Digital Display
Pulse Amplitude:
Calculated from measuring
the diff erence between the
peak and minimum pulse
pressure, averaged over 5
seconds.
Pulse Frequency Rate:
represents the number of
pulses per minute.
NOTE: Display numbers are
for reference only.
PAC-SAL-LIT-734 (1)
2024-01
Mean Airway Pressure
(MAP): is the average
pressure (amplitude +
CPAP) in the lungs over 5
seconds.
Usage Timer: total usage
D
time of the current session
Troubleshooting
Digital display off
The IPV® pressure must be >2.5 cmH
before the display turns on. Consider occluding the end of
the Phasitron, allowing the device to measure at least 2.5
cmH
O to wake digital display. Check connections and/or
2
digital display batt ery.
Nebulizer not aerosolizing
Ensure there is liquid present in the nebulizer cup and check
for fl ow out of the nebulizer baff le. Disconnect the yellow
tubing and verify fl ow. If fl ow is present, turn the Gas
Source knob to the OFF position and reconnect the yellow
tubing to the nebulizer bowl. If the issue persists, contact
product support.
Slow or no percussions
Check that the inlet gas source is properly connected.
Check the device to ensure the Gas Source is turned on to
the correct source, make sure Amplitude is not in the OFF
position, and adjust Frequency knob for higher rate. If the
issue persists, contact service.
NOTE:
Refer to the Instruction Manual for the
IPV® 1 for warnings, cautions, and additional
information. Always follow hospital protocol and
use good clinical judgement when delivering IPV®
therapy. Absolute contraindications for IPV®
usage include untreated tension pneumothorax
and/or an untrained or unskilled operator.