HP 6034A User Manual page 92

System dc power supply
Hide thumbs Also See for 6034A:
Table of Contents

Advertisement

ENSURE THAT THERE IS NO
LOAD ON 6034A
(
2
)
PROGRAM-
0'*
wrt
7 0 5?"M 1
10
OUTPUT 70S
>"Ml
G"
G
20
END
(3)
(4)
A3UI3-6 SHOULD BE LOW
PROGRAM
0 S
w r t
705?"M2
10 OUTPUT 705
"M2 G,!
G
20 END
READBACK COMPARATORS TEST
REFERENCES CAN BE CONTROLLED WITH RPG, OUTPUT CAN BE
VARIED FROM OPEN TO SHORT CIRCUIT. OUTPUT OF COMPARATOR
SHOULD BE HIGH WHEN NON-INVERTING INPUT ( + ) IS MORE
POSITIVE THAN INVERTING INPUT (-).
FAULT HI
CHECK A3U11 AND ASSOCIATED COMPONENTS. USE S.A. (FIG 5-15)
TO CHECK A3U15.
FAULTS H3 and 4
CHECK A3U8 AND ASSOCIATED COMPONENTS. USE S.A. (FIG 5-15)
TO CHECK A3U6.
OVP TEST
GO TO FIGURE 5-35
HP-IB TEST
GO TO FIGURE 5-38
Figure 5-31.
Self Test Failures (Sheet 3 of 3)
5-46

Advertisement

Table of Contents
loading

Table of Contents