Xylem wtw MIQ/R6 Operating Manual page 14

Iq sensor net
Table of Contents

Advertisement

Installation
Connecting lines
to the terminal
strip
Fig. 3-1
14
All manuals and user guides at all-guides.com
1
Open the module.
2
2 2
1
X18
X17
X16
X15
X14
240 VAC
240 VAC
2 A AC
2 A AC
3
R6
R5
Segment 2
Terminal strip with the relay connections
2
Screw the cable gland (pos. 1 in Fig. 3-1) with the sealing ring (pos. 2)
into the module housing.
3
Loosen the coupling ring (pos. 3 in Fig. 3-1).
4
Feed the line through the cable gland in the module housing.
5
Connect the wires to the terminal strip. While doing so, pay attention to
the specifications on the label located under the terminal strip.
6
Tighten the coupling ring (pos. 3 in Fig. 3-1).
WARNING
No free wires are allowed to project into the housing. Oth-
erwise, there is a danger that areas safe to contact could
come into contact with dangerous voltages. Always cut
off any wires that are not in use as closely as possible to
the cable gland.
X13
X12
X11
X10
X9
X8
X7
X6
240 VAC
240 VAC
240 VAC
240 VAC
2 A AC
2 A AC
2 A AC
2 A AC
R4
R3
R2
R1
SENSORNET 2
Segment 1
Relay output
ba75541d07
MIQ/R6
X5
X4
X3
X2
X1
ON
EIN
OFF
AUS
SENSORNET 1
03/2018