The countertop where the analyzer is placed shall be able to withstand at least 40 kg
of weight.
4.2.2 Power requirements
WARNING
•
Make sure the analyzer is properly grounded.
•
Before turning on the analyzer, make sure the input voltage meets the requirements.
CAUTION
•
Using a pin-board may bring the electrical interference and the analysis results may be
unreliable. Please place the analyzer near the electrical outlet to avoid using the pinboard.
•
Please use the original power cable shipped with the analyzer. Using other power cable
may damage the analyzer or cause unreliable analysis results.
Power specification table:
Voltage
(100 V–240 V~) ± 10%
4.2.3 General environment
•
Optimal operating temperature: 50°F~86°F (10°C~30°C)
•
Optimal operating humidity: 30%~85%
•
Atmospheric pressure: 70 kPa~106 kPa
•
The environment shall be as free as possible from dust, mechanical vibrations, loud
noises, and electrical interference.
•
It is advisable to evaluate the electromagnetic environment prior to operation of this
analyzer.
•
Do not use this analyzer in close proximity to sources of strong electromagnetic
radiation (e.g., unshielded intentional RF sources), as these may interfere with the
proper operation.
•
Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close, such as a lab centrifuge.
•
Do not place the analyzer in direct sunlight or in front of a source of heat or drafts.
•
The environment must be ventilated.
•
Install on a level surface.
•
For indoor use only.
Vers.: 20230710ENG
Page 25
Input Power
300 VA
Frequency
(50 Hz/60 Hz) ± 1 Hz