Lode BWSS Manual

Lode BWSS Manual

Body weight support system

Advertisement

Quick Links

Manual - Body Weight Support System
Rehabilitation
Lode BV
Zernikepark 16
NL 9747 AN
Groningen
The Netherlands
t: +31 50 571 28 11
f: +31 50 571 67 46
@: ask@lode.nl
www.lode.nl
M932822 / VERSION: 1.0 - 2020-07-23

Advertisement

Table of Contents
loading

Summary of Contents for Lode BWSS

  • Page 1 Manual - Body Weight Support System Rehabilitation Lode BV Zernikepark 16 NL 9747 AN Groningen The Netherlands t: +31 50 571 28 11 f: +31 50 571 67 46 @: ask@lode.nl www.lode.nl M932822 / VERSION: 1.0 - 2020-07-23...
  • Page 2: Table Of Contents

    9 - Problem solving 9 - Problem solving 9.1 - The BWSS does not work. The display on the control unit remains blank. 9.2 - The lift mechanism does not work. The display on the control unit is on. 10 - Maintenance...
  • Page 3 13 - Science 13 - Science 14 - Specifications 14 - Specifications...
  • Page 4: Intended Use

    3 - Maintenance The BWSS should be calibrated once a year. In case any damage is observed of the BWSS Lode B.V. or his representative should be informed in order to execute the necessary repair(s). Service of the BWSS is restricted to factory-trained personnel only.
  • Page 5: Precautions

    4 - Precautions Read this manual before using the device and follow it carefully. The test subject should not wear loose or dangling clothing while using the device. Do not unweight until you are sure the vest is properly secured. Set up and operate the device on a solid level surface.
  • Page 6 Before each use, inspect the vest for damage. Do not use if the integrity is questionable. Do not release the test person when the system is under tension, unless in an emergency situation. Not suitable for use in the presence of flammable anaesthetics.
  • Page 7: Introduction

    Optional side chairs for the therapist are available which can be mounted on the side bars. The BWSS is easy to move and can be used in a variety of operating environments. The design is user friendly, for both the test subject and the operator of the system.
  • Page 8: Description Of The Body Weight Support System

    7 - Description of the Body Weight Support System 7.1 - Parts The BWSS is a solid mobile device that can be used to reduce the body weight acting on a test persons lower limbs in a variety of operating environments.
  • Page 9: Control Unit

    7.2 - Control unit The “Control unit” (part 5) consists of the following elements: Control unit BWSS 1. Support weight indication display (Kg) 2. Up button. Increase support weight / Lift a person up 3. Down button. Decrease support weight / Put a person on his feet 4.
  • Page 10: Power Supply Panel

    2. Mains connector 3. Fuse box (two fuses inside) 4. Protective Earth connector 7.4 - Adjustments The BWSS can be fitted with optional side handle bars and optional therapist seats. The side handle bars are adjustable for test subjects of all sizes and ages.
  • Page 11: Using The Bwss

    8 - Using the BWSS This chapter describes how to use the BWSS. Execute the following steps for performing an bodyweight support test with the BWSS after the BWSS is installed. 8.1 - Adjust the Harness To adjust the harness read the separate harness manual, within the harness Note: The patients do need to wear appropriate cloths for the training before putting on the harness.
  • Page 12: Attaching The Harness To The Balancer

    8.3 - Attaching the harness to the balancer Each shoulder strap at the harness has a D-ring that attaches with quick release hooks to the balancer. Adjust the upright posture of the patient by moving the D-ring back and forth on the straps. With the individual adjustment the patient can be prevented of falling over to the front of the back.
  • Page 13: Putting On The Harness For Wheelchair Patients

    The patient is now ready to be unweighted. Before unweighting anyone, be sure the waist belt is snug and that the safety snaps are closed completely. - Place the wheelchair under the Lode BWSS. For treadmills, you may need to use a ramp to place the person under the system.
  • Page 14: Lift A Test Subject From A Wheelchair

    8.6 - Select the weight to be unloaded during the training At this point, the test person is wearing the Harness and is connected to the BWSS. The desired body weight support range should have been selected at this point. When the test person was lifted from a wheel chair.
  • Page 15: Release A Test Person Form The Bwss

    8.7 - Release a test person form the BWSS At this point, the test subject is wearing the Harness and is connected to the BWSS. The desired body weight support range should have been selected at this point. 1. Unlock the fall stop rope.
  • Page 16: Emergency Release Of A Test Person From The Bwss

    8.9 - Emergency release of a test person from the BWSS In some cases a person must be released from the BWSS immediately. Execute the following steps to release a test person immediately form the BWSS. You will need at least two persons to executed this procedure: 1.
  • Page 17: Problem Solving

    1. Turn the system off wait 10 seconds and turn the system back on to reset the internal motor controller. 2. When necessary release the test person from the BWSS. (read 4.9 Emergency release a test person form the BWSS) 3.
  • Page 18: Maintenance

    10 - Maintenance The BWSS is designed mechanically and electronically for durability and accuracy. The load cell in the BWSS needs to be calibrated once a year. The calibration is performed using calibration equipment specially designed by the manufacturer. Cleaning Cleaning The BWSS can be cleaned with a cloth dampened with non-aggressive liquid.
  • Page 19: Accessories

    Harness BWSS Small Harness BWSS Large Harness BWSS Medium, BWSS standard supplied with BWSS Height adjustment 18 Height adjustment 36 cm for BWSS / Mobile cm for BWSS / Mobile Fallstop Fallstop Partnumber: 932824 Partnumber: 932824 Partnumber: 932833 Partnumber: 932833...
  • Page 20: Faq

    Is Lode recommending special disinfection procedures for the Lode products due to COVID-19? Yes, Due to the fact that the scientific research on the spread of Covid-19 is not yet conclusive Lode recommends to disinfect and clean all Lode products and their accessories (cuffs, heart rate belts,...
  • Page 21 14 - Specifications Accuracy Weight support steps 0.1 kg 0.2 lbs Comfort Versatile Multi-Spring Adjustability (VSMA) Min. 0 kg Versatile Multi-Spring Adjustability (VSMA) 80 kg Max. Allowed user weight 160 kg 352.7 lbs Maximum lifting weight 160 kg 352.7 lbs Dimensions Product length (cm) 160 cm...

Table of Contents