Siemens SITRANS LR110 Operating Instructions Manual page 47

Radar transmitters
Table of Contents

Advertisement

Diagnostics and troubleshooting
10.4 Status messages according to NE 107
Failure
Text message
F013
no measured
value available
F017
Adjustment span
too small
F025
Error in the
linearization
table
F036
No operable
software
F040
Error in the
electronics
F080
General software
error
F105
Determine
measured value
F260
Error in the
calibration
F261
Error in the
transmitter
settings
F265
Measurement
function
disturbed
Function check
Text message
C700
Simulation active
40
Code
No measured value in the
switch-on phase or during
operation
Adjustment not within
specification
Index markers are not
continuously rising, for
example illogical value pairs
Checksum error if software
update failed or aborted
Limit value exceeded in signal
processing
Hardware error
General software error
The transmitter is still in
the switch-on phase, the
measured value could not yet
be determined
Checksum error in the
calibration values
Error in the EEPROM
Error during setup
Auto false echo suppression
faulty
Error when carrying out a
reset
Program sequence of the
measuring function disturbed
Code
A simulation is active
Cause
Check or correct installation
and/or parameter settings
Clean the antenna system
Change adjustment according
to the limit values (difference
between min. and max.
≥ 10 mm)
Check linearization table
Delete table/Create new
Repeat software update
Send transmitter for repair
Restart transmitter
Send transmitter for repair
Restart transmitter
Wait for the end of the switch-
on phase
Duration up to 3 minutes
depending on the
measurement environment
and parameter settings
Send transmitter for repair
Repeat setup
Carry out a reset
Device restarts automatically
Cause
Finish simulation
Rectification
State in CMD 48
Byte 5, Bit 0 of
Byte 0 ... 5
Byte 5, Bit 1 of
Byte 0 ... 5
Byte 5, Bit 2 of
Byte 0 ... 5
Byte 5, Bit 3 of
Byte 0 ... 5
Byte 5, Byte 5,
Bit 4 of
Byte 0 ... 5
Byte 5, Byte 5,
Bit 5 of
Byte 0 ... 5
Byte 5, Byte 5,
Bit 6 of
Byte 0 ... 5
Byte 4, Bit 0 of
Byte 0 ... 5
Byte 4, Bit 1 of
Byte 0 ... 5
Byte 4, Bit 3 of
Byte 0 ... 5
Rectification
State in CMD 48
Simulation
Active" in
Operating Instructions, 10/2023, A5E49829006