The countertop where the analyzer is placed shall be able to withstand at least 40 kg
of weight.

4.2.2 Power requirements

WARNING
Make sure the analyzer is properly grounded.
Before turning on the analyzer, make sure the input voltage meets the requirements.
CAUTION
Using a pin-board may bring the electrical interference and the analysis results may be
unreliable. Please place the analyzer near the electrical outlet to avoid using the pinboard.
Please use the original power cable shipped with the analyzer. Using other power cable
may damage the analyzer or cause unreliable analysis results.
Power specification table:
Voltage
(100 V–240 V~) ± 10%

4.2.3 General environment

Optimal operating temperature: 50°F~86°F (10°C~30°C)
Optimal operating humidity: 30%~85%
Atmospheric pressure: 70 kPa~106 kPa
The environment shall be as free as possible from dust, mechanical vibrations, loud
noises, and electrical interference.
It is advisable to evaluate the electromagnetic environment prior to operation of this
analyzer.
Do not use this analyzer in close proximity to sources of strong electromagnetic
radiation (e.g., unshielded intentional RF sources), as these may interfere with the
proper operation.
Do not place the analyzer near brush-type motors, flickering fluorescent lights, and
electrical contacts that regularly open and close, such as a lab centrifuge.
Do not place the analyzer in direct sunlight or in front of a source of heat or drafts.
The environment must be ventilated.
Install on a level surface.
For indoor use only.
Vers.: 20230710ENG
Page 25
Input Power
300 VA
Frequency
(50 Hz/60 Hz) ± 1 Hz