Removing The A5 Daughter Or A6 Measurement Assemblies - HP EPM E4418A Service Manual

Hide thumbs Also See for EPM E4418A:
Table of Contents

Advertisement

Removing the A5 Daughter or A6 Measurement Assemblies

Note
4. Position sensor flex
circuit into connector
in direction of arrow
HP EPM-441A/442A Service Guide
1. Disconnect the flex circuit from measurement assembly.
Care should be taken when disconnecting the flex circuit from the
measurement assembly.
The flex circuit assembly is released by pushing the connector tab
forward and lifting.
To replace the flex circuit, loop it as shown on page 5-16, and
connect the flex circuit as shown in the figures below.
2. Slide out the measurement assembly from the side of the power
meter.
3. The daughter assembly is removed vertically.
1. Connector closed
Assembly and Disassembly Guidelines
A5
2. Slide connector clamp
in direction of arrow
5.
6. Push clamp closed in
direction of arrow
Replaceable Parts
A6
3. Rotate connector
clamp in direction of
arrow
5-11