Table 2. Required Test Equipment
Model 532A/B
Equipment
Critical Specifications
Recommended HP Model
Oscilloscope
Osciiloscynchroscope
130C
Adapter (2)
Coax to waveguide
Band of interest
281A
Adapter
(Waveguide to waveguide)
Bands of interest
MP292B (M-band tests only)
Coaxial Attenuator
12 dB in 1 dB steps 200 MHz
355C
Waveguide Attenuator
Continuously variable to 20 dB
375A
Crystal Detector
Sensitivity: 0.3 mV/|uW
424A (423A with 281A adapter
may be substituted for G-, J-,
H-, and X-band tests)
Thermistor Mount
Band of interest. Matches Power Meter
486A
Power Meter
Matches Thermistor Mount
431B/C
Electronic Counter
Frequency: to 200 MHz
5245L with 5253B plug-in
Transfer Oscillator
Stable 200 MHz oscillator with 3.95 GHz
mixer and video amplifier
540B
Directional Coupler
Band of interest: 3 dB
752A
Harmonic Mixer
Input Frequency: 12.4-18.0 GHz
Output: 0.4 mV with 0 dBm input
932A (M- and P-bandtests only)
Sweep Oscillator
Band of interest
Leveling: remote, -V proportional to RF
Modulation: INT AM 1000 Hz square wave
8690 mainframe with:
G- and J-band H01-8693B
H- and X-band H01-8694A/B
M-band 8695A (special)
P-band 8695A
Low-Pass Filter
Passband: Band of interest
Reject: Harmonics
G-band Microphase LTP 6000
J-band Microphase LTP 8200
H-band Microphase LTP 10500
X-band HP X362A
M-band HP M362A
P-band HP P362A
Isolator
Isolation: > 30 dB
G-band Ferrotec I-152L
J-band Ferrotec I-153L
H-band Ferrotec I-154L
X-band Ferrotec I-155L
M-band Ferrotec I-744L
P-band Ferrotec I-156L
Table 3.
Test Card
i
Hewlett Packard Model 532A/B
Date
Frequency Meter
Tested by
FREQUENCY DIAL CALIBRATION
Maximum Dial Error
%
Specification
%
POWER DIP
High End of Band
dB Specifications > 1 dB
Center of Band
dB Specifications > 1 dB
Low End of Band
dB Specifications > 1 dB
00245-5