CHAPTER 4
========
MAl NTE NANCE
==============
INTRODUCTION
4. 1
This Chapter contains functional tests and adjustment procedures which will
assist in maintaining satisfactory performance and serve as an aid to general
faul t diagnos is.
NOTE: Calibration of the Amplifier Divider Assembly 19-0630 and the Attenuator
Assembly 11-0683 (901 0/9011) requires specialised equipment. If these items
are not functioning correctly it is recommended that they be despatched to
Racal Instruments Limited, or to an authorized repair agency for recalibration.
REMOVAL OF CASE
CAUTION:
4.2
(1)
(2)
(3)
iV\ains voltages are exposed when the case is removed.
Switch off the a.c. supply at the bench outlet and disconnect power lead
from the instrument rear panel socket.
Place the instrument face downwards on the bench and slacken off the
four pozidrive screws at the rear (one in each of the four plastic retainers).
Carefully lift the case off the instrument.
TABLE 4.1
FUSELlNK DATA
Ref.
Function
Rating
Part Number
FS1
+220V d.c.
Not used
FS2
+12V d.c.
750mA
23-0005
FS3
+5V d.c.
1.5A
23-0007
See Note Below
FS4
-5 AV d.c.
500mA
23-0004
FSs
-12V d.c.
500mA
23-0004
r
40V
250 mA (anti-surge)
23-0031}
FS6
Mains 110y
500
mA
(anti-surge) 23-0022
Rear Panel
NOTE: If it is required
to
check fuse-links FS2 to FS5 the case must be removed.
F1 (Remote Control Assy)
100mA
23-0001
9010-9013
4 - 1