Racal Instruments 9010 Service Manual page 20

Decade divider

Advertisement

TEST EQUIPMENT REQUIRED
4.3
TABLE 4.2
TEST INSTRUMENTS
TEST EQUIPMENT
SPECIFICATION
SUITABLE MODEL
20 000
Q/V
olt AC/DC
Multimeter
Voltage Ranges 0-2!5OV
20k
AVO Model 8
f. s. d.
Frequency Meter (Counter)
10 Hz - 60 MHz. 1V Sensi-
Racal 9037, 9021, 9022,
tivity. Resolution 1 Hz.
9023 etc.
Signal Generator
lMHz to > 600 MHz.
lmV to> lV into 500.
o
sci Hoscope
Bandwidth d.c.
to
>50MHz.
Tektronix 453 or
454-
Sensivitity 50 mV/cm.
Electroni c Voltmeter
50Q input impedance.
Racal-Airmec Type 301A
10 Hz - 60 MHz •
...
Coaxial Leads (500)
{Quantity twa with BNC
connector at each end.
3 ft. (1 metre) long
Quanti ty one with BNC-
to-open end.
BNC IT I Piece
Quantity one, with 500
termi nati ng pad.
POWER SUPPLY CHECK
4.4
(l)
Remove the instrument case (para. 4.2) and check fuses (Table 4.1).
(2)
The instrument should be prepared for use as instructed in page 2-1.
NOTE: If an untested power supply unit is being fitted it is recommended that the
Amplifier Divider and the Divider Buffer Assemblies be removed from the
instrument until the power supply output voltages have been initially
adjusted.
(3)
Cheek that the mains supply voltage is satisfactory and that the voltage
Selector (rear panel) is correctly set.
(4)
Set the POWER switch to ON and note that the amber ponel lamp
i Ilumi nates.
901 0-9013
4 - 2