9.2.6 Maintenance setup (for administrators only)
Tap MENU > SETUP > MAINTENANCE to set up the following content:
•
Standby: set up the wait time (between 30 and 60 min) before the analyzer enters the
standby status.
•
Probe cleanser maintenance: set up the start time (between 00:00 and 23:59) and the
reminding interval (between 5 and 60 min) for the daily probe cleanser maintenance.
9.2.7 Reagent setup
Tap MENU > SETUP > REAGENT SETUP to replace the reagents. You can also directly select
REAGENT SETUP from the utility button area.
See section "10.2 Reagent management" for more details.
9.2.8 Gain setup (for administrators only)
When the analyzer reports the HGB blank voltage abnormal error and you cannot remove the
error by pressing the "Remove Error" button, adjust the HGB gains to correct the HGB blank
voltage.
Tap MENU > SETUP > GAIN SETUP > WB to enter the "Gain Setup" screen of whole blood.
Adjust the HGB default gain in the "HGB" set text box until the HGB blank voltage is in the
range of [4.30 – 4.50].
9.2.9 Auto startup/shutdown time (for administrators only)
Administrators may set up the auto startup and shutdown time for the analyzer on the "Auto
Startup/Shutdown" screen. When you have set the auto startup/shutdown time, the analyzer
automatically starts and shutdown at the set time.
1. Tap MENU > SETUP > AUTO STARTUP/SHUTDOWN to enter the corresponding
screen.
2. Select the required dates by taping on "Auto Startup" and "Auto Shutdown".
Vers.: 20230710ENG
Page 48