OPERATOR'S CHECKS
AND ADJUSTMENTS
INTRODUCTION
To verify the operation and accuracy of your instru-
ment before making measurements, perform the following
check
and
adjustment
procedures.
If adjustments are
required beyond the scope of these operator's checks
and adjustments, refer the instrument to a qualified service
technician.
Before
proceeding with these instructions, refer to
"Preparation for Use" (Section 2).
Verify that the POWER switch is OFF (push button
out), then plug the power cord into the power outlet.
If indications specified in these procedures cannot be
obtained,
refer the
instrument to
a qualified service
technician.
INITIAL SETUP
1. Press in the POWER switch button (ON) and allow
the instrument to warm up (20 minutes is recommended
for maximum accuracy).
2. Set instrument controls to obtain a baseline trace:
Display
READOUT
INTENSITY
Midrange between ""MIN""
and fully clockwise
FOCUS
Midrange
Vertical
POSITION
Midrange
MODE
CH 1
BW LIMIT
Off (button out)
CH 1 VOLTS/DIV
10 mV
CH 1 Input Coupling
1 MQ GND
Horizontal
A AND B SEC/DIV
Locked together at 1 ms
SEC/DIV VAR
Calibrated detent
POSITION
Midrange
10X MAG
Off (button out)
Trigger
HOLDOFF
Fully counterclockwise
LEVEL
Midrange
MODE
AUTO LVL
SOURCE
VERT
COUPLING
DC
SLOPE
+
3. Adjust the INTENSITY and READOUT INTENSITY
controls for desired display and readout brightness and best
trace definition.
4. Adjust
the
Vertical
and
Horizontal
POSITION
controls to position the trace within the graticule area.
TRACE ROTATION ADJUSTMENT
1. Preset instrument controls and obtain a baseline trace
as described in "Initial Setup."
2. Use the Channel 1 POSITION control to move the
baseline trace to the center horizontal graticule line.
NOTE
Normally, the resulting trace will be parallel to the
center horizontal graticule line, and the Trace Rota-
tion adjustment should not be required.