Performance
Check/Calibration
DD
501
TABLE
5-1
Test
Equipment
Required
Description
Minimum
Specifications
Usage
Examples
of
Applicable
Test
Equipment
1.
Test-oscilloscope
system
(dual-trace)
Bandwidth, dc
to
65
MHz;
minimum
deflection
factor,
50 mV/div; accuracy,
with-
in
3%.
Used
throughout
pro-
cedure
as a
waveform
monitor.
a.
Tektronix
465
Oscilloscope,
with 2
each P6065A, 10X
probes.
b.
Tektronix
7603
Oscilloscope
with
7A26
Amplifier,
7B50 Time
Base,
and
2
each P6053B, 10X
Probes.
2.
Sine-wave
generator
Output frequency
range,
50
kHz
to
100
MHz;
output
am-
plitude
range, 0.030
V
to
2
V.
Used
throughout
pro-
cedure
as
signal
source.
a.
Tektronix
Type
191
Constant-
Amplitude
Signal Generator.
3.
Square-wave
generator
Output frequency
range,
1
kHz
to
100
kHz; output
am-
plitude
range,
0.5
V
to
1
V.
Used
throughout
pro-
cedure
as
signal
source.
a.
Tektronix
PG
501 Pulse
Gen-
erator.
b.
Tektronix
FG
501 Function
Generator.
4.
Digital
counter
Frequency
range, 0
to
1
MHz;
accuracy,
±1
count.
Capable
of
ratio
measurement.
Delay count accuracy
check.
a.
Tektronix
DC
503
Universal
Counter.
5.
Power module
Tektronix
TM
500-Series
Used
throughout
pro-
cedure
to
supply
power
to
the
DD
501.
a.
Tektronix
TM
503
Power
Module.
6.
Plug-in extension'
Tektronix
TM
500-
and
5000-
Series plug-in extender.
Allows access
to
inter-
a.
Tektronix
Part
Number
nal
adjustments
and
test
067-0645-02
Calibration
points
in
the
DD
501.
Fixture.
7.
Input
RC
normalizer
Time
constant,
1
MO
X
20
pF;
connector,
BNC.
Used
to
normalize
the
start
and
events
input
RC
product.
a.
Tektronix
Part
Number
067-0538-00
Calibration
Fixture.
8.
Dual
input
coupler
Connectors,
BNC;
cable
lengths,
matched
within
0.1
inch.
Insertion of identical
signals
to
two
inputs
simultaneously.
a.
Tektronix
Part
Number
067-0525-00
Calibration
Fixture.
9.
Termination
(2
each)
Impedance, 50
0;
accuracy,
±2%;
connectors,
BNC.
Used
throughout
pro-
cedure
to
properly
ter-
minate output
signals.
a.
Tektronix
Part
Number
011-0049-01.
10.
Adapter
Probe
tip
to
BNC
male.
Com-
patible
with
P6053B and
P6065A
probes.
Trigger
sensitivity
check.
a.
Tektronix
Part
Number
013-0084-01.
11.
Adapter
T
connector; connectors,
BNC.
Delay count accuracy
check. Recycle
rate
and
throughput
time check.
a.
Tektronix
Part
Number
103-0030-00.
Used
(or
calibration only.
NOT
used
for
performance
check.
5-2
REV
B,
JUN
1978

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DD 501 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents