Advanced Bionics CI-5293 Instructions For Use Manual page 57

Table of Contents

Advertisement

i
ieC 60601
mmuNity
t
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Note 1: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
C
omPliaNCe
l
l
ª
eSt
eVel
eVel
10 V/m
e
e
leCtromagNetiC
NViroNmeNt
Portable and mobile RF communications equipment should
be used no closer to any part of the Naída CI, including
cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.Recommended separation distance:
d = 1.2√P < 800 MHz
d = 2.3√P ≥ 800 MHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic site
survey,
should be less than the compliance level in each
b
frequency range.
Interference may occur in the vicinity of
c
equipment marked with the following symbol:
- g
uidaNCe
55