INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NIHON KOHDEN 1550K and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Lode NIHON KOHDEN 1550K

  • Page 1 Service documentation Lode BV Zernikepark 16 NL 9747 AN Groningen The Netherlands t: +31 50 571 28 11 f: +31 50 571 67 46 @: ask@lode.nl www.lode.nl INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15...
  • Page 2: Table Of Contents

    1 - List of symbols used 1 - List of symbols used 2 - Validity 2 - Validity 3 - Setup instructions - Nihon Kohden 1550K 3 - Setup instructions - Nihon Kohden 1550K 3.1 - Cable selection 3.2 - Software setup 3.3 - Ergometer setup...
  • Page 3: List Of Symbols Used

    This interface form covers * All Corival ergometers starting with serialnumber 9609xx-20150001 If your treadmill or ergometer is not covered by this interface form please contact Lode service service@lode.nl INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15...
  • Page 4: Setup Instructions - Nihon Kohden 1550K

    Switch off the ergometer and the system. You will need the RS-232 cable (part no. 930730) to establish communication between the Lode ergometer and your system. The lay- out of the cable is as shown in the picture. Connect the RS232 cable to the System.
  • Page 5: Software Setup

    The system is now ready for communication with the ergometer. 3.3 - Ergometer setup 3.3 - Ergometer setup The following protocol needs to be set in the Lode ergometer to communicate with the external equipment. Bicycle ergometer: Ergoline P10 The next chapters describe the procedures to setup these protocols depending on the type of ergometer.
  • Page 6: Setting Up The Hardware

    4 - Setting up the hardware Setting up the hardware Lode ergometers have two ways of connecting external devices. The cable selection depends on the connection possibilities of the external device. The previous chapter tells you which cable you need. This can be a RS232 cable or a USB cable.
  • Page 7: Connecting With The Usb Cable

    If there is no other connection available you can also use a USB cable to connect. 1. You will need an USB A to B cable (Lode part no. 930916) to establish communication between the Lode ergometer and your stress system or ergospirometrysystem.
  • Page 8: Treadmill Or Ergometer With 3.5" Touch Screen Unit

    Swipe with your finger over the screen as indicated with the red arrow in the picture to go to the main menu. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 8/22...
  • Page 9 The menu "Service" is only intended for service purposes to be executed by factory trainend personel. Select "Terminal Settings" to change the communication protocol. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 9/22...
  • Page 10 Select the Communication protocol as desribed in "Ergometer setup" chapter. Select "Default" for the baudrate. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 10/22...
  • Page 11 Select "RS232" or "USB" depending in which cable you have connected to the Lode ergometer. Select the button with the house to go back to the main menu. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 11/22...
  • Page 12 Select "Terminal" to set the ergometer in terminal mode. Your device is now ready for external control. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 12/22...
  • Page 13: Treadmill Or Ergometer With 7" Touch Screen Control Unit

    *Note that for this instruction the graphic view for a Lode Corival is used. The view for a treadmill might difer a bit but the procedure is the same. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 13/22...
  • Page 14 "Terminal". Select the communication protocol ad described in the "Ergometer setup" chapter. Use the scroll button to view the complete list. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 14/22...
  • Page 15 Select "Default" for the baudrate. Select "RS232"or "USB" depending on the type of cable that you connected to the Lode ergometer. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 15/22...
  • Page 16 - Go to the settings menu - Select "Default Start Menu" - Select "Terminal" - Switch the ergometer off and on. It will boot directly into terminal mode. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 16/22...
  • Page 17: Nummerical Control Unit - Treadmill Only

    7 - Nummerical Control Unit - Treadmill only 7 - Nummerical Control Unit - Treadmill only 1. Press and hold key #1 and #2 as shown on "Figure 1" for 10 seconds. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 17/22...
  • Page 18 3. By using the "+" and "-" keys for the grade adjustment #3 you can change the "0" behind terminal to "1" to activate the Terminal mode. 4. Press enter #4 and "prot id" will be shown on the display, for Lode protocol this should be set to "1". Set the protocol as described in the "Ergometer setup" chapter.
  • Page 19: Treadmill Or Bicycle Without Control Unit

    Connect an USB cable from the computer to the USB port of the treadmill or use a RS232 cable from the computer to the RS232 port Start the Lode Service Manager and login with the same account details that you use to login on the website.
  • Page 20 6. Switch the device off, wait 10 seconds 7. Switch the device on The screen as shown left should show up within 5 seconds after powering up. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 20/22...
  • Page 21 Lode ergometer. Select default for the baudrate. Note that Default should always be selected unless different specified by your dealer. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 21/22...
  • Page 22 Your device is now ready for use with the external controller. Note that the device is automatically switched to terminal mode when no manual controller is present at the device itself. INTERFACE SHEET - NIHON KOHDEN 1550K - CPET ERGOMETER / VERSION: 1.0 - 2016-01-15 22/22...