M967905 - Manual Angio imaging with fixation set
Imaging
Lode BV
Zernikepark 16
NL 9747 AN Groningen
The Netherlands
t: +31 50 571 28 11
f: +31 50 571 67 46
@: ask@lode.nl
www.lode.nl
M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24

Advertisement

Table of Contents
loading

Summary of Contents for Lode M967905

  • Page 1 M967905 - Manual Angio imaging with fixation set Imaging Lode BV Zernikepark 16 NL 9747 AN Groningen The Netherlands t: +31 50 571 28 11 f: +31 50 571 67 46 @: ask@lode.nl www.lode.nl M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24...
  • Page 2: Table Of Contents

    Table of contents Table of contents 1 - Intended Use 1 - Intended Use 2 - Precautions 2 - Precautions 2.1 - Warnings for using the optional wheeled frame with the Angio ergometer 3 - Contra Indication 3 - Contra Indication 3.1 - Free area 4 - Validity 4 - Validity...
  • Page 3: Intended Use

    If, at any time during exercise, the test subject feels faint, dizzy, or experiences pain, stop the test and he or she should be consulted by the physician. M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 3/21...
  • Page 4 Tighten / replace any loose or worn components prior to use. Do not place any loose parts or objects on the device before and during use. M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 4/21...
  • Page 5 The installation should only be done by authorized persons who follow the installation instructions. Do not spill any liquid over the device. M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 5/21...
  • Page 6 The operator should have knowledge about the Warnings and Cautions before using the device. The (support of the) Control Unit can’t be used as handlebar or support. M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 6/21...
  • Page 7: Warnings For Using The Optional Wheeled Frame With The Angio Ergometer

    Be aware of the stability when lifting the Angio on top of the wheeled frame. Before installing the Angio on the wheeled frame the wheels should be blocked. M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 7/21...
  • Page 8: Contra Indication

    The free area must also include the area for emergency dismount. Where equipment is positioned adjacent to each other the value of the free area may be shared. M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 8/21...
  • Page 9: Validity

    4 - Validity 4 - Validity This manual covers all Angio versions starting with the following serial number: 9679xx Angio imaging: S/N 20170001 - Date of issue: 2017-01-31 M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 9/21...
  • Page 10: Description Of Parts

    (rpm). Note: be aware of hands near the display-arm! Tighten the knobs of the display-arm to prevent dropping. M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 10/21...
  • Page 11: Description Of Parts Angio With Fixation Set

    4.3 - Angio - description of parts bottom side 4.3 - Angio - description of parts bottom side 2 - cable connections 6 - Identification plate M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 11/21...
  • Page 12: Installation

    Note. There is place for the thick AC cord in the middle and 2 other thinner cables on the sides. 4. Put the ergometer back in upright position. M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 12/21...
  • Page 13: Angio With Fixation Set

    1. Connect the device with the supplied cable to the external device. You can control the device with e.g. an exercise ECG device or Metabolic Cart. 2. Download the interfacing instructions from the internet at www.lode.nl after you have registered your device.
  • Page 14: External Control With Lem Or Lcrm

    7.2 - External control with LEM or LCRM 1. If your device has an optional RS232 or USB port, connect it directly to the PC on which the Lode software is installed. If your device has a RJ45 connector plug connect it with an UTP cable to the special interface connector that was delivered with your device and to a USB port in your PC.
  • Page 15: Maintenance

    (grounded) wall socket. External Input External Control Connector Class I MEDICAL EQUIPMENT Type B Electrical Safety IEC 60601-1 General warning 9 - Maintenance 9 - Maintenance M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 15/21...
  • Page 16: Cleaning

    Lode cannot advise a specific cleaning agent, since local recipes may differ. The operator should clean the device (handgrips, cushions) after each exercise test. M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 16/21...
  • Page 17: Calibration

    11 - Calibration 11 - Calibration The ergometer should be calibrated once a year. In case any damage is observed of the ergometer Lode B.V. or his representative should be informed in order to execute the necessary repair(s). Service of the ergometer is restricted to factory-trained personnel only.
  • Page 18: Accessories

    For large feet sizes Angio ergometer application cycling Wheeled frame For different leg/arm length Pedal shoes for foot fixation Pedal shoes for children Extra large Pedal shoes M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 18/21...
  • Page 19 USB to Serial converter Usability pack Radiology Partnumber: 226012 Partnumber: 226012 Partnumber: 967830 Partnumber: 967830 Easy connection Programming functions USB to Serial converter Usability pack Radiology M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 19/21...
  • Page 20: Specifications

    V AC 100 - 240 V Phases Frequency 50/60 Hz Power consumption 160 W Power cord IEC 60320 C13 with CEE 7/7 plug Power cord NEMA M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 20/21...
  • Page 21 Minimum air pressure storage & transport 50 kPa Max. humidity storage & transport 95 % Min. humidity storage & transport 10 % Order info Partnumber 967905 M967905 - MANUAL ANGIO IMAGING WITH FIXATION SET / VERSION: 1.0 - 2017-02-24 21/21...