THE FOLLOWING
TESTS SHOULD BE DONE ONLY
IF THE LISTEN FUNCTION
WORKS PROPERLY. AFTER
ANY
REPAIR, RETURN TO TALK TEST SECTION OF
MAIN TREE.
IF THE
SYSTEM IS HUNG UP, TESTS 6 AND 7 MAY INDICATE STUCK
HANDSHAKE
LINES. NOTE THAT
IF
CHKA4U5) IS CORRECTLY
DECODING
TALK
COMMANDS, A4DS2 WILL REFLECT TALK STATE REGARDLESS OF FRONT - PANEL TLK LED,
I. TALK MONITOR SEE FIGURE 5-40
2. TRANSMIT ENABLE MONITOR SEE FIGURE 5-40
3. HP-18
INTERRUPT
LOOP SEE FIGURE 5-38
USING
THE PROGRAMS LISTED IN FIGURE 5 - 40, CHECK THAT AN INTERRUPT
OCCURS FOR
BOTH TALK
AND TRANSMIT ENABLE.
4. LOAD/SHIFT PROTECTION CIRCUIT SEE FIGURE 5-39
5. TALK HANDSHAKE SEE FIGURE 5-43
6. MULTIPLEXERS AND DATA SHIFTER/OUTPUT SUFFER CIRCUITS SEE FIGURE 5-42
7. TRANSCEIVERS
SEE FIGURE 5-37
Figure 5-33.
HP-1B Talk Tests
THE FOLLOWING TESTS SHOULD BE DONE ONLY IF THE LISTEN AND TALK FUNCTIONS WORK PROPERLY.
AFTER ANY REPAIR, RETURN TO LISTEN TEST SECTION OF MAIN TREE.
1.
SRQ TRANSCEIVER SEE FIGURE 5-37.
IT IS NECESSARY ONLY TO CHECK SRQ TRANSCEIVER
2.
S.P.T.E. MONITOR SEE FIGURE 5 44.
3.
HP IP INTERRUPT LOOP SEE FIGURE 5 38.
USE THE S.P.T.E. MONITOR PROGRAM LISTED IN FIGURE 5-44.
4.
LOAD/SHIFT PROTECTION SEE FIGURE 5 39.
USE THE S.P.T.E. MONITOR PROGRAM LISTED IN FIGURE 5-44.
5.
DATA SHIFTER SEE FIGURE 5-42.
Figure 5-34.
Serial Poll Tests