Racal Instruments 9057 Technical Manual page 42

Frequency period meter
Table of Contents

Advertisement

9057/9059
Equipment Required:
(a)
Signal Generator and 50 ohm pad.
(b)
Electronic Voltmeter.
(c)
15pF Capacitor.
Procedure
(1)
On the 9057 (or 9059 Channel 'A') set the Sensitivity Switch to 0.01 -lOY.
(2)
Connect the signal generator to the input of the electronic voltmeter with
a 15pF capacitor shunting the 50 ohm input of the voltmeter this simulates
the instrument input conditions.
(3)
Set the signal generator output to exactly 1 OmV on the voltmeter. Note
the exact reading on the signal generator control.
(4)
Connect the signal generator via the BNC 'T' piece with the 50 ohm load
to the input socket of the 9057 or (9059 Channel 'A').
(5)
Set the 9057/9059 Function Switch to FREQUENCY or FREQUENCY 'A'.
(6)
Set the signal generator frequency and the 9057 Push Button Range Switch
to the settings listed in Table
3.
Check that in each instance a stable and
accurate display is obtained.
TABLE
3
9057 and 9059 Channel 'A' Sensitivity Check
Push Button Range
Input Frequencies
Remarks
1 Os kHz (4)
10Hz
.....
Sensitivity switch
1s
kHz (3)
100Hz, 100kHz
set to 0.01 -lOY
1 OOms MHz (2)
100 kHz, 1 MHz
lOms MHz (1)
1 MHz,
10 MHz
Input Level to be
Continue the
>
not higher than
frequency sweep
lOmV r.m.s.
in 10 MHz steps
to 80 MHz
~
{7)
On Model 9059 transfer the signal generator to the Channel 'B' input
socket (50 ohm load not required).
6-6