power socket.
Page 19
Haag-Streit AG Before each Operation | 6 Manual cleaning Front lens Materials required – Disposable gloves – Damp cloths – Dry cotton or microfiber cloth – Cleaning agent (glutaraldehyde-based or alcohol-based) – Cleaning agent (pH 6 − 10) Process –...
6 | Before each Operation Haag-Streit AG 6.2.2 Disinfection Parts that are intended to be used/touched by the user are colored in black in the "Visual Guide for Cleaning and Disinfection", such as handles, push bars, knobs, oculars and framing of monitors and displays.
Haag-Streit AG Before each Operation | 6 6.2.3 Visual Guide for Cleaning and Disinfection Color Note Green Parts that are intended for cleaning and disinfection. Black Parts that are intended to be touched by the user. They must be cleaned and disinfected before and after each surgical intervention.
7 | Operation of C.MOR HD and C.MOR HD³ Haag-Streit AG 7 Operation of C.MOR HD and C.MOR HD³ NOTICE Incorrect camera settings may result in poor image quality or no image display. a) After adjusting the camera settings, make sure that the image reproduction on HD MIOS or the external monitor works.
Haag-Streit AG Operation of C.MOR HD and C.MOR HD³ | 7 Button Function Create and Save Still Images Open and Select Stored Scene Files. 7.2 Switch the CCU on and off The camera head is automatically switched on or off with the surgical microscope system.
7 | Operation of C.MOR HD and C.MOR HD³ Haag-Streit AG 7.3.1 Menu Settings Menu Function Default Note value Video Adjust Pedestal Adjusts the black level. Increasing the value brightens black sections. Normally not used. Video Level Adjusts the camera image signal level. In- creasing the value produces a brighter im- age.
Page 25
Haag-Streit AG Operation of C.MOR HD and C.MOR HD³ | 7 Menu Function Default Note value AE Mode Shutter Control AUTO Sets the shutter speed mode to automatic or manual. Manual Shutter Adjusts the shutter speed, if Shutter Control is set to FIX. Increasing the value reduces the brightness of the image.
Page 26
7 | Operation of C.MOR HD and C.MOR HD³ Haag-Streit AG Menu Function Default Note value Video setting Format 50 Hz Sets the type of the VBS output signal. SDI Setting* 1080i Sets the type of the HDI SDI output signal.
Haag-Streit AG Operation of C.MOR HD and C.MOR HD³ | 7 Menu Function Default Note value Foot Switch Foot Switch NONE Carries out different operations for foot switches in use. It is possible to use up to 4 foot switches. White Balance WB Mode Sets the white balance mode.
7 | Operation of C.MOR HD and C.MOR HD³ Haag-Streit AG 7.3.4 Adjust Video Frequency Using the Video Setting submenu, the output signals for VBS, DVI and HD SDI can be adjus- ted. For the Y/C signal, no manual setting is possible.
Haag-Streit AG Operation of C.MOR HD and C.MOR HD³ | 7 7.3.5 Choose White Balance Mode In case of incorrect color reproduction, a white balance can be perfomed. The white balance func- tion can be set to automatic or manual. With manual white balance, it is possible to adjust the val- ues for the red and blue gain.
7 | Operation of C.MOR HD and C.MOR HD³ Haag-Streit AG 7.3.7 Increase Image Details It is recommended to activate the detail enhancement of the camera image, so that the image will be clearer and easier to view. To change settings in the setup menu, the menu lock must be deactivated: 1.
Haag-Streit AG Operation of C.MOR HD and C.MOR HD³ | 7 Store a scene file 1. Make sure that all settings that are to be saved in this scene file are correctly adjusted. 2. Select the menu item Scene File.
Haag-Streit AG 7.5 Adjust Light Sensitivity (GAIN) C.MOR HD and C.MOR HD³ can automatically adjust the brightness of the camera image via gain function. The gain of a camera provides a light amplification. Despite changing lighting conditions, the camera image is displayed with almost constant brightness.
9. To activate the menu lock, set the value of the menu item Menu Lock to ON. 7.7 Adjust Image and Monitor Settings (BARS) C.MOR HD and C.MOR HD³ can output a color bar test pattern, so that image and monitor settings (e.g., contrast, brightness) can be adjusted easier.
The error may be due to a systematic fault in the medical device and must be attended by Haag-Streit. Please contact your local representative or Haag-Streit, stating the product name, type and serial number together with a description of the error and the circumstances under which the error oc- curs.
9.1 General Perform maintenance in accordance with the instructions for use of the surgical microscope sys- tem. We recommend to have the surgical microscope system regularly serviced by your Haag-Streit na- tional representative. 9.2 Maintenance Intervals Maintenance tasks which are not described in these instructions must be conducted by authorized service technicians.
Only use approved accessories and spare parts. b) Do not exceed the maximum load capacity of the carrier unit. c) Do not use additional multiple sockets or extension cables for power connection of the sur- gical microscope system. C.MOR HD Name Power pack 4401129...
EN 60 601-1 Appliance class I equipment IEC 60 601-1-2 UMDNS 12-539 GMDN 12539 Degree of protection by enclosure (IP-Code) IP20 Technical Data C.MOR HD C.MOR HD³ 4000011 4000012 Mechanical data Dimensions CCU Height [mm] Width [mm] Depth [mm] Weight [kg]...
Page 38
11 | Technical Data Haag-Streit AG Technical Data C.MOR HD C.MOR HD³ Type ⅓″, 1-Chip, CMOS ⅓″, 3-Chip, CMOS Effective pixels 1920 x 1080 1920 x 1080 Scan method 1920 x 1080 50/59.94 1920 x 1080 50/59.94 progressive progressive Height – width - ratio 16:9 16:9 Horizontal resolution...
Directive on Waste of Electrical and Electronic Equipment and may not be disposed together with waste from private households. The manufacturer shall take the surgical microscope system and its components back for proper recycling or disposal. Please contact your local Haag-Streit sales representative. 1500 4000089 02010_web...
13 | Appendix Haag-Streit AG 13 Appendix 13.1 Specifications on Electromagnetic Compatibility Phenomenon Test standard Level Emissions CISPR 11 Class A Harmonic currents IEC 61000-3-2 Class A, stationary Voltage fluctuations and flicker IEC 61000-3-3 Discharge of static electricity IEC 61000-4-2 ±8 kV contact...
Haag-Streit AG Glossary Glossary HIBC AE is the abbreviation for Automatic Exposure. HIBC is the abbreviation of the term "Health In- This function automatically controls the exposure dustry Bar Code". It was introduced to label of the camera lens via the aperture value and medical devices and make them forgery-proof the shutter speed of the lens.
Page 42
Glossary Haag-Streit AG VBS ist the abbreviation for Video Blanking Sync. The VBS signal is a video signal for black and white picture transmission, which is com- posed of the video/picture signal (V), the blank- ing signal (B) and the synchronization signal (S).