M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23

Advertisement

Table of Contents
loading

Summary of Contents for Lode M967900

  • Page 1 M967900 - Manual Angio single Cardio Pulmonary Lode BV Zernikepark 16 NL 9747 AN Groningen The Netherlands t: +31 50 571 28 11 f: +31 50 571 67 46 @: ask@lode.nl www.lode.nl M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23...
  • Page 2: Table Of Contents

    Table of contents 1 - Intended Use 1 - Intended Use 2 - Precautions 2 - Precautions 3 - Contra Indication 3 - Contra Indication 3.1 - Free area 4 - Validity 4 - Validity 4.1 - Description of parts 4.2 - Angio - description of parts bottom side 5 - Installation 5 - Installation...
  • Page 3: Intended Use

    If, at any time during exercise, the test subject feels faint, dizzy, or experiences pain, stop the test and he or she should be consulted by the physician. M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 3/20...
  • Page 4 Inspect the device for worn or loose components prior to use. Tighten / replace any loose or worn components prior to use. Do not place any loose parts or objects on the device before and during use. M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 4/20...
  • Page 5 If the test person feels faint stop exercising immediately. The installation should only be done by authorized persons who follow the installation instructions. Do not spill any liquid over the device. M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 5/20...
  • Page 6 The operator should have knowledge about the Warnings and Cautions before using the device. The (support of the) Control Unit can’t be used as handlebar or support. M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 6/20...
  • Page 7 This device should only be sold by, or under the supervision of authorized persons. Take care that the cable of the control unit cannot get stuck between the cranks when using the ergometer. M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 7/20...
  • Page 8: Contra Indication

    The free area must also include the area for emergency dismount. Where equipment is positioned adjacent to each other the value of the free area may be shared. M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 8/20...
  • Page 9: Validity

    4 - Validity This manual covers all Angio versions starting with the following serial number: 9679xx Angio cpet: S/N 20170001 - Date of issue: 2017-01-31 M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 9/20...
  • Page 10: Description Of Parts

    (rpm). Note: be aware of hands near the display-arm! Tighten the knobs of the display-arm to prevent dropping. M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 10/20...
  • Page 11: Angio - Description Of Parts Bottom Side

    4.2 - Angio - description of parts bottom side 2 - cable connections 6 - Identification plate M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 11/20...
  • Page 12: Installation

    Note. There is place for the thick AC cord in the middle and 2 other thinner cables on the sides. 4. Put the ergometer back in upright position. M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 12/20...
  • Page 13: Terminal - External Control

    6.2 - External control with LEM or LCRM 1. If your device has an optional RS232 or USB port, connect it directly to the PC on which the Lode software is installed. If your device has a RJ45 connector plug connect it with an UTP cable to the special interface connector that was delivered with your device and to a USB port in your PC.
  • Page 14: List Of Symbols Used

    The equipment has a safety earth (ground) connection and must be connected to an earthed (grounded) wall socket. External Input M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 14/20...
  • Page 15: Maintenance

    Opening of the equipment by unauthorized agencies is not allowed and will terminate any claim to warranty. Lode will make available on request circuit diagrams, component part lists, descriptions, calibration instructions, or other information that will assist authorized service personnel to repair those parts of the ergometer that are designated by Lode as repairable by authorized service personnel.
  • Page 16 If you are not certain about the discolouring effect of a cleaning agent, you may try a little on a part of the product that is not visible during normal use. Lode cannot advise a specific cleaning agent, since local recipes may differ. The operator should clean the device (handgrips, cushions) after each exercise test.
  • Page 17: Calibration

    The ergometer should be calibrated once a year. In case any damage is observed of the ergometer Lode B.V. or his representative should be informed in order to execute the necessary repair(s). Service of the ergometer is restricted to factory-trained personnel only.
  • Page 18: Accessories

    Pedal shoes for childen Optimal force Versatile ergometry Handgrip fixation set cycling application Pedal shoes for foot fixation Pedal shoes for children For different leg/arm length Handgrips (pair) Handgrip fixation set M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 18/20...
  • Page 19 USB to Serial converter Partnumber: 226012 Partnumber: 226012 Easy connection USB to Serial converter M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 19/20...
  • Page 20: Specifications

    10 % Set Mode Manual operation mode Preset protocol operation mode External control unit Selfdesigned protocol operation mode Dimensions Product weight 40 kg 88.2 lbs Order info Partnumber 967900 M967900 - MANUAL ANGIO SINGLE SET / VERSION: 1.0 - 2017-02-23 20/20...