Cleaning And Preventive Maintenance; Daily; Weekly; Monthly - Human HumaMax Micro User Manual

Table of Contents

Advertisement

ClEANINg AND prEvENTIvE mAINTENANCE

9 CLEANING AND PREVENTIVE MAINTENANCE

9.1 Daily

No daily cleaning is required, except in the case of accidental
glass breakage or a large amount of spillage in the bowl.

9.2 Weekly

Clean the bowl and the accessories with a cotton wool pad
dipped in 70% alcohol solution.

9.3 Monthly

Maintenance for drive shaft. You can wipe the drive shaft
with soft cloth, and then apply a thin coat of silicon grease.
All cleaning should
be performed with
the centrifuge discon-
nected from the power
outlet.
Never use metallic
implements to clean
the rotor or the inserts.
After cleaning the acces-
sories, always rinse them
with clean water and
dry.
Don´t clean the Lid
with alcohol to pre-
vent dolling