2415-UW ULTRASONIC CLEANER
6.
Repair
REPLACE THE INJECTOR
Do not press down on the detergent flow
valve when loosening or tightening the
joint to the injector. Undue pressure can
break the joint.
Remove the fauity injector.
injector port (see Figure 6-5),
flow valve and the injector.
2.
Before connecting the detergent flow vaive to the
new injector, cut and discard the hose barb from the
3.
Chamber (bevel) the inside diameter edge of the
4.
Usenowrenches when installing the injector. Teflon
REMOVE
HOSE BARB
DO NOT OVERTIGHTEN
OTHERWISE FITTING WILL
BREAK
CUT
HERE
(CHAMFER
THE CUT
EDGE)
FIGURE 6-5. REPLACE THE INJECTOR
tape and hand tightening are all that is necessary.
350593